The INFILOOP® Fixed Loop UHMWPE Suture Titanium Button is indicated in surgical procedures for soft tissue fixation to the bone. It is used for fixation of tendons and ligaments during orthopaedic reconstruction procedures such as Knee ligament reconstruction which includes anterior cruciate ligament (ACL), posterior cruciate ligament (PCL), medial collateral ligament (MCL), and lateral collateral ligament (LCL).
Device Story
Device consists of titanium button and UHMWPE suture loop; used for soft tissue-to-bone fixation during arthroscopic knee ligament reconstruction (ACL, PCL, MCL, LCL). Surgeon inserts device to secure tendons/ligaments to bone. Provides mechanical fixation; replaces need for traditional fasteners. Benefits include accurate fixation and compatibility with arthroscopic approaches. Performance verified via pull-out and cyclic loading tests.
Clinical Evidence
Bench testing only. Included biocompatibility (sensitization, reactivity, pyrogenicity, systemic toxicity, cytotoxicity, implantation), sterility, shelf-life, pull-out testing, and cyclic loading. Literature review supported the use of pure UHMWPE suture material over polyester-blended sutures.
Technological Characteristics
Materials: Titanium button, UHMWPE suture loop. Single-use, sterile (ETO). Dimensions/specs within range of predicates. Mechanical fixation principle. No software or energy source.
Indications for Use
Indicated for soft tissue-to-bone fixation in orthopedic reconstruction procedures, specifically knee ligament reconstruction (ACL, PCL, MCL, LCL).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Conmed Linvatec XO Button™ (Continous Loop) (K070780)
{1}------------------------------------------------
November 8, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.
Healthium Medtech Limited Pankaj Dawar, PhD Deputy General Manager Regulatory Affair 472-D, 13th Cross, 4th Phase, Peenya Industrial Area Bangalore, Karnataka 560058 India
Re: K232886
Trade/Device Name: INFILOOP® Fixed Loop UHMWPE Suture Titanium Button Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: September 13, 2023 Received: September 18, 2023
Dear Dr. Dawar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Jesse Muir -S Jesse Muir -S
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232886
#### Device Name
INFILOOP® Fixed Loop UHMWPE Suture Titanium Button
Indications for Use (Describe)
The INFILOOP® Fixed Loop UHMWPE Suture Titanium Button is indicated in surgical procedures for soft tissue fixation to the bone. It is used for fixation of tendons and ligaments during orthopaedic reconstruction procedures such as Knee ligament reconstruction which includes anterior cruciate ligament (ACL), posterior cruciate ligament (PCL), medial collateral ligament (MCL), and lateral collateral ligament (LCL).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in large, bold, sans-serif letters, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller, sans-serif font. The logo is simple and modern, and the red circle adds a pop of color.
510(k) Summary INFILOOP® Fixed Loop UHMWPE Suture Titanium Button
Image /page/4/Picture/2 description: The image contains the alphanumeric string "K232886" in the upper left corner. Below this is the word "Healthium" in a stylized font. To the right of the word is a logo consisting of two overlapping circles, one orange and one blue, with a white cross shape in the center.
## 510(k) Summary based on the 21 CFR 807.92
| Date Prepared | 12 September 2023 |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K232886 |
| Submitter | Healthium Medtech Limited<br>472-D, 13th Cross, 4th Phase,<br>Peenya Industrial Area,<br>Bangalore, India, 560058<br>Ph: +91 - 80 - 41868000 |
| Contact Person | Pankaj Dawar<br>Deputy General Manager<br>Regulatory Affairs<br>Ph: +91 - 80 - 41868000<br>pankaj.d@healthiummedtech.com |
| Name of Device | INFILOOP® Fixed Loop UHMWPE Suture Titanium Button |
| Common Name | Fixed Loop UHMWPE Suture Titanium Button |
| Product Code | MBI |
| Classification Name | 21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener |
| Regulatory Class | II |
| Predicate Device | Conmed Linvatec XO Button™ (Continous Loop) (K070780)-Primary Predicate<br>RIGIDLOOPTM Cortical Fixation System (K130814)-Additional Predicate |
| Reference Device | No reference device used |
| Purpose of Submission | Traditional 510(k) is submitted to obtain clearance for the INFILOOP® Fixed Loop<br>UHMPWE suture Titanium Button |
| Device Description | The use of the INFILOOP® Fixed Loop UHMWPE Suture Titanium Button provides<br>the orthopaedic surgeon a means of accurate fixation in ligament reconstructive<br>surgery of soft tissue to bone fixation. The fixation device allows for arthroscopic<br>ligament reconstruction approaches. |
| Intended Use/Indications for Use | The INFILOOP® Fixed Loop UHMWPE Suture Titanium Button is indicated in<br>surgical procedures for soft tissue fixation to the bone. It is used for fixation of<br>tendons and ligaments during orthopaedic reconstruction procedures such as Knee<br>ligament reconstruction which includes anterior cruciate ligament (ACL), posterior<br>cruciate ligament (PCL), medial collateral ligament (MCL), and lateral collateral<br>ligament (LCL) |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Sironix. The logo consists of the word "SIRONIX" in a sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the text "Arthroscopy Solutions" in a smaller font.
Image /page/5/Picture/2 description: The image contains the text "K232886" at the top left corner. Below the text is a logo for a company called "Healthium". The logo consists of the company name in a stylized font and a circular graphic with a cross shape in the middle.
Page 2 of 4
## Comparison of the Technological Characteristics with Predicate Device
| S. No | Parameters | Conmed Linvatec<br>XO Button™<br>(Continous Loop)<br>(K070780)<br>(Primary Predicate) | RIGIDLOOP™<br>Cortical Fixation<br>System<br>(K130814)<br>(Additional Predicate) | INFILOOP® Fixed<br>Loop UHMWPE<br>Suture Titanium<br>Button<br>(Subject Device) | Comments |
|----------|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| 1. | Manufacturer | ConMed Linvatec | Medos International<br>SARL | Healthium Medtech<br>Limited | - |
| 2. | Product Code | MBI (Classification<br>Product Code)<br>GAT (Subsequent<br>Product Code) | MBI | MBI | Similar to<br>predicate<br>device |
| 3. | Regulation<br>Number | 21 CFR 888.3040 | 21 CFR 888.3040 | 21 CFR 888.3040 | Similar to<br>predicate<br>device |
| 4. | Classification | Class II | Class II | Class II | Similar to<br>predicate<br>device |
| 5. | Intended Use/<br>Indications for<br>Use | The Conmed Linvatec<br>XO Button™ with<br>continuos loop is<br>intended to provide<br>suspension fixation for<br>soft tissue to bone in<br>the repair of the<br>natural ligament or<br>tendon disruption or<br>reconstruction of a<br>ligament using soft<br>tissue grafts. Examples<br>of such procedures<br>include anterior<br>cruciate ligament,<br>posterior cruciate<br>ligament, medial<br>collateral ligament,<br>and lateral collateral<br>ligament. | The RIGIDLOOP™<br>Cortical Fixation<br>System is used for<br>the fixation of soft<br>tissue to bone in<br>orthopedic<br>procedures such as<br>ACL repair | The INFILOOP®<br>Fixed Loop<br>UHMWPE Suture<br>Titanium Button is<br>indicated in surgical<br>procedures for soft<br>tissue fixation to the<br>bone. It is used for<br>fixation of tendons<br>and ligaments during<br>orthopaedic<br>reconstruction<br>procedures such as<br>Knee ligament<br>reconstruction which<br>includes anterior<br>cruciate ligament<br>(ACL), posterior<br>cruciate ligament<br>(PCL), medial<br>collateral ligament<br>(MCL), and lateral<br>collateral ligament<br>(LCL). | Similar to<br>predicate<br>device |
| 6. | Button and<br>Loop Material | Button: Titanium<br>Loop: UHMWPE | Button: Titanium<br>Loop: UHMWPE | Button: Titanium<br>Loop: UHMWPE | Similar to<br>predicate<br>device |
| S.<br>No | Parameters | Conmed Linvatec<br>XO Button™<br>(Continous Loop)<br>(K070780)<br>(Primary Predicate) | RIGIDLOOP™<br>Cortical Fixation<br>System<br>(K130814)<br>(Additional<br>Predicate) | INFILOOP® Fixed<br>Loop UHMWPE<br>Suture Titanium<br>Button<br>(Subject Device) | Comments |
| 7. | Suture<br>Material | Polyester, USP#5 UHMWPE,<br>USP#3-4 | UHMWPE,<br>USP#5 Polyester,<br>USP#5 | UHMWPE,<br>USP#5 UHMWPE,<br>USP#5 | Refer the<br>SE<br>Analysis |
| 8. | Single<br>Use/Reuse | Single Use | Single Use | Single Use | Similar to<br>predicate<br>device |
| 9. | Shelf Life | 5 Years | 5 Years | 5 Years | Similar to<br>predicate<br>device |
| 10. | Sterilization<br>Method | ETO | ETO | ETO | Similar to<br>predicate<br>device |
| 11. | Performance<br>Data | Pull Out Testing Cyclical Loading | Pull Out Testing | Pull Out Testing Cyclical Loading | Similar to<br>predicate<br>device |
| 12. | Safety Data | No Data Available | No Data<br>Available | Skin Sensitization Intracutaneous reactivity Material Mediated Pyrogenicity Acute Systemic Toxicity In-vitro cytotoxicity Bone Implantation Intramuscular Implantation Bacterial Endotoxin USP<br><85> MRI Compatibility | - |
Registered Office: 472/D, 13th Cross, 4th Phase, Peenya Industrial Area, Bengaluru – 560058, India
www.healthiummedtech.com | CIN : U03311KA1992PLC013831
{7}------------------------------------------------
510(k) Summary INFILOOP® Fixed Loop UHMWPE Suture Titanium Button
Image /page/6/Picture/1 description: The image shows the logo for Sironix Arthroscopy Solutions. The word "SIRONIX" is written in a bold, sans-serif font, with the "O" in "SIRONIX" being a red circle. Below the word "SIRONIX" is the phrase "Arthroscopy Solutions" in a smaller, sans-serif font. The logo is simple and modern, and the red circle adds a pop of color.
Image /page/6/Picture/2 description: The image contains the text 'K232886' in the upper left corner. Below this text is the word 'Healthium' in a stylized font. To the right of 'Healthium' is a logo consisting of a blue circle with a white cross inside, partially obscured by an orange circle with a white cross inside.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Healthium, a medical device company. The logo consists of the word "Healthium" in a blue, sans-serif font, with a stylized graphic to the right. The graphic is composed of two overlapping circles, one orange and one blue, with a white cross shape in the center where they intersect. The text "K232886" is located in the upper left corner of the image.
## SE Analysis
Predicate devices are preloaded with one UHMWPE suture and one Polyester Suture, whereas the subject device is preloaded with UHMWPE sutures. From the literature study, it can be inferred that the Polyester suture had lower ultimate load than all groups of sutures used in the study except the suture composed of polyester and UHMWPE (P<. 05).
Pure UHMWPE suture had higher ultimate failure load than sutures composed of either polyester or polyester plus UHMWPE (P<.05). Predominant failure mode was suture cutting through the meniscus for the groups except for polyester suture which failed by suture rupture. This literature study shows that Pure UHMWPE material has better ultimate load than the material used in predicate and this change is not raising any questions on the safety and efficacy of the subject device.
### Comparison and summary of the technological characteristics
The proposed INFILOOP® Fixed Loon UHMWPE Suture Titanium Button (referred to as the "subjected device") shares similar basic design features and materials of construction for the button and loop when compared to the predicate devices. The specifications and dimensions of the subjected device fall within the range of those of the predicate devices. Minor differences in technological characteristics do not raise any safety or effectiveness concerns.
### Comparison and summary of the safety, performance and non-clinical testing data
Verification activities were performed on the implant and its predicates. Testing assessments included biocompatibility, pyrogenicity (both material andotoxin) in accordance with USP <85>, sterility, shelf life, pull-out testing, and cyclic loading (as mentioned in the comparison table). The results of performance testing have demonstrated that the proposed devices are suitable for their intended use.
## Conclusion
The INFILOOP® Fixed Loop UHMWPE Suture Titanium Button is substantially equivalent to the predicate devices. Any differences between the subject device and the predicate devices do not raised questions concerning safety and effectiveness.
Based on the indications for use, technological characteristics, and the summary of data submitted, it is determined that the proposed device is substantially equivalent to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.