K240835 · Skeletal Dynamics, Inc. · MBI · Jul 19, 2024 · Orthopedic
Device Facts
Record ID
K240835
Device Name
Suture Button Repair System
Applicant
Skeletal Dynamics, Inc.
Product Code
MBI · Orthopedic
Decision Date
Jul 19, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Suture Button Repair System is intended for fixation of bone and soft tissue in upper extremity orthopedic procedures requiring ligament or tendon reconstruction in the hand, wrist, forearm, elbow and shoulder.
Device Story
Suture Button Repair System consists of suture button implants and associated instruments for orthopedic fixation. Used in upper extremity procedures (hand, wrist, forearm, elbow, shoulder) for ligament or tendon reconstruction. Implants provided sterile for single use; instruments provided non-sterile for end-user sterilization. Device functions as bone fixation fastener to secure soft tissue to bone. Operated by orthopedic surgeons in clinical/OR settings. Output is mechanical fixation of tissue to bone, facilitating healing and reconstruction. Benefits include stable fixation during recovery.
Clinical Evidence
Bench testing only. Included sterilization validation (ISO 11135-1), biocompatibility (ISO 10993-1), packaging stability (ISO 11607-1), and mechanical performance testing (cyclic loading and ultimate load to failure).
Technological Characteristics
Metallic bone fixation fastener. Materials and performance requirements consistent with predicate. Sterilization via ethylene oxide per ISO 11135-1. Mechanical performance evaluated via cyclic loading and ultimate load to failure testing.
Indications for Use
Indicated for fixation of bone and soft tissue in upper extremity orthopedic procedures requiring ligament or tendon reconstruction in the hand, wrist, forearm, elbow, and shoulder.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
K023021 — MODIFICATION TO SMITH & NEPHEW TWINFIX TI QUICK T · Smith & Nephew, Inc. · Oct 4, 2002
K141263 — TOGGLELOC AND JUGGERLOC IMPLANTS/INSTRUMENTS/KITS · Biomet Manufacturing Corp · Nov 19, 2014
K020159 — SMITH & NEPHEW TWINFIX TI QUICK T · Smith & Nephew, Inc. · Mar 26, 2002
K993261 — TALON ANCHOR SNAP-PAK · Mitek Products · Dec 2, 1999
K141011 — G-FORCE GEN 2 SUTURE ANCHOR SYSTEM · Wrightmedicaltechnologyinc · Jul 30, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
July 19, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Skeletal Dynamics Inc Alexandra Rodriguez Rojas Regulatory Affairs Manager 7300 North Kendall Drive Suite 800 Miami, Florida 33156
Re: K240835
Trade/Device Name: Suture Button Repair System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: June 17, 2024 Received: June 17, 2024
Dear Alexandra Rodriguez Rojas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Digitally signed by JESSE MUIR JESSE MUIR -S Date: 2024.07.19 14:06:58 -04'00'
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### 510(k) SUMMARY Skeletal Dynamic's Suture Button Repair System
#### Submitter
Skeletal Dynamics, Inc. 7300 N. Kendall Drive, Suite 800 Miami, FL 33156 Phone : 305-596-7585 E-mail : arodriguez(@skeletaldynamics.com Contact Person: Alexandra Rodriguez Rojas Date Prepared: July 18, 2024
# Name and Classification
Trade Name: Suture Button Repair System Common Name: Fastener, Fixation, Nondegradable, Soft Tissue Classification Name: Smooth or threaded metallic bone fixation fastener Classification Number: 21 CFR \$888.3040 Regulatory Class: Class II Product Code: MBI
### Primary Predicate Device
Riverpoint OrthoButton AL (K171060)
#### Reference Devices
Hs Fiber (Polyblend), River Bond, Riversilk (Silk), Riverpro (Polypropylene), Riverlon (Nylon) Model Varies by Size/Need (K100006) Geminus Volar Distal Radius Plate System (K182492) Headless Compression Screw System (K143624) Mini Tightrope Ft Repair Kit (K071978)
#### Device Description
The Suture Button Repair System is comprised of the following Suture Button configurations in conjunction with accessories: (1) Distal Bicep Suture Button Repair configuration, (2) IOL Suture Button-Repair configuration and (3) CMC Suture Button Repair configuration. The implants are provided sterile for single use. The instruments are provided non-sterile and shall be sterilized by the end user.
#### Indications for Use
The Suture Button Repair System is intended for fixation of bone and soft tissue in upper extremity orthopedic procedures requiring ligament or tendon reconstruction in the hand, wrist, forearm, elbow and shoulder.
#### Performance Testing
Non-clinical performance testing for the Suture Button Repair System included sterilization validation per ISO11135-1 - Sterilization of Health Care Products Ethylene Oxide Part 1: Requirements for the Development, Validation, and Routine Control of a Sterilization Process
{4}------------------------------------------------
for medical devices, biocompatibility testing per ISO10993-1 - Biological Evaluation of Medical Devices, stability testing on the product and packaging per ISO 11607-1- Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems. The Suture Button Repair System was also evaluated for strength and elongation during cyclic loading and ultimate load to failure conditions.
# Substantial Equivalence and Comparison of Technical Characteristics
The Skeletal Dynamics Suture Button Repair System is as safe and effective as the previously cleared OrthoButton AL. The Skeletal Dynamics Suture Button Repair System has the same intended use, the same principles of operation, and similar technical characteristics as the predicate device. Both the Suture Button Repair System and the predicate device are sterilized using the same processes, are composed of the same materials, and are tested per the same performance requirements. Bench top performance testing demonstrated substantially equivalent mechanical performance as the predicate device. The minor difference in technical characteristics is limited to the Suture Button configurations. This difference does not raise new questions of safety or effectiveness; the Suture Button Repair System is substantially equivalent to the currently marketed predicate device.
# Conclusions
The Skeletal Dynamic's Suture Button Repair System is substantially equivalent to the predicate device identified in this premarket notification.
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240835
Device Name Suture Button Repair System
Indications for Use (Describe)
The Sutton Repair System is intended for fixation of bone and soft tissue in upper extremity orthopedic procedures requiring ligament or tendon reconstruction in the hand, wrist, forearm, elbow and shoulder.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.