← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K232492

# 2.7 mm GRYPHON Flex Knotless PEEK Anchor; 2.7 mm GRYPHON Flex Knotless Biocomposite Anchor (K232492)

_Medos International SARL · MBI · Oct 6, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K232492

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Oct 6, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The GRYPHON™ Flex Knotless Anchor is indicated for use for reattachment of soft tissue to bone for the following procedures: Shoulder: Bankart Repair, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction, Superior Capsule Reconstruction (Glenoid Only); Hip: Acetabular Labral Repair; Knee: MCL, LCL, Iliotibial Band Tenodesis; Foot/Ankle: Lateral Stabilization, Medial Stabilization.

## Device Story

The GRYPHON™ Flex Knotless Anchor is a two-piece orthopedic fixation system consisting of a permanent anchor and a disposable inserter. The device is used by surgeons in clinical settings to reattach soft tissue to bone. The anchor is preloaded on the inserter; a Nitinol distal tip retains the suture and anchor during mallet-driven insertion into bone. The inserter is available in straight or curved configurations to improve maneuverability within the joint space. By providing a knotless fixation method, the device simplifies soft tissue repair procedures, potentially reducing surgical time and improving outcomes in shoulder, hip, knee, and foot/ankle repairs.

## Clinical Evidence

Bench testing only. Performance evaluation included in-vitro fixation strength at 0, 6, 12, 13, and 26 weeks, insertion torque, torque to failure, and cyclic/post-cyclic fixation at time zero. Biological safety evaluated per ISO 10993-1; sterilization validated per ANSI/AAMI/ISO 11135:2014; non-pyrogenicity confirmed per ANSI/AAMI ST-72:2011.

## Technological Characteristics

2.7mm knotless suture anchor. Materials: PEEK or Biocomposite (Polylactic Acid and Tricalcium Phosphate). Two-piece design with permanent anchor and disposable inserter (straight or curved). Sterilization: Ethylene Oxide (EO) per ANSI/AAMI/ISO 11135:2014. Single-use.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex PushLock Anchors ([K101679](/device/K101679.md))

## Reference Devices

- HEALIX ADVANCE™ SP PEEK Anchor ([K182941](/device/K182941.md), [K201883](/device/K201883.md))
- HEALIX ADVANCE™ SP BIOCOMPOSITE Anchor ([K191242](/device/K191242.md), [K201883](/device/K201883.md))
- GRYPHON™ Anchor W/ Proknot Technology ([K140643](/device/K140643.md))
- ORTHOCORD® Suture ([K040004](/device/K040004.md))
- DYNACORD™ Suture ([K181182](/device/K181182.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 6, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medos International SARL % Elizabeth Messana Senior Regulatory Specialist DePuy Synthes Mitek Sports Medicine 325 Paramount Drive Raynham, Massachusetts 02767

## Re: K232492

Trade/Device Name: 2.7 mm GRYPHON™ Flex Knotless PEEK Anchor; 2.7 mm GRYPHON™ Flex Knotless Biocomposite Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, MAI Dated: August 16, 2023 Received: August 17, 2023

# Dear Elizabeth Messana:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sara S. Thompson -S

For

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health 

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232492

#### Device Name

GRYPHON™ Flex Knotless Anchor

#### Indications for Use (Describe)

The GRYPHON™ Flex Knotless Anchor is indicated for use for reattachment of soft tissue to bone for the following procedures:

|            | Indication                                     | PEEK | BIOCRYL |
|------------|------------------------------------------------|------|---------|
| Shoulder   | Bankart Repair                                 | X    | X       |
|            | SLAP Lesion Repair                             | X    | X       |
|            | Capsular Shift or Capsulolabral Reconstruction | X    | X       |
|            | Superior Capsule Reconstruction (Glenoid Only) | X    | X       |
| Hip        | Acetabular Labral Repair                       | X    | X       |
| Knee       | MCL                                            | X    | X       |
|            | LCL                                            | X    | X       |
|            | Iliotibial Band Tenodesis                      | X    | X       |
| Foot/Ankle | Lateral Stabilization                          | X    | X       |
|            | Medial Stabilization                           | X    | X       |

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

EF PSC Publishing Services (301) 443-6740

{5}------------------------------------------------

# 510(k) SUMMARY GRYPHON™ Flex Knotless Anchors Date Prepared: 8/15/2023

| Submitter's<br>Name and<br>Address | DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br><br>On behalf of:<br>Medos International SARL<br>Chemin-Blanc 38, Le Locle Neuchatel<br>CH 2400, Switzerland                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                        |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Contact Person                     | Elizabeth P. Messana<br>Senior Regulatory Affairs Specialist<br><br>DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Telephone: 508-828-3150<br>Email: emessan1@its.jnj.com |
| Name of Medical<br>Device          | Proprietary Name:<br>A) GRYPHON™ Flex Knotless PEEK Anchor<br>B) GRYPHON™ Flex Knotless BIOCOMPOSITE Anchor<br><br>Classification Name:<br>A) Smooth or threaded metallic bone fixation fastener<br>B) Single/multiple component metallic bone fixation appliances<br>and accessories<br><br>Product Code:<br>A) MBI<br>B) MAI<br><br>Common Name:<br>Suture Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                        |
| Substantial Equivalence            | The GRYPHON Flex Knotless PEEK and BIOCOMPOSITE<br>Anchors are substantially equivalent to:<br>K101679, Arthrex PushLock Anchors (BIOCOMPOSITE)              Reference Devices:<br>K182941 and K201883- HEALIX ADVANCE™ SP PEEK<br>Anchor (MEDOS International), K191242 and K201883 HEALIX ADVANCE™ SP<br>BIOCOMPOSITE Anchor (MEDOS International), K140643 GRYPHON™ Anchor W/ Proknot Technology<br>(DePuy Mitek), K040004 ORTHOCORD® Suture (DePuy Mitek), K181182 DYNACORD™ Suture (MEDOS International).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                        |
| Device Classification              | GRYPHON Flex Knotless PEEK Anchor is classified as:<br>Fastener, Fixation, Nondegradable, Soft Tissue, classified as Class<br>II, product code MBI, regulated under 21 CFR 888.3040.<br><br>GRYPHON Flex Knotless BIOCOMPOSITE Anchor is classified as:<br>Single/multiple component metallic bone fixation appliances and<br>accessories, classified as Class II, product code MAI, regulated<br>under 21 CFR 888.3030.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                        |
| Device Panel                       | Orthopedic Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                        |
| Device Description                 | The proposed GRYPHON Flex Knotless PEEK and<br>BIOCOMPOSITE Anchors is a line extension to the currently<br>marketed GRYPHON Anchor family. The proposed product is a<br>knotless anchor intended for fixation of soft tissue to bone in<br>various soft tissue repair procedures. In its finished good<br>configuration, the proposed device is a two-piece design; comprised<br>of a permanent implant (anchor) component and a disposable<br>(inserter) component. The 2.7mm anchor implant is preloaded on<br>the inserter and a Nitinol distal tip retains the suture and the anchor<br>while the device is malleted into the bone. The design offers<br>surgeons maneuverability within the joint space. The disposable<br>inserter will be offered in two design configurations, straight or<br>curved to accommodate varying anatomies.<br><br>The proposed GRYPHON Flex Knotless PEEK and<br>BIOCOMPOSITE Anchors will be available in a 2.7 mm size,<br>molded from either Polyetheretherketone (PEEK) material or |                                                        |

{6}------------------------------------------------

{7}------------------------------------------------

|                               | Absorbable Biocryl Biocomposite ((Polylactic Acid (PLA) and Tricalcium Phosphate (TCP)) material.                                                                                                                                                |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                               | Both the GRYPHON Flex Knotless PEEK and BIOCOMPOSITE Anchors are provided sterile via Ethylene Oxide (EO) sterilization and are for single use only.                                                                                             |
| Technological Characteristics | The suture anchor design, principal of operation and indications for use are similar when compared to the predicate device- Arthrex PushLock (BIOCOMPOSITE) Anchors, (K101679). A subset of indications will be pursued for the proposed device. |
|                               | The PEEK material of the proposed device is identical to that present in the reference device- K182941 and K201883- HEALIX ADVANCE™ SP PEEK Anchor.                                                                                              |
|                               | The Biocomposite material of the proposed device is identical to that present in the reference device- K191242 and K201883- HEALIX ADVANCE™ SP BIOCOMPOSITE Anchor                                                                               |

The GRYPHON™ Flex Knotless Anchor is indicated for use for reattachment of soft tissue Indications for to bone for the following procedures: Use

|            | Indication                                     | PEEK | BIOCRYL |
|------------|------------------------------------------------|------|---------|
| Shoulder   | Bankart Repair                                 | X    | X       |
|            | SLAP Lesion Repair                             | X    | X       |
|            | Capsular Shift or Capsulolabral Reconstruction | X    | X       |
|            | Superior Capsule Reconstruction (Glenoid Only) | X    | X       |
| Hip        | Acetabular Labral Repair                       | X    | X       |
| Knee       | MCL                                            | X    | X       |
|            | LCL                                            | X    | X       |
|            | Iliotibial Band Tenodesis                      | X    | X       |
| Foot/Ankle | Lateral Stabilization                          | X    | X       |
|            | Medial Stabilization                           | X    | X       |

#### Non-clinical Verification activities were performed on the proposed device and its predicate. Non-Testing clinical performance testing included evaluation of in-vitro fixation strength at zero, six, twelve, thirteen and twenty-six weeks, insertion torque, torque to failure and cyclic & post cyclic fixation at time zero.

{8}------------------------------------------------

| Ethylene Oxide Sterilization was validated according to ANSI/AAMI/ISO 11135: 2014 to a<br>SAL of 1 x 10-6.                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                            |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EO residuals were tested per AAMI/ANSI/ISO 10993-7:2008.                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                            |
| Biological safety was evaluated according to ISO 10993-1 for devices having direct patient<br>contact, inclusive of raw materials of components, manufacturing processes, sterilization, and primary<br>packaging materials.                   |                                                                                                                                                                                                                                                                                                                                                            |
| The proposed device has been determined to be non-pyrogenic per the requirements set forth<br>in ANSI/AAMI ST-72:2011, United States Pharmacopeia (USP), and European<br>Pharmacopeia (EP) using the bacterial endotoxin testing (BET) method. |                                                                                                                                                                                                                                                                                                                                                            |
| Safety and<br>Performance                                                                                                                                                                                                                      | Results of non-clinical performance testing have demonstrated that the proposed devices are<br>suitable for their intended use.                                                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                | Based on similarities in the indications for use, technological characteristics, and<br>performance testing in comparison to the predicate devices, the proposed GRYPHON Flex<br>Knotless PEEK and BIOCOMPOSITE Anchors have demonstrated that they are substantially<br>equivalent to the predicate device under the Federal Food, Drug and Cosmetic Act. |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K232492](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K232492)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
