MSA HIP SYSTEM

K113107 · Global Manufacturing Technology · MEH · Dec 15, 2011 · Orthopedic

Device Facts

Record IDK113107
Device NameMSA HIP SYSTEM
ApplicantGlobal Manufacturing Technology
Product CodeMEH · Orthopedic
Decision DateDec 15, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MSA Hip System is indicated for cementless use and is intended to replace a hip joint. The device is intended for: - skeletally mature individuals undergoing primary surgery for total hip replacement - patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fracture of the femur. - patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis - patients suffering from disability due to previous fusion - patients with acute femoral neck fractures

Device Story

Modular hip prosthesis system; replaces hip joint. Components: short stem (Ti-6Al-4V), modular neck (CoCr alloy), and femoral head (metal or ceramic). Stem features proximal plasma spray coating; optional hydroxyapatite coating. Device implanted by orthopedic surgeons in clinical/OR setting. System provides structural replacement for damaged hip joints; restores joint function; alleviates pain. Ceramic femoral head addition represents design modification to existing system.

Clinical Evidence

Bench testing only. Burst strength testing performed on the ceramic femoral head to verify mechanical integrity.

Technological Characteristics

Modular hip prosthesis. Materials: Ti-6Al-4V (ASTM F136) stem, CoCr alloy (ASTM F1537) neck, ceramic or metal femoral head. Surface: Commercially pure titanium plasma spray (ASTM F1580) on proximal stem; optional hydroxyapatite coating. Non-software device.

Indications for Use

Indicated for skeletally mature individuals requiring primary total hip replacement due to severe hip pain/disability from rheumatoid/osteo/traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, nonunion of femoral fracture, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, previous fusion, or acute femoral neck fractures.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 1 5 2011 K113107 (pg 1/2) #### 5. 510(k) Summary | Contact: | Mr. Steve Banks<br>Director | |--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------| | Prepared By: | Musculoskeletal Clinical Regulatory Advisers, LLC<br>1331 H Street, NW, 12th Floor<br>Washington, DC 20005<br>Phone: (202) 552-5800<br>Fax: (202) 552-5798 | | Date Prepared: | October 19, 2011 | | Device Trade Name: | MSA Hip System | | Manufacturer: | Global Manufacturing Technology<br>8-10 Resolution Drive<br>Unanderra NSW 2526<br>AUSTRALIA<br>Phone: (011) 61 (2) 8887 0100 | | Classification: | 21 CFR 888.3353 | | | Hip joint metal/ceramic/polymer semi-constrained<br>cemented or nonporous uncemented prosthesis | | | Class II | | Product Code: | MEH, LZO | # Reason for Special 510(k) Submission: The purpose of this Special 510(k) is to add ceramic femoral heads to the MSA Hip System. There have been no changes to the intended use of the device or its fundamental scientific technology. # Indications For Use: The MSA Hip System is indicated for cementless use and is intended to replace a hip joint. The device is intended for: - · skeletally mature individuals undergoing primary surgery for total hip replacement - · patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fracture of the femur. {1}------------------------------------------------ - · patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis - · patients suffering from disability due to previous fusion - · patients with acute femoral neck fractures ### Device Description: The MSA Hip System is a modular hip system consisting of a short stem which mates with a modular neck. The hip stems are made of Ti-6AI-4V (ASTM F136) with a commercially pure titanium plasma spray coating (ASTM F1580) on the proximal end. The stems are available with and without a hydroxyapatite coating. The modular necks are made of CoCr alloy (ASTM F1537). The MSA Hip System can be mated with metal or ceramic femoral heads and acetabular components. #### Predicate Devices: The modified MSA Hip System is substantially equivalent to the predicate MSA Hip System (K102172) with respect to indications, design, and function. #### Substantial Equivalence: The company performed burst strength testing of the ceramic femoral head. The test results demonstrate that the pre-determined acceptance criteria identified in the Design Control Activities Summary were met. #### Conclusion The MSA Hip System when mated with a ceramic femoral head is substantially equivalent to previously cleared devices with respect to its indications for use, design, and function. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a human figure. # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Global Manufacturing Technology % Musculoskeletal Clinical Regulatory Advisers, LLC Ms. Hollace Rhodes Director, Orthopaedic Regulatory Affairs 1331 H St Northwest Washington, DC 20005 DEC 1 5 2011 Re: K113107 Trade/Device Name: MSA Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis porous uncemented prosthesis Regulatory Class: Class II Product Code: MEH, LZO Dated: November 16, 2011 Received: November 17, 2011 Dear Ms. Rhodes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ ### Page 2 - Ms. Hollace Rhodes device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, incerely yours, for Peter Vrooman Mark N. Melkerson Mark N. Melker Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use 4. 510(k) Number (if known): 长川3107 (pg ١/١) Device Name: MSA Hip System The MSA Hip System is indicated for cementless use and is intended to replace a hip joint. The device is intended for: - skeletally mature individuals undergoing primary surgery for total hip . replacement - patients suffering from severe hip pain and disability due to rheumatoid arthritis, ● osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fracture of the femur. - patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital . femoral epiphysis - patients suffering from disability due to previous fusion . - . patients with acute femoral neck fractures イ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for MxM (Division Sign-Off) Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113107
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...