ZODIAC SPINAL FIXATION SYSTEM

K051286 · Alphatec/Nexmed · MNH · Jun 17, 2005 · Orthopedic

Device Facts

Record IDK051286
Device NameZODIAC SPINAL FIXATION SYSTEM
ApplicantAlphatec/Nexmed
Product CodeMNH · Orthopedic
Decision DateJun 17, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

It is intended that this device, in any system configuration be removed after development of solid fusion mass. Hook component indications are limited to T7-L5. Sacral screw indications are limited to the sacrum only. 1) The ZODIAC™ Spinal Fixation System, when used as a hook and sacral screw fixation system (nonpedicle screw) is intended for: a. Patients having fractures of the thoracic and lumbar spine. b. Patients having deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolisthesis, and acute pars fracture allowing spondylolisthesis). 2) The ZODIAC™ Spinal Fixation System, when used as a pedicle screw system in the thoraco-lumbo-sacral region of the spine is intended for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). 3) In addition, the ZODIAC™ Spinal Fixation System, when used as a pedicle screw fixation system is intended for: a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint. 4) The ZODIAC™ Spinal Fixation System, when used as a laminar hook and bone screw system is intended for: a. Patients having fractures of thoracic and lumbar spine. b. Patients having thoracolumbar deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolistheses and acute pars fracture allowing spondylolisthesis).

Device Story

ZODIAC™ Spinal Fixation System is a rigid spinal construct composed of implantable components (hooks, screws, rods). Used in thoraco-lumbo-sacral spine to provide stabilization during fusion. Implanted by surgeons in clinical settings. System provides mechanical support to the spine; intended for removal after solid fusion mass development. Benefits include stabilization of spinal segments in patients with fractures, deformities, or degenerative conditions.

Clinical Evidence

Bench testing only.

Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V ELI, ASTM F 136), commercially pure titanium (CP Grade 4, ASTM G67), and stainless steel (ASTM F 138). System consists of hooks, screws, and rods for rigid spinal fixation.

Indications for Use

Indicated for patients with thoracic/lumbar spine fractures, spinal deformities (scoliosis, kyphoscoliosis), spondylolisthesis (isthmic, degenerative, acute pars fracture), spinal tumors, dislocations, or failed previous fusion. Pedicle screw use requires objective evidence of neurologic impairment for degenerative spondylolisthesis. Specific indications for severe grade 3/4 spondylolisthesis at L5-S1. Hook use limited to T7-L5; sacral screws limited to sacrum.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K051286 r"1/2 # ZODIAC™ Spinal Fixation System 510(k) SUMMARY May 2005 | Company: | Alphatec Manufacturing, Inc.<br>6110 Corte Del Cedro<br>Carlsbad, CA 92009 USA<br>Telephone: (760) 431-9286<br>Fax: (760) 431-9132 | |-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Ellen Yarnall, Director of RA/QA | | Trade/Proprietary Name: | ZODIAC™ Spinal Fixation System | | Common Name: | Pedicle Screw Spinal System | | Classification Name: | Spinal Interlaminal Fixation Orthosis (888.3050)(KWP)<br>Spinal Pedicle Fixation (888.3070)(MNI)<br>Orthosis, Spondylolythisis Spinal Fixation (888.3070)(MNH) | ### Product Description: The ZODIAC™ Spinal Fixation System is comprised of various types and sizes of implantable components that are assembled to create a rigid spinal construct. System components are offered in titanium alloy, Ti-6Al-4V ELI (ASTM F 136). System rods are also available in commercially pure titanium, CP Grade 4 (ASTM G67). This submission seeks clearance for stainless steel (ASTM F 138) system components. #### Indications for Use: It is intended that this device, in any system configuration be removed after development of solid fusion mass. Hook component indications are limited to T7-L5. Sacral screw indications are limited to the sacrum only. - 1) The ZODIAC™ Spinal Fixation System, when used as a hook and sacral screw fixation system (nonpedicle screw) is intended for: - a. Patients having fractures of the thoracic and lumbar spine. - b. Patients having deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). - c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolisthesis, and acute pars fracture allowing spondylolisthesis). - 2) The ZODIAC™ Spinal Fixation System, when used as a pedicle screw system in the thoraco-lumbo-sacral region of the spine is intended for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). - 3) In addition, the ZODIAC™ Spinal Fixation System, when used as a pedicle screw fixation system is intended for: {1}------------------------------------------------ K051284 p²/₂ # ZODIAC™ Spinal Fixation System 510(k) SUMMARY May 2005 - a. Patients receiving only autogenous bone graft. - b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint. - 4) The ZODIAC™ Spinal Fixation System, when used as a laminar hook and bone screw system is intended for: - a. Patients having fractures of thoracic and lumbar spine. - Patients having thoracolumbar deformity (i.e., idioscoliosis, neuromuscular scoliosis, ﻗ or kyphoscoliosis with associated paralysis or spasticity). - c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolistheses and acute pars fracture allowing spondylolisthesis). #### Substantial Equivalence: Substantial equivalence for the ZODIAC™ Spinal Fixation System manufactured from stainless steel material is the same as the ZODIAC Spinal Fixation System previously cleared under K033090 and K042673. #### Performance Data Performance data were submitted to characterize the ZODIAC™ Spinal Fixation System manufactured from stainless steel. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with three wavy lines that form a stylized human figure. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. JUN 17 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Ellen A. Yarnall Director of Regulatory Affairs Alphatec Manufacturing Incorporated 6110 Corte Del Cedro Carlsbad, California 92009 Re: K051286 Trade/Device Name: ZODIAC™ Spinal Fixation System Regulation Number: 21 CFR 888.3050, 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI, KWP Dated: May 17, 2005 Received: May 18, 2005 Dear Ms. Yarnall: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Ms. Ellen A. Yarnall This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and w yourse ough finding of substantial equivalence of your device to a legally premarket notification "The results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you dosire spective at (240) 276-0120 . Also, please note the regulation entitled, Colliation of Complance market notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general information on Jour responser Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, s. Hipt Rhodes Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE | 510(k) Number (if known): | K051286 | |---------------------------|---------| |---------------------------|---------| ZODIAC™ Spinal Fixation System Device Name: ## Indications for Use: It is intended that this device, in any system configuration be removed after development of solid fusion mass. Hook component indications are limited to T7-L5. Sacral screw indications are limited to the sacrum only. - 1) The ZODIAC™ Spinal Fixation System when used as a hook and sacral screw fixation system (nonpedicle screw) is intended for: - a. Patients having fractures of the thoracic and lumbar spine. - b. Patients having deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). - c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolisthesis, and acute pars fracture allowing spondylolisthesis). - 2) The ZODIAC™ Spinal Fixation System, when used as a pedicle screw system in the thoraco-lumbo-sacral region of the spine is intended for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). - 3) In addition, the ZODIAC™ Spinal Fixation System, when used as a pedicle screw fixation system is intended for: - a. Patients receiving only autogenous bone graft. - b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint. - 4) The ZODIAC™ Spinal Fixation System, when used as a laminar hook and bone screw system is intended for: - a. Patients having fractures of thoracic and lumbar spine. - b. Patients having thoracolumbar deformity (i.e., idioscoliosis, neuromuscular scoliosis, or kyphoscoliosis with associated paralysis or spasticity). 1/2 {5}------------------------------------------------ - c. Patients having spondylolisthesis (i.e., isthmic spondylolisthesis, degenerative spondylolistheses and acute pars fracture allowing spondylolisthesis). Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR . Over-The Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW. THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Hypt Rurdis ign-(Division Division of General. Restorative, and Neurological Devices 510(k) Number K051286 2/2 000016
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...