PREVISION HIP SYSTEM WITH RECON RING

K102424 · Aesculap Implant Systems, LLC · MEH · Dec 9, 2010 · Orthopedic

Device Facts

Record IDK102424
Device NamePREVISION HIP SYSTEM WITH RECON RING
ApplicantAesculap Implant Systems, LLC
Product CodeMEH · Orthopedic
Decision DateDec 9, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Prevision Hip System is intended for cementless revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The Prevision Hip System is intended for revision only. The Recon Ring is intended to bridge the areas of acetabular bone loss in patients with acetabular bone deficiency. In addition, the Recon Ring is intended to provide support for an all polyethylene acetabular implant in a cemented application. Diagnostic indications for use of this device include acetabular dysplasia, osteoporosis, protrusion acetabuli, cystic acetabular roof, reconstruction in cases of defects after fracture, acetabular loosening, tumors or revision surgery. The general indications associated with the use of the Recon Ring in total hip arthroplasty include: Advanced joint destruction resulting from degenerative, posttraumatic or rheumatoid arthritis. Fracture or avascular necrosis of the femoral head, Failed previous surgery, e.g. osteosynthesis, joint reconstruction, arthrodesis, hemi-arthroplasty and total hip replacement.

Device Story

Modular hip prosthesis system consisting of proximal body, distal stem, and tension nut; components assembled by surgeon in OR or in situ to allow independent sizing. Designed for cementless revision hip arthroplasty; Recon Ring component bridges acetabular bone loss and supports cemented all-polyethylene acetabular implants. Manufactured from titanium alloy; proximal body features plasma-sprayed (Plasmapore) and calcium phosphate (u-CaP) coatings. Used by orthopedic surgeons in clinical settings to restore hip function in patients with significant bone deficiency or failed prior joint reconstructions.

Clinical Evidence

Bench testing only. Performance evaluated in accordance with ISO 7206 and ASTM F 2068 standards. Results demonstrated similarity to legally marketed predicate devices.

Technological Characteristics

Modular hip system; titanium alloy construction; proximal body plasma-sprayed (Plasmapore) with secondary calcium phosphate (u-CaP) coating. Recon Ring manufactured from titanium. Designed for cementless use. Testing performed per ISO 7206 and ASTM F 2068.

Indications for Use

Indicated for patients requiring cementless revision hip arthroplasty due to poor bone stock or deficiencies of the femoral head, neck, or proximal femur; also indicated for patients with acetabular bone deficiency requiring bridge support or cemented all-polyethylene acetabular implant support. Includes patients with acetabular dysplasia, osteoporosis, protrusion acetabuli, cystic acetabular roof, post-fracture defects, loosening, tumors, or failed prior surgeries (e.g., arthrodesis, hemi-arthroplasty).

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102424 (pg 1/2) Page 1 of 2 #### B. 510(k) SUMMARY (as required by 21 CFR 807.92) DEC - 9 2010 Prevision Hip System with Recon Ring December 3, 2010 | COMPANY: | Aesculap Implant Systems, LLC<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 3005673311 | |----------|-------------------------------------------------------------------------------------------------------------------------------------| |----------|-------------------------------------------------------------------------------------------------------------------------------------| CONTACT: Kathy A. Racosky 610-984-9291 (phone) 610-791-6882 (fax) Kathy.racosky@aesculap.com TRADE NAME: Prevision Hip System COMMON NAME: Femoral Hip Prosthesis CLASSIFICATION NAME: Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calcium-Phosphate Prosthesis, Hip, Semi-constrained, Metal/Polymer, Cemented Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented or Uncemented 888.3353, 888.3350, 888.3360, 888.3353, 888.3390, REGULATION NUMBER: PRODUCT CODE: MEH, JDI, LWJ, LZO, KWY #### SUBSTANTIAL EQUIVALENCE Aesculap Implant Systems, LLC believes that the Prevision Hip System with Recon Ring is substantially equivalent to: - Zimmer, Inc. ZMR Hip System (K992667 & K994286) ● - Intermedics Orthopedics Inc., Burch/Schneider Reinforcement Cage (K960678) . - Aesculap Implant Systems Metha Hip System (K071916) . - Aesculap Implant Systems Bicontact (K040191) . # DEVICE DESCRIPTION The Prevision Hip System is a modular system which consists of a proximal body, distal stem and a tension nut. The three individual components utilize a modular junction. The individual components may be assembled by the surgeon in the operating room or in situ to allow independent sizing of the proximal body and distal stem. All proximal components are able to mate with all distal components providing a comprehensive range of combinations offering optimal flexibility. It is designed for use with currently available Aesculap Implant Systems femoral heads, acetabular components and Bipolar cups. {1}------------------------------------------------ ### Page 2 of 2 The Prevision Hip System is manufactured from Titanium Alloy and intended for cementless use. The proximal body is plasma sprayed (Plasmapore®) with a secondary coating of Calcium Phosphate (u-CaP®). The Recon Ring will be offered in three sizes with left and right configurations. The Recon Ring is manufactured from Titanium. It is designed to be used with the PE acetabular cup cleared in BiContact (K040191). # PURPOSE FOR PREMARKET NOTIFICATION The purpose for this submission is to gain marketing clearance for the Aesculao Implant Systems Prevision Hip System with Recon Ring. ### INDICATIONS FOR USE The Prevision Hip System is intended for cementless revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The Prevision Hip System is intended for revision only. The Recon Ring is intended to bridge the areas of acetabular bone loss in patients with acetabular bone deficiency. In addition, the Recon Ring is intended to provide support for an all polyethylene acetabular implant in a cemented application. Diagnostic indications for use of this device include acetabular dysplasia, osteoporosis, protrusion acetabuli, cvstic acetabular roof, reconstruction in cases of defects after fracture, acetabular loosening, tumors or revision surgery. The general indications associated with the use of the Recon Ring in total hip arthroplasty include: - Advanced joint destruction resulting from degenerative, posttraumatic or rheumatoid . arthritis. - . Fracture or avascular necrosis of the femoral head, - Failed previous surgery, e.g. osteosynthesis, joint reconstruction, arthrodesis, hemi-● arthroplasty and total hip replacement. # TECHNOLIGICAL CHARACTERISTICS(compared to Predicate(s)} The components of the Aesculap Implant Systems Prevision Hip System with Recon Ring are offered in a similar range of shapes and sizes as the predicate devices. The material used for the Aesculap Implant Systems device is the same as that used to manufacture the predicate devices. # PERFORMANCE DATA Testing of the Aesculap Implant Systems Prevision Hip System was performed in accordance with ISO 7206 and ASTM F 2068 and the results were found to be similar to other legally marketed predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Aesculap Implant Systems LLC % Ms. Kathy A. Racosky 3773 Corporate Parkway Center Valley, PA 18034 DEC - 9 2010 Re: K102424 : Trade/Device Name: Prevision Hip System with Recon Ring Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: MEH, JDI, LWJ, LZO, KWY Dated: December 3, 2010 Received: December 6, 2010 Dear Ms. Racosky: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 910(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regarly manatinent date of the Medical Device Amendments, or to commerce province to May 26, 1976, the enasting with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approval of a premarket approval app and Cosment Act (Act) that to not require approvinces of the general controls provisions of the Act. The You may, merelore, market the device, subject to the genirements for annual registration, listing of general controls provisions of the rict liveling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Prease note: CDICT does not over be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it es an be If your device is classified (Scc above) fine viar regulations affecting your device can be may be subject to additional controls. Existing major regulations addition, EDA m may be subject to additions: Existing mays. Existing mays found in the Code of I cucral Rogarations) This Levice in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s Issualitat equr device of a success at that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Ms. Kathy A. Racosky I uge 2 - Ms. Ratily in Not regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm1. Sincerely yours, Sincerely yours, {} Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT A. DEC - 9 2010 **510(k) Number:** K102424(pg 1/1) ### Device Name: Prevision Hip System with Recon Ring #### Indications for Use: The Prevision Hip System is intended for cementless revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The Prevision Hip System is intended for revision only. The Recon Ring is intended to bridge the areas of acetabular bone loss in patients with acetabular bone deficiency. In addition, the Recon Ring is intended to provide support for an all polyethylene acetabular implant in a cemented application. Diagnostic indications for use of this device include acetabular dysplasia, osteoporosis, protrusion acetabuli, cystic acetabular roof, reconstruction in cases of defects after fracture, acetabular loosening, tumors or revision surgery. The general indications associated with the use of the Recon Ring in total hip arthroplasty include: - Advanced joint destruction resulting from degenerative, posttraumatic or rheumatoid ● arthritis. - Fracture or avascular necrosis of the femoral head, � - Failed previous surgery, e.g. osteosynthesis, joint reconstruction, arthrodesis, hemi-. arthroplasty and total hip replacement. | Prescription Use | X | |-----------------------------|---| | and/or Over-the-Counter Use | | (per 21 CFR 801.109) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Apr M. Wilkerson (Division Sign-Off (Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K102424
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...