The EBI® PI48 Spinal Fixation System is a single use spinal fixation device for pedicle screw fixation and a non-pedicle hook and sacral/iliac screw fixation system of the non-cervical spine. When used as a pedicle screw fixation system, in the non-cervical spine of skeletally mature patients, the System is intended to provide immobilization of spinal segments, as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocations, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system in skeletally mature patients, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5 - S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass. When used as a posterior hook and sacral/iliac screw fixation system, the levels of attachment are the lumbar and thoracic spine, and screw fixation is limited to the sacrum and ilium. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), pseudarthrosis, scoliosis, spondylolisthesis, fracture, previous failed fusion, or tumor resection. When used as an anterior fixation system, the levels of attachment are the anterolateral bodies of the lumbar and thoracic spine. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), pseudarthrosis, scoliosis, spondylolisthesis, fracture, previous failed fusion, or tumor resection.
Device Story
Single-use spinal fixation system; utilizes rods, screws, couplers, and hooks to immobilize and stabilize spinal segments. Used in thoracic, lumbar, and sacral spine; applied by surgeons in clinical settings. Provides mechanical support as adjunct to fusion; facilitates stabilization for various spinal pathologies. Benefits patient by correcting deformities and stabilizing unstable segments.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Technological Characteristics
Constructed from titanium alloy (Ti-6Al-4V ELI) per ASTM F136 and unalloyed titanium per ASTM F67-00. Mechanical fixation system consisting of rods, screws, couplers, and hooks.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion in the non-cervical spine. Treats degenerative disc disease, degenerative spondylolisthesis with neurologic impairment, fracture, dislocations, scoliosis, kyphosis, spinal tumor, pseudarthrosis, and stenosis. Specific criteria for severe L5-S1 spondylolisthesis (Grades 3-4) using autogenous bone graft.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
EBI® Omega 21™ Spinal Fixation System
Webb-Morley Spine System
Related Devices
K033312 — EBI TOP LOADING MAS SPINAL FIXATION SYSTEM · Ebi, L.P. · Jan 13, 2004
K201568 — Calvary Spine Pedicle Screw System · Calvary Spine, LLC · Nov 12, 2020
K992784 — NEXT GENERATION RODDING SYSTEM · United States Surgical, A Division of Tyco Healthc · Oct 27, 1999
K980184 — CD SPINAL SYSTEM · Sofamor Danek USA,Inc. · Dec 18, 1998
K152248 — CD HORIZON Spinal System · Medtronic Sofamor Danek USA, Inc. · Nov 5, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 2 7 2003
03/423 3
page 1 of 2
510(k) Summary
This 510(k) Summary for the EB1® P148 Spinal Fixation System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act.
- Contact Person: Frederic Testa, RAC 1. Submitter: EBI, L.P. Telephone: (973) 299-9300, ext.2208 100 Interpace Parkway Parsippany, NJ 07054
Date prepared: May 5, 2003
| 2. Proprietary Name: | EBI® P148 Spinal Fixation System |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Spinal Fixation Device |
| Classification Names: | Spondylolisthesis Spinal Fixation Device System<br>Spinal Intervertebral Body Fixation Orthosis<br>Spinal Interlaminal Fixation Orthosis |
## 3. Predicate or legally marketed devices that are substantially equivalent:
- EBI® Omega 21™ Spinal Fixation System .
- Webb-Morley Spine System .
- 4. Description of the device: The EBI® P148 Spinal Fixation System is a single use spinal fixation device that utilizes rods, screws, couplers and hooks.
- 5. Intended Use: The EBI® P148 Spinal Fixation System is a single use spinal fixation device for pedicle screw fixation and a non-pedicle hook and sacral/iliac screw fixation system of the non-cervical spine.
{1}------------------------------------------------
031423 3
page 20f
When used as a pedicle screw fixation system, in the non-cervical spine of skeletally mature patients, the System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies) degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocations; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system in skeletally mature patients, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5 - S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.
When used as a posterior hook and sacral/iliac screw fixation system, the levels of attachment are the lumbar and thoracic spine, and screw fixation is limited to the sacrum and ilium. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); pseudarthrosis; stenosis; scoliosis; spondylolisthesis: fracture: previous failed fusion: or tumor resection.
When used as an anterior fixation system, the levels of attachment are the anterolateral vertebral bodies of the lumbar and thoracic spine. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); pseudarthrosis; stenosis; scoliosis; spondylolisthesis; fracture; previous failed fusion: or tumor resection.
{2}------------------------------------------------
K0 31423
page 3 of 3
- 6. Materials: The components of the System are manufactured from titanium alloy (Ti-6A1-4V ELI) per ASTM F136 and unalloyed titanium per ASTM F 67-00.
- 7. Comparison of the technological characteristics of the device to predicate
devices: There are no significant differences between the proposed EBI® Omega21™ Spinal Fixation System and the currently marketed EBI Omega21 Spinal Fixation System. It is substantially equivalent* to the predicate devices in regards to intended use, materials, and function.
<sup>*</sup> Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully market approval or reclassification and is not interpreted as an admission or any other type of evidence in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)]
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, possibly representing water or movement.
Public Health Service
JUN 2 7 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Jon Caparotta, RAC Manager, Regulatory Affairs EBI, L.P. 100 Interpace Parkway Parsippany, NJ 07054
Re: K031423
Trade/Device Name: EBI® P148 Spinal Fixation System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: MNH, KWP, KWQ, MNI Dated: May 5 2003 Received: May 6, 2003
Dear Mr. Caparotta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Jon Caparotta, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
for Mark N. Millheam
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Statement of Indications for Use
510(k) Number (if known): k031423
Page I of
Device Name: EBI® P148 Spinal Fixation System
Indications For Use:
The EBI® PI48 Spinal Fixation System is a single use spinal fixation device for pedicle screw fixation and a non-pedicle hook and sacral/iliac screw fixation system of the non-cervical spine.
When used as a pedicle screw fixation system, in the non-cervical spine of skeletally mature patients, the System is intended to provide immobilization of spinal segments, as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocations, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system in skeletally mature patients, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5 - S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.
When used as a posterior hook and sacral/iliac screw fixation system, the levels of attachment are the lumbar and thoracic spine, and screw fixation is limited to the sacrum and ilium. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), pseudarthrosis, scoliosis, spondylolisthesis, fracture, previous failed fusion, or tumor resection.
When used as an anterior fixation system, the levels of attachment are the anterolateral bodies of the lumbar and thoracic spine. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), pseudarthrosis, scoliosis, spondylolisthesis, fracture, previous failed fusion, or tumor resection.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.