← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K974049

# INNOVASIVE LINX BT 10MM LIGAMENT FASTENER (K974049)

_Innovasive Devices, Inc. · MBI · Feb 18, 1998 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K974049

## Device Facts

- **Applicant:** Innovasive Devices, Inc.
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Feb 18, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The 10mm Ligament Fasteners are intended for use in the fixation of ligament and tendon bone block grafts in cruciate ligament reconstruction surgeries of the knee.

## Device Story

The Innovasive 10mm Ligament Fastener is a bone fixation device used in knee cruciate ligament reconstruction. It consists of an outer threaded sleeve and a central pin. During surgery, the sleeve is inserted into bone; the central pin is then driven into the sleeve, causing the sleeve to expand and anchor into the bone. The pin features a side-plate with transverse pegs to secure bone-patella-bone grafts. A temporary suture is used to guide the pin into the sleeve and is subsequently discarded. The device is intended for use by surgeons in a clinical setting to provide mechanical fixation of soft tissue grafts.

## Clinical Evidence

Bench testing only. Mechanical testing compared the ultimate holding strength of the 10mm Ligament Fastener to the predicate device in a bone model, demonstrating equivalent holding strength.

## Technological Characteristics

Device consists of an outer threaded sleeve and a central expansion pin. Dimensions: 10mm outside diameter, 20mm length. Fixation is achieved via mechanical expansion of the sleeve within the bone. No software or electronic components.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- 8mm Innovasive Devices Ligament Fastener ([K970316](/device/K970316.md))
- Mitek Ligament Anchor

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the text "K974049" on the top line and "510(K) SUMMARY" on the second line. The text is black and the background is white. The text appears to be handwritten or in a font that mimics handwriting.

## 1. SUBMITTER:

Innovasive Devices, Inc. 734 Forest St. Marlborough, MA 01752 Telephone: 508-460-8229

Contact: Stephen M. Page, Manager of Regulatory Affairs Date Prepared: October 22, 1997

## 2. DEVICE:

ుండి 10:5

Innovasive 10mm Ligament Fastener Classification Name: Single/multiple component bone fixation appliances and accessories. Trade Name: innovasive Devices 10mm Ligament Fastener

### 3. PREDICATE DEVICE:

The predicate devices used to determine substantial equivalence for the Innovasive Devices 10mm Ligament Fastener were (1) the 8mm Innovasive Devices Ligament Fastener marketed by Innovasive Devices, Marlborough, MA, and (2) the Mitek Ligament Anchor marketed by Mitek Surgical, Norwood, MA.

#### 4. DEVICE DESCRIPTION:

The Innovasive 10mm Ligament Fastener utilizes a central pin placed inside an outer sleeve resulting in the expansion of the outer sleeve and the ultimate fixation of the device into bone. The Innovasive Devices 10mm Ligament Fastener will be offered in a single size, 10mm outside diameter and 20mm long.

The outer sleeve has threads on its exterior to hold to the bone and a central ID designed to accept the pin component.

The central pin has ribs along its length designed to expand the outer sleeve as it is placed down the sleeve inside diameter. In addition, the central pin incorporates a "side-plate" with two transverse pegs to hold a bone plug for bone-patella-bone fixation of the soft tissue graft. The front of the pin has a through-hole to accept a piece of suture to guide the pin into the sleeve during the deployment of the device. The suture is then removed and discarded.

{1}------------------------------------------------

#### 5. INTENDED USE:

The 10mm Ligament Fasteners are intended for use in the fixation of ligament and tendon bone block grafts in cruciate ligament reconstruction surgeries of the knee.

# 6. COMPARISON OF CHARACTERISTICS:

The Innovasive 10mm Ligament Fastener utilizes the same basic design for fixation into bone as the currently marketed 8mm Ligament Fastener (K970316). This design utilizes a central pin placed inside an outer sleeve resulting in the expansion of the outer sleeve and the ultimate fixation of the device into bone.

The outer sleeve is identical for both the 8mm and 10mm Ligament Fasteners. However, the central pin for the existing 8mm Ligament Fastener has a radius at the eyelet designed for direct fixation of the soft tissue graft; whereas the central pin for the proposed 10mm Ligament Fastener incorporates a "side-plate" with two transverse pegs to hold a bone plug for bone-patella-bone fixation of the soft tissue graft.

# 7. PERFORMANCE DATA:

The following performance data was provided in support of the substantial equivalence determination:

- 1. Mechanical Testing: Comparison of the ultimate holding strength in a bone model compared to the predicate device. The Innovasive 10mm Ligament Fastener holding strength was found to be equivalent to the strength of the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird in flight, composed of three curved lines. The bird is oriented towards the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Stephen M. Page Manager of Regulatory Affairs Innovasive Devices, Inc. 734 Forest Street 01752 Marlborough, Massachusetts

FEB | 8 1998

K974049 Re : LinX BT 10mm Ligament Fastener Trade Name: Regulatory Class: II Product Codes: MBI, HRX, and HWC Dated: January 23, 1998 Received: January 26, 1998

Dear Mr. Page:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with ---the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{3}------------------------------------------------

Page 2 - Mr. Stephen M. Page

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,

a M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE 10mm Ligament Fastener

The 10mm Ligament Fasteners are intended for use in the fixation of ligament and tendon bone block grafts in cruciate ligament reconstruction surgeries of the knee.

**Prescription Use**
(Per 21 CFR 801.109)

pcooddo

Sion-Off Divis Division of General Restorative De 974049 510(k) Numb

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K974049](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K974049)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
