It is intended that this device, in any system configuration be removed after development of solid fusion mass. Hook component indications are limited to T7-L5. Sacral screw indications are limited to the sacrum only. 1) The ZODIAC Polyaxial Pedicle Screw System, when used as a hook and sacral screw fixation system (nonpedicle screw) is intended for: a. Patients having fractures of the thoracic and lumbar spine. b. Patients having deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolisthesis, and acute pars fracture allowing spondylolisthesis). 2) The ZODIAC Polyaxial Pedicle Screw System, when used as a pedicle screw system in the thoraco-lumbo-sacral region of the spine is intended for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). 3) In addition, the ZODIAC Polyaxial Pedicle Screw System, when used as a pedicle screw fixation system is intended for: a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint. 4) The ZODIAC Polyaxial Pedicle Screw System, when used as a laminar hook and bone screw system is intended for: a. Patients having fractures of thoracic and lumbar spine. b. Patients having thoracolumbar deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolistheses and acute pars fracture allowing spondylolisthesis).
Device Story
ZODIAC Polyaxial Pedicle Screw System is a spinal fixation construct; includes polyaxial pedicle screws, laminar hooks, sacral screws, and adjustable transverse connectors. Used by surgeons in clinical/OR settings to provide stabilization and torsional rigidity during spinal fusion procedures. Components implanted to support spine until solid fusion mass develops; device intended for removal post-fusion. System accommodates varying patient anatomy through range of sizes. Clinical benefit includes stabilization of spinal segments in patients with fractures, deformities, or degenerative conditions. Surgeon selects components based on patient pathology; constructs are assembled intraoperatively to achieve desired fixation.
Clinical Evidence
Bench testing only. Mechanical and dynamic testing performed on transverse connectors within spinal constructs to evaluate performance against predicate devices. No clinical data provided.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V ELI, ASTM F136) or stainless steel (ASTM F138). Components: Polyaxial pedicle screws, laminar hooks, sacral screws, adjustable transverse connectors. Mechanical fixation system; no energy source or software.
Indications for Use
Indicated for patients with thoracic/lumbar/sacral spine fractures, deformities (scoliosis, kyphoscoliosis), spondylolisthesis (isthmic, degenerative, acute pars fracture), degenerative spondylolisthesis with neurologic impairment, dislocation, spinal tumor, or failed previous fusion. Hook use limited to T7-L5; sacral screws limited to sacrum. Requires autogenous bone graft for specific spondylolisthesis cases.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
K033090 — ICON POLYAXIAL PEDICLE SCREW SYSTEM · Alphatec Mfg., Inc. · Mar 26, 2004
K042673 — ZODIAC POLYAXIAL SPINAL FIXATION SYSTEM · Alphatec/Nexmed · Oct 27, 2004
K071890 — ZODIAC 4.0 POLYAXIAL SPINAL FIXATION SYSTEM · Alphatec Spine, Inc. · Nov 9, 2007
K093077 — ZODIAC POLYAXIAL SPINAL FIXATION SYSTEM · Alphatec Spine, Inc. · Dec 23, 2009
K051286 — ZODIAC SPINAL FIXATION SYSTEM · Alphatec/Nexmed · Jun 17, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
### K061321 (P9 | of 2) ZODIAC™ Polyaxial Pedicle Screw System 510(k) SUMMARY
July 5, 2006
Company: Alphatec Spine, Inc. 2051 Palomar Airport Road#100 Carlsbad, CA 92011 USA Telephone: (760) 431-9286 Fax: (760) 431-9132
JUL - 7 2006
| Contact Person: | Paula Morgan, Director of Regulatory Affairs |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Proprietary Name: | ZODIAC Polyaxial Pedicle Screw System |
| Common Name: | Pedicle Screw Spinal System |
| Classification Name: | Spinal Interlaminal Fixation Orthosis (888.3050)<br>Spondylolisthesis Spinal Fixation Device (888.3070)<br>Pedicle Screw Spinal System (888.3070) |
#### Product Description:
Adjustable bridges function as transverse connectors. Transverse connectors are used in spinal constructs to add torsional rigidity. The devices come in a range of sizes to accommodate the patient's anatomy. The transverse connectors will be used as part of the ZODIAC Polyaxial Pedicle Screw System. Components are manufactured from titenium alloy, Ti-6AI-4V ELI (ASTM F 136) or stainless steel (ASTM F138).
#### Indications for Use:
It is intended that this device, in any system configuration be removed after development of solid fusion mass. Hook component indications are limited to T7-L5. Sacral screw indications are limited to the sacrum only.
- l) The ZODIAC Polyaxial Pedicle Screw System, when used as a hook and sacral screw fixation system (nonpedicle screw) is intended for:
- Patients having fractures of the thoracic and lumbar spine. a.
- b. Patients having deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity).
- c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolisthesis, and acute pars fracture allowing spondylolisthesis).
7.1
{1}------------------------------------------------
# ZODIAC™ Polyaxial Pedicle Screw System 510(k) SUMMARY July 5, 2006
- 2) The ZODIAC™ Polyaxial Pedicle Screw System, when used as a pedicle screw system in the thoraco-lumbo-sacral region of the spine is intended for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis).
- 3) In addition, the ZODIAC™ Polyaxial Pedicle Screw System, when used as a pedicle screw fixation system is intended for:
- a. Patients receiving only autogenous bone graft.
- b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint.
- 4) The ZODIAC™ Polyaxial Pedicle Screw System, when used as a laminar hook and bone screw system is intended for:
- a. Patients having fractures of thoracic and lumbar spine.
- b. Patients having thoracolumbar deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity).
- c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolistheses and acute pars fracture allowing spondylolisthesis).
#### Substantial Equivalence:
The ZODIAC™ Polyaxial Pedicle Screw System is substantially equivalent to the following predicate devices:
| Trade/Proprietary Name | Manufacturer | Clearance |
|--------------------------|--------------|-----------|
| UCR Spinal System | SeaSpine | K043232 |
| Moss-Miami Spinal System | Depuy Inc. | K982320 |
| Universal Spine System | Synthes | K022949 |
#### Performance Data:
Mechanical and dynamic testing of the transverse connectors as part of a spinal construct using the ZODIAC™ Polyaxial Pedicle Screw System was performed. The test results demonstrate that the mechanical performance of the ZODIAC™ Polyaxial Pedicle Terrew System is at least comparable to, if not better than, those of the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 7 2006
Alphatec Spine, Inc. % Paula Morgan Director of Regulatory Affairs 2051 Palomar Airport Road, Suite 100 Carlsbad, California 92011
Re: K061321
Trade/Device Name: ZODIAC Polyaxial Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNI, MNH, KWP Dated: June 9, 2006 Received: June 13, 2006
Dear Ms. Morgan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Barbara Buchup
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
#### 510(k) Number (if known): K061321
## Device Name: ZODIAC Polyaxial Pedicle Screw System
#### Indications for Use:
It is intended that this device, in any system configuration be removed after development of solid fusion mass. Hook component indications are limited to T7-L5. Sacral screw indications are limited to the sacrum only.
- 1) The ZODIAC Polyaxial Pedicle Screw System, when used as a hook and sacral screw fixation system (nonpedicle screw) is intended for:
- a. Patients having fractures of the thoracic and lumbar spine.
- b. Patients having deformity (i.e., idioscoliosis, neuromuscular scollosis, or kyphoscoliosis with associated paralysis or spasticity).
- c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis. degenerative spondylolisthesis, and acute pars fracture allowing spondylolisthesis).
- 2) The ZODIAC Polvaxial Pedicle Screw System, when used as a pedicle screw system in the thoraco-lumbo-sacral region of the spine is intended for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis).
- 3) In addition, the ZODIAC Polyaxial Pedicle Screw System, when used as a pedicle screw fixation system is intended for:
- a. Patients receiving only autogenous bone graft.
- b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint.
- 4) The ZODIAC Polyaxial Pedicle Screw System, when used as a laminar hook and bone screw system is intended for:
- a. Patients having fractures of thoracic and lumbar spine.
b. Patients having thoracolumbar deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). (Division Sign-Off)
> Division of General, Restorative, and Neurological Devices
{5}------------------------------------------------
- c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolistheses and acute pars fracture allowing spondylolisthesis).
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The Counter Use
# (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aarbera Buerem for mxm
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number_K061321
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.