K091193 · Isotis Orthobiologics, Inc. · MBP · Aug 10, 2009 · Orthopedic
Device Facts
Record ID
K091193
Device Name
ACCELL EVO3C
Applicant
Isotis Orthobiologics, Inc.
Product Code
MBP · Orthopedic
Decision Date
Aug 10, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Accell Evo3c™ is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the posterolateral spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.
Device Story
Accell Evo3c™ is a bone void filler composed of human demineralized bone matrix (DBM) and cancellous bone chips. The DBM provides osteoinductive potential, while the cancellous chips serve as an osteoconductive scaffold. The device is used by surgeons to fill skeletal voids or gaps in the posterolateral spine, extremities, and pelvis. It is supplied as a particulate/dispersed material. The device is processed to ensure viral inactivation through donor screening, serologic testing, tissue cleaning, and terminal sterilization. It functions as a graft extender or void filler to support bone healing in non-load-bearing applications.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including quantitative in vitro assays for osteoinductive potential (validated against athymic mouse models) and viral inactivation validation studies using a panel of viruses.
Technological Characteristics
Composition: Human demineralized cortical bone matrix (DBM) and cancellous bone chips. Osteoconductive scaffold and osteoinductive potential. Processing: Antimicrobial, antiviral, and antiseptic solutions per AATB recommendations. Sterilization: Terminal sterilization. Form factor: Particulate/dispersed bone graft material.
Indications for Use
Indicated for patients requiring bone void filling or grafting in the posterolateral spine, extremities, and pelvis. Used for surgically created defects or traumatic bone injuries. Not for use in areas intrinsic to skeletal stability.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
### 5.0 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | IsoTis OrthoBiologics, Inc. | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| TRADE NAME: | ACCELL EVO3C™ | |
| COMMON NAME: | Bone Void Filler | |
| CLASSIFICATION<br>NAME: | Resorbable calcium salt bone void filler | AUG 10 2009 |
| DEVICE<br>CLASSIFICATION: | Class II | |
| PRODUCT CODE | MQV, MBP | |
| PREDICATE DEVICES: | Accell Family of products - K061880 (Accell DBM100,<br>Accell TBM®, Accell A2i and Accell Connexus®),<br>DYNAGRAFT® II - K040419, ORTHOBLAST® II | |
K050642 , DBX DBM Mix - K063676, Grafton DBM Orthoblend - K051195
#### 5.1 Substantially Equivalent To:
The Accell Evo3c™ is substantially equivalent in intended use, principal of operation and technological characteristics to the current Accell Family of products (Accell DBM100, Accell TBM®, Accell A2i and Accell Connexus®); 510(k) - K061880, DYNAGRAFT® II; 510(k) -K040419, ORTHOBLAST® II; 510(k) – K050642, DBX DBM Mix; 510(k) – K063676, and Grafton DBM Orthoblend 510(k) – K051195.
#### 5.2 Description of the Device Subject to Premarket Notification:
The Accell Evo3c™ is an extension to the Accell Family of products cleared under 510(k) K061880. The Accell Family of products, including Accell Evo3c™, contain human demineralized bone matrix (DBM) in particulate and dispersed forms. The Accell Evo3c™ is the same as Accell A2i, submitted under 510(k) K061880, with the exception of the inclusion of cancellous bone chips which are added as an osteoconductive scaffold. The Accell A2i is marketed under the brand name "Accell Evo3™," hence the "c" in Accell Evo3c™ denotes the inclusion of cancellous chips in the formulation.
# 5.3 Indication for Use:
Accell Evo3c™ is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the posterolateral spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.
{1}------------------------------------------------
#### 5.4 Technical Characteristics:
The Accell Evo3c™ and Accell Family of products (K061880) utilize ground and demineralized, human cortical bone (DBM) in the formulation of the product. The Accell Evo3c™ includes cancellous chips which act as an osteoconductive scaffold. The DBM component exhibits osteoinductive potential in validated animal and/or in vitro models. It is unknown how the osteoinductive potential, measured in these validated models, will correlate with clinical performance in human subjects.
#### 5.5 Performance Data:
All necessary verification testing has been performed for the Accell Evo3c™ product to assure substantial equivalence to the predicate device.
#### 5.6 Osteoinductive Potential
Each lot of DBM used to manufacture the Accell Evo3c™ and the Accell Family of products is tested for osteoinductive potential using a quantitative in vitro assay. The in vitro assay has been validated to correlate to an athymic mouse osteoinductive potential assay. It is unknown how osteoinductive potential measured via the in vitro or athymic mouse assays will correlate with human clinical performance.
#### 5.7 Viral Inactivation Validation
The methods for processing of the DBM contained in Accell Evo3c™ and the Accell Family of products were evaluated for their viral inactivation potential. A selected panel of viruses representing various virus types, sizes, shapes and genomes were evaluated. The viral inactivation testing demonstrated suitable inactivation potential of the processing methods for a wide range of potential viruses.
Viral Inactivation Validation: The methods for processing of the DBM contained in Accell Evo3c were evaluated for their viral inactivation potential. A selected panel of viruses representing various virus types, sizes, shapes and genomes were evaluated. The viral inactivation testing demonstrated suitable inactivation potential of the processing methods for a wide range of potential viruses. . The Evo3c product contains cancellous bone which has been added to the demineralized bone matrix. Since the cancellous bone chips are not demineralized, the degree of viral inactivation of this component is not fully known The cancellous bone has been processed in antimicrobial, antiviral, and antiseptic solutions which are recommended by the American Association of Tissue Banks (AATB) to reduce the risk of transmissible viral diseases from human tissue products. The risk of disease transmission with the cancellous bone component remains low due to multiple safeguards including donor screening, serologic testing, tissue cleaning process, and terminal sterilization of the finished device.
#### 5.8 Basis for Determination of Substantial Equivalence:
Upon reviewing the safety and efficacy information provided in this submission and comparing
{2}------------------------------------------------
ﻟﺪ
intended use, principle of operation and overall technological characteristics, the Accell Evo3c™, has been determined by IsoTis OrthoBiologics to be substantially equivalent to an existing legally marketed device.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Public Health Service
IsoTis OrthoBiologics, Inc. Division of Integra LifeSciences Corporation % Ms. Carroll Councilman Director Quality Assurance and Regulatory Affairs 2 Goodyear, Suite A Irvine, California 92618
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## AUG 1 0 2009
Re: K091193
Trade/Device Name: Accell Evo3cTM Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler devices Regulatory Class: II Product Code: MQV, MBP Dated: June 2, 2009 Received: June 3, 2009
Dear Ms. Councilman :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Carroll Councilman
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Сваловкин
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K091193
Device Name:
Accell Evo3c™
Indications for Use:
Accell Evo3c™ is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the posterolateral spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div> <span style="font-size: 2em;">✓</span> </div> (Per 21 CFR 801.109) | OR | Over-The-Counter Use |
|------------------|--------------------------------------------------------------------------|----|--------------------------|
| | | | (Optional Format 1-2-96) |
| | (Division Sign-Off) | | Page 1 of 1 |
| | Division of Surgical, Orthopedic, and Restorative Devices | | |
| 510(k) Number | K091193 | | |
CONFIDENTIALISOTIS ORTHOBIOLOGICS, INC. ACCELL EVO3C™
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.