The Actis DuoFix Hip Prosthesis - Collarless is a forged titanium (Ti-6Al-4V) femoral stem for hip arthroplasty. It features a sintered commercially pure titanium bead porous coating (Porocoat) and a plasma-sprayed hydroxyapatite (HA) coating to facilitate cementless fixation. The device is a line extension of the Actis DuoFix system, modified to be collarless. It is used by orthopedic surgeons in clinical settings for total or hemi-hip arthroplasty. The stem is compatible with unipolar, bipolar, and modular metal/ceramic femoral heads. By providing a collarless design, it offers surgeons an alternative anatomical fit for patients requiring hip joint reconstruction. The device is implanted during surgery to restore joint function and alleviate pain associated with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical tests were conducted. Substantial equivalence is supported by non-clinical bench testing, including distal fatigue testing (ISO 7206-4), neck fatigue testing (ISO 7206-6), biocompatibility evaluation (ISO 10993 series), MRI safety evaluation (ASTM F2503, F2213, F2119, F2182), and bacterial endotoxin testing (ANSI/AAMI ST 72).
Technological Characteristics
Forged titanium alloy (Ti-6Al-4V) femoral stem; Porocoat (sintered commercially pure titanium beads) porous coating; plasma-sprayed hydroxyapatite (HA) coating. Cementless fixation. Sterilized via gamma irradiation (SAL 10^-6). Packaged in double PETG blister with Tyvek lid. 10-year shelf life.
Indications for Use
Indicated for total hip replacement in patients with severely painful/disabled joints due to osteoarthritis, rheumatoid arthritis, or congenital hip dysplasia; avascular necrosis of the femoral head; acute traumatic fracture of the femoral head/neck; or ankylosis. Indicated for hemi-hip arthroplasty in patients with acute fracture of the femoral head/neck or fracture dislocation that cannot be reduced/treated with internal fixation; avascular necrosis; non-union of femoral neck fractures; high subcapital/femoral neck fractures in the elderly; degenerative arthritis of the femoral head; or pathology of the femoral head/neck/proximal femur. Cementless use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 20, 2021
DePuy Ireland UC % Erin Combs Senior Regulatory Affairs Specialist Depuy Orthopaedics, Inc 700 Orthopaedic Drive Warsaw, Indiana 46582
Re: K210581
Trade/Device Name: Actis DuoFix Hip Prosthesis- Collarless Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: MEH, LPH, KWL, KWY Dated: April 30, 2021 Received: May 3, 2021
Dear Erin Combs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
William Jung, Ph.D. Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K210581
Device Name Actis DuoFix Hip Prosthesis - Collarless
#### Indications for Use (Describe)
#### INDICATIONS FOR USE
Total hip replacement or hip arthroplasty is indicated in the following conditions:
1. A severely painful and/or disabled joint from osteoarthritis, rheumatoid arthritis, or congenital hip dysplasia.
- 2. Avascular necrosis of the femoral head.
- 3. Acute traumatic fracture of the femoral head or neck.
- 4. Certain cases of ankylosis.
Hemi-hip arthroplasty is indicated in the following conditions:
- 1. Acute fracture of the femoral head or neck that cannot be appropriately reduced and treated fixation.
- 2. Fracture dislocation of the hip that cannot be appropriately reduced and treated with internal fixation.
- 3. Avascular necrosis of the femoral head.
- 4. Non-union of femoral neck fractures.
- 5. Certain high subcapital and femoral neck fractures in the elderly.
- 6. Degenerative arthritis involving only the femoral head in which the acetabulum does not require replacement.
7. Pathology involving only the femoral head/neck and/or proximal femur that can be adequately treated by hemi-hip arthroplasty.
The ACTIS™ DUOFIX™ Hip Prosthesis - Collarless is indicated for cementless use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(K) SUMMARY
(As required by 21 CFR 807.92 and 21 CFR 807.93)
| Submitter Information | |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Ireland UC |
| Address | Loughbeg, Ringaskiddy<br>Co. Cork, IRELAND |
| Establishment Registration<br>Number | 3015516266 |
| Name of contact person | Elaine Pears |
| e-mail address | epears@its.jnj.com and DePuySynthesJointsRegulatoryAffairs@its.jnj.com |
| Alternative contact person | Clare Hill |
| e-mail address | chill7@its.jnj.com |
| Work mobile | +44 7795 389956 (UK time zone) |
| Date prepared | 16th February 2021 |
| Name of device | |
| Trade or proprietary name | DePuy Actis DuoFix Hip Prosthesis – Collarless |
| Common or usual name | Uncemented Hip Prosthesis |
| Classification name | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented<br>prosthesis;<br>Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous<br>uncemented prosthesis;<br>Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis;<br>Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented<br>prosthesis. |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | 21 CFR 888.3358; 888.3353; 888.3360; 888.3390 |
| Product Code(s) | MEH, LPH, KWL, KWY |
| Legally marketed device(s) to<br>which equivalence is claimed | Primary predicate - DePuy Actis DuoFix Hip Prosthesis K160907, cleared July<br>19, 2016<br>Secondary predicate - DePuy Summit DuoFix Hip Prosthesis K170339,<br>cleared April 26, 2018<br>Reference predicate for hydroxyapatite (HA) coating process step – DePuy<br>Actis DuoFix Hip Prosthesis K202472, cleared October 21, 2020 |
| Reason for 510(k) submission | The purpose of this 510K submission is to obtain market clearance for DePuy<br>Actis DuoFix Hip Prosthesis - Collarless. |
| Device description | The DePuy Actis DuoFix Hip Prosthesis – Collarless is a forged titanium (Ti-6Al-4V) femoral stem designed to be used as one component of a system of prostheses in hip arthroplasty. The stems are compatible with both unipolar and bipolar femoral heads intended for hemi-hip arthroplasty and with modular metal and ceramic femoral heads intended for total hip arthroplasty<br>The Actis DuoFix Hip Prosthesis – Collarless implants have sintered commercially pure titanium bead porous coating (Porocoat®) and a thin layer of plasma-sprayed hydroxyapatite (HA) coating. The stem consists of a wide range of stem neck designs and sizes allowing an accurate anatomical match for each patient.<br>The Actis DuoFix Hip Prosthesis – Collarless is designed as a line extension to the primary predicate Actis DuoFix Hip Prosthesis (K160907) to meet the need of surgeons whose preference is to use a femoral stem without a collar for hip arthroplasty. |
| Intended use of the device | Total hip arthroplasty and hemi-hip arthroplasty |
| Indications for use | Total hip replacement is indicated in the following conditions:<br>1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.<br>2. Avascular necrosis of the femoral head.<br>3. Acute traumatic fracture of the femoral head or neck.<br>4. Certain cases of ankylosis.<br>Hemi-hip arthroplasty is indicated in the following conditions:<br>1. Acute fracture of the femoral head or neck that cannot be appropriately reduced and treated with internal fixation.<br>2. Fracture dislocation of the hip that cannot be appropriately reduced and treated with internal fixation.<br>3. Avascular necrosis of the femoral head.<br>4. Non-union of femoral neck fractures.<br>5. Certain high subcapital and femoral neck fractures in the elderly.<br>6. Degenerative arthritis involving only the femoral head in which the acetabulum does not require replacement.<br>7. Pathology involving only the femoral head/neck and/or proximal femur that can be adequately treated by hemi-hip arthroplasty. |
{6}------------------------------------------------
{7}------------------------------------------------
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICES | | | | |
|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Characteristics | Subject Device:<br>DePuy Actis DuoFix<br>Hip Prosthesis -<br>Collarless | Primary Predicate Device:<br>DePuy Actis DuoFix Hip Prosthesis<br>K160907 | Secondary Predicate Device:<br>DePuy Summit DuoFix Hip Prosthesis K170339 | |
| Intended Use<br>and Indications<br>for Use | Same as primary<br>predicate | Intended for use in total hip arthroplasty and hemi-hip<br>arthroplasty.<br>Total hip replacement is indicated in the following<br>conditions:<br>1. A severely painful and/or disabled joint from<br>osteoarthritis, traumatic arthritis, rheumatoid arthritis,<br>or congenital hip dysplasia.<br>2. Avascular necrosis of the femoral head.<br>3. Acute traumatic fracture of the femoral head or neck.<br>4. Certain cases of ankylosis.<br>Hemi-hip arthroplasty is indicated in the following<br>conditions:<br>1. Acute fracture of the femoral head or neck that cannot<br>be appropriately reduced and treated with internal<br>fixation.<br>2. Fracture dislocation of the hip that cannot be<br>appropriately reduced and treated with internal<br>fixation.<br>3. Avascular necrosis of the femoral head.<br>4. Non-union of femoral neck fractures.<br>5. Certain high subcapital and femoral neck fractures in<br>the elderly.<br>6. Degenerative arthritis involving only the femoral head<br>in which<br>the acetabulum does not require<br>replacement.<br>7. Pathology involving only the femoral head/neck<br>and/or proximal femur that can be adequately<br>treated by hemi-hip arthroplasty. | Intended for use in total hip arthroplasty and hemi-hip<br>arthroplasty.<br>Total hip replacement is indicated in the following<br>conditions:<br>1. A severely painful and/or disabled joint from<br>osteoarthritis, traumatic arthritis, rheumatoid arthritis, or<br>congenital hip dysplasia.<br>2. Avascular necrosis of the femoral head.<br>3. Acute traumatic fracture of the femoral head or neck.<br>4. Failed previous hip<br>surgery including joint<br>reconstruction,<br>fixation,<br>arthrodesis.<br>internal<br>hemiarthroplasty, surface replacement arthroplasty, or<br>total hip replacement.<br>5. Certain cases of ankylosis.<br>Hemi-hip replacement is indicated in the following<br>conditions:<br>1. Acute fracture of the femoral head or neck that cannot<br>be appropriately reduced and treated with internal<br>fixation.<br>2. Fracture dislocation of the hip that cannot be<br>appropriately reduced and treated with internal fixation.<br>3. Avascular necrosis of the femoral head.<br>4. Non-union of femoral neck fractures.<br>5. Certain high subcapital and femoral neck fractures in<br>the elderly.<br>6. Degenerative arthritis involving only the femoral head<br>in which the acetabulum does not require replacement.<br>7. Pathology involving only the femoral head/neck and/or<br>proximal femur that can be adequately treated by hemi-<br>hip arthroplasty. | |
| | | | | |
| Material | Same as primary<br>predicate | Femoral stem: Wrought titanium alloy<br>Porous coating: Commercially pure unalloyed<br>Titanium sintered bead porous coating (Porocoat®)<br>HA coating: Hydroxyapatite powder | Same as primary predicate | |
| Fixation | Same as primary<br>predicate | Uncemented | Same as primary predicate | |
| Stem Size | Same as primary<br>predicate | 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 | 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 | |
| Neck Offset | Same as primary<br>predicate | Standard, High | Same as primary predicate | |
| Collar | Same as secondary<br>predicate | Collared | Collarless | |
| Sterile Method | Same as primary<br>predicate | Gamma irradiation | Same as primary predicate | |
| SAL | Same as primary<br>predicate | 10⁻⁶ | Same as primary predicate | |
| Packaging | Same as primary<br>predicate | Double PETG blister with Tyvek peel lid | Same as primary predicate | |
| Shelf Life | Same as primary<br>predicate | 10 years | Same as primary predicate | |
{8}------------------------------------------------
{9}------------------------------------------------
The subject device [Acts DuoFix Hip Prosthess] is identical in design to the primary predicate device [Actis DuoFix Hip Prosthesis (K160907) with the single exception that, in common with the secondary predicate [Summit DuoFix Hip Prosthesis (K170339)], there is no collar on the neck of the device.
The subject device has the same intentialy and secondary predicate devices. The subject and both predicate devices are made from the same materials and have the same method of fixation. Sterility Assurance Level (SAL), packaging materials and shelf-life are the same across the subject and predicate devices.
Indications for use and range of stem sizes and neck offsets available are identical for the subject device and the primary predicate.
## SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
Distal fatigue testing (to ISO 7206-4: 2010), neck fairge testing (to ISO 7206-6:2013) and biocompatibility evaluation (to ISO 1099-1:2018, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-10:2017 and ISO 10993-18:2020) were carried out as part of design verification for the devices in order to demonstrate substantial equivalence of safety and efficacy with the predicate devices.
Magnetic Resonance Imaging safety evaluation testing was performed using ASTM F2503-13, and the tests evaluated the worst-case components and constructs for magnetically induced force (ASTM F2213-17), image artefact (ASTM 2119-07) and RF heating (ASTM F2182-11a). The testing concluded that there are no safety issues related to magnetic field interactions identified in the labelling.
The proposed devices also meet the requirement of bacterial endotoxin testing as specified in ANSI/AAMI ST 72:2019.
Hydroxyaatie characterization data, as recommended in the FDA Guidance 510(k) Information needed for Hydrosy, while hands, is provided.
SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No clinical tests were conducted to demonstrate substantial equivalence.
### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The subject DePuy Actis DuoFix Hip Prosthess devices are substantially equivalent to the predicate DePuy Actis DuoFix and DePuy Summit DuoFix hip prostheses (K160907 and K170339).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.