← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K244025

# Argo Knotless GENESYS Anchor (K244025)

_Conmed Corporation · MBI · Jan 28, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K244025

## Device Facts

- **Applicant:** Conmed Corporation
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Jan 28, 2025
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.

## Device Story

The Argo Knotless GENESYS Anchor is an implantable bone fixation device used in arthroscopic or open orthopedic surgeries. It consists of a threaded bioabsorbable anchor, a PEEK-Optima eyelet, a #2 UHMWPE retention suture, and a single-use driver. The device requires a pre-prepared bone hole. During surgery, the physician inserts the anchor into the bone; the eyelet secures soft tissue (ligaments, tendons, or joint capsules) to the bone to stabilize the repair. The system supports up to six sutures or one bioresorbable reinforced implant. The device is supplied sterile for single use. By providing a secure mechanical attachment, it facilitates tissue healing during the postoperative immobilization period.

## Clinical Evidence

Bench testing only. Performance testing included insertion, cyclic, and ultimate pull displacement resistance. User validation was conducted using cadaveric medial collateral ligament/lateral ligament models. Packaging and labeling were also validated.

## Technological Characteristics

Materials: Bioabsorbable anchor, PEEK-Optima eyelet, #2 UHMWPE suture. Principle: Threaded mechanical bone fixation. Form factor: Threaded anchor with delivery driver. Connectivity: None. Sterilization: Ethylene oxide (ETO) to SAL 10^-6. Biocompatibility: Per ISO 10993-1 and FDA G95-1.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Argo Knotless® GENESYS™ SP Anchor, models SPKx ([K240090](/device/K240090.md))
- Argo Knotless® GENESYS™ Anchor, models KBCx ([K240090](/device/K240090.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 28, 2025

ConMed Corporation Dionne Sanders Sr. Manager, Regulatory Affairs 525 French Road Utica, New York 13502

Re: K244025

Trade/Device Name: Argo Knotless GENESYS Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: December 24, 2024 Received: December 27, 2024

Dear Dionne Sanders:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# CHRISTOPHER FERREIRA -S

Christopher Ferriera, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K244025

Device Name

Argo Knotless® GENESYSTM Anchor

Indications for Use (Describe)

Argo Knotless® GENESYSTM Anchor

The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures.

The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue. in conjunction with appropriate postoperative immobilization throughout the healing period.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) SUMMARY Argo Knotless® GENESYS ™ Anchor

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary for K244025.

#### l. SUBMITTER

Manufacturer: CONMED Corporation 525 French Road Utica, NY 13502

Official Contact Person: Dionne Sanders, JM, MS, CQA, RAC 525 French Road Utica, NY 13502 (O) 813-997-8126

Date Prepared: December 24, 2024

#### II. DEVICE NAME

III.

| Device Name:         | Argo Knotless® GENESYS ™ Anchor                                                                    |
|----------------------|----------------------------------------------------------------------------------------------------|
| Classification Name: | Fasteners, Fixation, Biodegradable, Soft Tissue<br>Fasteners, Fixation, Nondegradable, Soft Tissue |
| Regulatory Class:    | Class II, per 21 CFR Part 888.3030; 888.3040                                                       |
| Product Codes:       | MAI, MBI                                                                                           |
| PREDICATE DEVICE     |                                                                                                    |
| Device Name:         | Argo Knotless® GENESYS ™ SP Anchor, models SPKx<br>Argo Knotless® GENESYS ™ Anchor, models KBCx    |
| Company Name:        | CONMED Corporation                                                                                 |

#### IV. DEVICE DESCRIPTION

Argo Knotless® GENESYS ™ Anchor: The Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK Optima material. The device features a single use driver, a threaded anchor, a PEEK-Optima eyelet, a #2 UHMWPE, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures or one (1) bioresorbable reinforced implant. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.

{5}------------------------------------------------

#### V. INTENDED USE/ INDICATIONS FOR USE

## Argo Knotless® GENESYS™ Anchor

The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures.

The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.

#### VI. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A summary of the technological characteristics between the proposed and the predicate device is provided.

|                                        | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                        | Argo Knotless® GENESYS™ Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Argo Knotless® GENESYS™ Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 510k Number                            | K244025                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K240090                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended<br>Use/Indications for<br>Use | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Argo Knotless® GENESYS™ Anchor<br>The biocomposite suture anchor is intended to<br>reattach soft tissue to bone in orthopedic<br>surgical procedures.<br><br>The Argo Knotless® GENESYS™ Anchor may be<br>used in either arthroscopic or open surgical<br>procedures. After the suture is anchored to the<br>bone, it may be used to reattach soft tissue,<br>such as ligaments, tendons, or joint capsules to<br>the bone. The suture anchor system thereby<br>stabilizes the damaged soft tissue, in<br>conjunction with appropriate postoperative<br>immobilization throughout the healing period. |
| Contraindications                      | 1. Pathological conditions of bone<br>which would adversely affect the<br>Argo Knotless® GENESYS™ Anchor.<br>2. Pathological conditions in the<br>soft tissue to be repaired or<br>reconstructed which would<br>adversely affect suture fixation.<br>3. Physical conditions that would<br>eliminate, or tend to eliminate,<br>adequate implant support or<br>retard healing.<br>4. Conditions which tend to limit<br>the patient's ability or willingness<br>to restrict activities or follow | 1. Pathological conditions of bone which would<br>adversely affect the Argo Knotless® GENESYS™<br>Anchor.<br>2. Pathological conditions in the soft tissue to be<br>repaired or reconstructed which would<br>adversely affect suture fixation.<br>3. Physical conditions that would eliminate, or<br>tend to eliminate, adequate implant support or<br>retard healing.<br>4. Conditions which tend to limit the patient's<br>ability or willingness to restrict activities or<br>follow directions during the healing period.<br>5. Direct attachment of artificial ligaments or<br>other implants    |

{6}------------------------------------------------

|                                | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device                                                                                                                                                                                                                                                 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | directions during the healing<br>period.<br>5. Direct attachment of artificial<br>ligaments or other implants<br>except for bioresorbable<br>reinforced implants.<br>6. Foreign body sensitivity, known<br>or suspected allergies to implant<br>and/or instrument materials.<br>7. This device is not approved for<br>screw attachment or fixation to<br>the posterior elements (pedicles)<br>of the cervical, thoracic or lumbar<br>spine. | 6. Foreign body sensitivity, known or suspected<br>allergies to implant and/or instrument materials.<br>7. This device is not approved for screw<br>attachment or fixation to the posterior elements<br>(pedicles) of the cervical, thoracic or lumbar<br>spine. |
| How Supplied                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        | Sterile Anchor on a delivery driver                                                                                                                                                                                                                              |
| Single Use<br>/Reusable        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        | Single-Use Only                                                                                                                                                                                                                                                  |
| Sterilization                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        | ETO                                                                                                                                                                                                                                                              |
| Shelf-Life                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        | 18-months                                                                                                                                                                                                                                                        |
| Principle of<br>Operation      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        | Soft tissue to bone fixation                                                                                                                                                                                                                                     |
| Biocompatibility               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        | In accordance with ISO 10993-1 and FDA# G95-1                                                                                                                                                                                                                    |
| Design                         | Non-self-punching anchor                                                                                                                                                                                                                                                                                                                                                                                                                    | Self-punching anchor<br>Non-self-punching anchor<br>Non-self-punching, fully loaded anchor                                                                                                                                                                       |
| Packaging                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        | Packaged as a single-unit device                                                                                                                                                                                                                                 |
| Instrumentation                | Same, with addition of reamer and<br>drill guide                                                                                                                                                                                                                                                                                                                                                                                            | Class 1, Exempt instrumentation for initiating a<br>bone hole.                                                                                                                                                                                                   |
| Use with<br>resorbable implant | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                         | Specific SKU                                                                                                                                                                                                                                                     |

#### VII. PERFORMANCE DATA

Testing and analysis have been completed to demonstrate that the Argo Knotless® GENESYS™ Anchor performs as intended and is substantially equivalent to the predicate device. Additional evaluation for this modification includes the following.

- . Performance Testing (Insertion, Cyclic and Ultimate Pull Displacement Resistance)
- User Validation (Cadaver Medial Collateral Ligament/Lateral Ligament)
- Packaging and Labeling User Validation

#### VIII. CONCLUSION

The Argo Knotless® GENESYS™ Anchor is substantially equivalent in design, intended use, performance, and fundamental principle of operation. There are no additional questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K244025](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K244025)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
