← Product Code [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH) · K061364

# PILOT POSTERIOR LUMBAR PLATING SYSTEM (K061364)

_Life Spine · MNH · Sep 5, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K061364

## Device Facts

- **Applicant:** Life Spine
- **Product Code:** [MNH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH.md)
- **Decision Date:** Sep 5, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Internal fixation implants are load-sharing devices intended to stabilize and hold an alignment in place until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing. The PILOT™ Plating System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization and immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities. The PILOT™ Posterior Lumbar Plating System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) spondylolisthesis, (3) trauma (i.e., fracture or dislocation) (4) spinal stenosis, (5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis) (6) tumor, (7) pseudoarthrosis, (8) failed previous fusion.

## Device Story

PILOT™ Posterior Lumbar Plating System; load-sharing internal fixation implant. Components: bone screws, plates, washers, nuts, collets. Used in thoracic, lumbar, and sacral spine to stabilize/immobilize spinal segments as adjunct to fusion. Implanted by surgeons in clinical settings. Provides mechanical support during bone healing; not intended to replace body structures or bear full body weight if bone healing is incomplete. Benefits: stabilization of spinal segments in patients with acute/chronic instabilities or deformities.

## Clinical Evidence

Bench testing only. Biomechanical testing conducted in accordance with ASTM F1717 to demonstrate substantial equivalence.

## Technological Characteristics

Materials: Medical grade titanium alloy (ASTM F136, Ti 6AL-4V-ELI). Components: bone screws, plates, washers, nuts, collets. Mechanical fixation system. Sold non-sterile.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the words "Life Spine" in a stylized font. The word "Life" is in bold, while the word "Spine" is in a thinner font. There is a curved line underneath the word "Spine", and a circled R symbol to the right of the word "Spine".

761264

#### 510(k) Summary of Safety and Effectiveness

| 510(k) Summary - PILOT™ Posterior Lumbar Plating System |                                                                                                                                        | SEP - 5 2006 |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Submitted By:                                           | Life Spine<br>2400 Hassell Road, Suite 370<br>Hoffman Estates, IL 60195<br>Telephone: 847-884-6117<br>Fax: 847-884-6118                |              |
| 510(k) Contact:                                         | Erin Malloy<br>Life Spine<br>2400 Hassell Road, Suite 370<br>Hoffman Estates, IL 60195<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |              |
| Date Prepared:                                          | August 2, 2006                                                                                                                         |              |
| Trade Name:                                             | PILOT™ Posterior Lumbar Plating System                                                                                                 |              |
| Common Name:                                            | Posterior Pedicle screw system                                                                                                         |              |
| Classification:                                         | 21 CFR 888.3070<br>Pedicle Screw Spinal System<br>Class II                                                                             |              |
| Device Product Code:                                    | MNI: Orthosis, Spinal, Pedical Fixation<br>MNH: Orthosis, spondylolisthesis spinal fixation                                            |              |
| Predicate Device:                                       | Predicate device information is included in this premarket<br>notification                                                             |              |

#### Device Description:

When implanted in the thoracic, lumbar, and/or sacral spine, the PILOT™ Plating System provides additional support during spinal fusion. The PILOT™ Plating System consists of bone screws, plates, washers, nuts, and collets in various shapes and sizes. The PILOTTM Plating System is manufactured from medical grade titanium and will be sold non-sterile.

#### Intended Use of the Device:

Internal fixation implants are load-sharing devices intended to stabilize and hold an alignment in place until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The PILOT™ Plating System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a handwritten sequence of characters, "K061364". The characters are written in black ink on a white background. The handwriting is somewhat stylized, with the numbers being slightly slanted.

and immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities.

The PILOT™ Postcrior Lumbar Plating System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) spondylolisthesis, (3) trauma (1.e., fracture or dislocation) (4) spinal stenosis, (5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis) (6) tumor, (7) pseudoarthrosis, (8) failed previous fusion,

#### Material:

The PILOT™ Plating System is manufactured from medical grade titanium alloy described by ASTM F136 (Ti 6AL-4V-ELI).

## Performance Data:

Biomechanical testing in accordance with ASTM F1717 was conducted to demonstrate substantial equivalence.

## Substantial Equivalence:

The Life Spine PILOT™ Plating System was shown to be substantially equivalent to a previously cleared device in indications for use, design, function, and materials used.

Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo features a stylized eagle with three lines forming its wings and head. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Life Spine % Ms. Erin Malloy Project Engineer 2400 Hassell Road, Suite 370 Hoffman Estates, Illinois 60195

SEP - 5 2006

Re: K061364

Trade/Device Name: PILOTTM Posterior Lumbar Plating System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: August 2, 2006 Received: August 4, 2006

Dear Ms. Malloy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Erin Malloy

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbara Buchund
for

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Life Spine. The words "Life Spine" are written in a bold, sans-serif font. A curved line underlines the word "Spine". A registered trademark symbol is located to the right of the word "Spine".

## Indications for Use

510(k) number (if known):

# Device Name: PILOT™ Posterior Lumbar Plating System

Internal fixation implants are load-sharing devices intended to stabilize and hold an alignment in place until normal healing occurs. Implants are not intended to replace anguinent in practures or bear the weight of the body in the presence of incomplete bone healing.

The PILOT™ Posterior Lumbar Plating System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally posterior poents. It provides stabilization and immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities.

The PILOT™ Posterior Lumbar Plating System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) spondylolisthesis, (3) trauma (i.e., fracture or dislocation) (4) spinal stenosis, (5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis) (6) tumor, (7) pseudoarthrosis, (8) failed previous fusion.

Prescription Use x (Part 21 CFR 801 Subpart D) And/Or

Over-the-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Division of General, Restorative. and Neurological Devices

510(k) Number K061364

page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K061364](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MNH/K061364)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).