← Product Code [MJW](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MJW) · K033046

# SUBTALAR PEG IMPLANT (K033046)

_Nexa Orthopedics, Inc. · MJW · Dec 23, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MJW/K033046

## Device Facts

- **Applicant:** Nexa Orthopedics, Inc.
- **Product Code:** [MJW](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MJW.md)
- **Decision Date:** Dec 23, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Futura Biomedical Subtalar Peg Implant is intended to treat hyperpronated foot and stabilize the subtalar joint. It is intended to block forward, downward and medial displacement of the talus, thus allowing normal subtalar joint motion but blocking excessive pronation and the resulting sequela. Examples include: Flat foot treatment in children and adolescents Congenital flat foot Unsuccessful long term orthopaedic treatment (shoes, insoles ... ) Tarsal coalitions Painful flat foot Supple deformity in posterior tibial tendon dysfunction Paralytic flat foot Subtalar instability This device is intended to be fixed with bonc cement.

## Device Story

Subtalar Peg Implant acts as a spacer for subtalar joint; implanted into calcaneus to maintain joint space; limits excessive pronation while allowing normal joint motion; blocks forward, downward, and medial talar displacement. Device is one-piece, available in 13 sizes and two configurations. Used by orthopedic surgeons in clinical settings. Implanted via surgical procedure; fixed with bone cement. Provides mechanical stabilization for patients with flat foot or subtalar instability; benefits include pain reduction and improved joint function.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on design characteristics, intended use, and material comparisons to the predicate device.

## Technological Characteristics

One-piece implant; materials: UHMWPE or Cobalt Chromium (CoCr) with titanium plasma spray (CpTi) on stem. Dimensions: 13 sizes, two configurations. Fixation: bone cement. Non-constrained design. Single-use.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Smith Subtalar Peg ([K792670](/device/K792670.md))
- Wright Medical STA-Peg implant

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 14, 2016

Ms. Louise M. Focht Futura Biomedical 990 Park Center Drive, Suite H Vista. California 92081

Re: K033046

Trade/Device Name: Subtalar Peg Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MJW Dated: September 29, 2003 Received: September 29, 2003

Dear Ms. Focht:

This letter corrects our substantially equivalent letter of December 23, 2003.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

Page 2 - Ms. Louise M. Focht

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 8091 ), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Mejkersons -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Page ***_ of _***

1033046 510 (k) Number (If Known):_ Device Name: Subtalar Peg Implant

### Indications for Use:

The Futura Biomedical Subtalar Peg Implant is intended to treat hyperpronated foot and stabilize the subtalar joint. It is intended to block forward, downward and medial displacement of the talus, thus allowing normal subtalar joint motion but blocking excessive pronation and the resulting sequela. Examples include:

Flat foot treatment in children and adolescents Congenital flat foot Unsuccessful long term orthopaedic treatment (shoes, insoles ... ) Tarsal coalitions Painful flat foot Supple deformity in posterior tibial tendon dysfunction Paralytic flat foot Subtalar instability

This device is intended to be fixed with bonc cement.

Prescription Use (Part 21 CFR 801 Subpant D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

2

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

.

...ion Siera Civi...

. :: "::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::

{3}------------------------------------------------

DEC 2 3 2003 -

K033046

PAGE 1 of 3

## 510 (k) Summary

Summary of 510 (k) safety and effectivencss information upon which the substantial cquivalence determination is based:

December 22, 2003 Prepared: Futura Biomedical Applicant: 990 Park Center Drive Vista, CA 92081

760-599-1670 Telephone: Fax: 760-599-1675 Louise M. Focht Contact:

Device Name: Device Trade Name: Device Classification: Reviewing Panel: Regulation Number Product Code: Predicate Device: Registration Number: Owner Operator Number: Subtalar Peg Implant Subtalar Peg Implant Class II Orthopedic

87 MJW K792670- Smith Subtalar Peg 2030833 9028319

#### Device Description:

The Futura Biomedical Implant is a one-picce device made of UHMWPE or Cobalt Chromium with titanium plasma spray on the stem, intended to be implanted into the calcancus of the foot. The implant is designed in 13 sizes and two configurations. The implant which is used in the treatment of excessive motion of the talus relative to the calcaneus acts as a spacer for the joint, maintaining the joint space, allowing for range of motion, but limiting excessive pronation.

#### Indications for Use:

The Futura Biomedical Subtalar Peg Implant is intended to treat hyperpronated foot and stabalize the subtalar joint. It is intended to block forward, downward and medial displacement of the talus, thus allowing normal subtalar joint motion but blocking excessive pronation and the resulting sequela. Examples include:

Flat foot treatment in children and adolescents Congenital flat foot Unsuccessful long term orthopaedic treatment (shoes, insoles ... ) Tarsal coalitions

{4}------------------------------------------------

K03 3x46
Page 2 of 3

Painful flat foot Supple deformity in posterior tibial tendon dysfunction Paralytic flat foot Subtalar instability

This device is intended to be fixed with bone cement.

# Comparison to Predicate Device:

The legally marketed predicate device to which this device is substantially equivalent is the Wright Medical STA-Peg implant.

Regulatory Class: II 87 MJW Product Code:

Table 1. Comparison of Futura Biomedical and Wright Medical STA-Peg Implant

| Item                | Futura Product                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Wright Medical Product                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name        | Subtalar Peg Implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | STA-Peg Implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Usc                 | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Fixation            | Stem in bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Stem in bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Constraint          | Non constrained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | non constrained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Material            | UHMWPE or CoCr/CpTi                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | UHMWPE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Sizes               | 5 sizes, two configurations (13 sizes)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 5 sizes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for use | The Futura Biomedical Subtalar Peg<br>Implant is intended to treat<br>hyperpronated foot and stabilize the<br>subtalar joint. It is intended to block<br>forward, downward and medial<br>displacement of the talus, thus<br>allowing normal subtalar joint<br>motion but blocking excessive<br>pronation and the resulting sequela.<br>Examples include:<br>Flat foot treatment in<br>children and adolescents<br>Congenital flat foot<br>Unsuccessful long term<br>orthopaedic treatment<br>(shoes, insolcs...)<br>Tarsal coalitions<br>Painful flat foot<br>Supple deformity in<br>posterior tibial tendon<br>dysfunction<br>Paralytic flat foot<br>Subtalar instability<br>This device is intended to be fixed<br>with bone cement | The STA-Peg implant is intended to<br>treat hyperpronated foot and stabilize<br>the subtalar joint. It is intended to<br>block forward, downward and medial<br>displacement of the talus, thus<br>allowing normal subtalar joint motion<br>but blocking excessive pronation and<br>the resulting sequela. Examples<br>include:<br>Flat foot treatment in<br>children and adolescents<br>Congenital flat foot<br>Unsuccessful long term<br>orthopaedic treatment (shoes,<br>insoles...)<br>Tarsal coalitions<br>Painful flat foot<br>Supple deformity in posterior<br>tibial tendon dysfunction<br>Paralytic flat foot<br>Subtalar instability |

{5}------------------------------------------------

K072046

PAGE 3 of 3

3

Similarities of the Futura Biomedical Subtalar Arthrorisis Implant and the Wright Medical STA-Peg Implant include:

Both devices are: intended for single use only; intended for surgical implantation longer Bour devices are: michood for single aco of the calcaneus of the foot, allowing normal than 50 days, both dovies are prinst excessive pronaton and resulting sequela; both Sublaial John model of industry standard materials are introduced in either uevices are made of made of marably sized; both devices have the same indications for use.

# Summary:

The device and the predicate device have similar design characteristics and intended use. The new device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MJW/K033046](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MJW/K033046)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
