← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K231322

# Fix2Lock(PEEK Self Punching) (K231322)

_Osteonic Co., Ltd. · MBI · Jun 7, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K231322

## Device Facts

- **Applicant:** Osteonic Co., Ltd.
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Jun 7, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; Orthopedic surgery for shoulders, knees, foot/ankle, hand/wrist and elbow.

## Device Story

Fix2Lock is a sterile, single-use orthopedic bone fixation fastener. It consists of a PEEK (polyether ether ketone) anchor, non-absorbable suture, and a driver assembly (shaft and handle). Used by surgeons in orthopedic procedures to reattach soft tissue to bone. The device is inserted into bone to provide mechanical fixation. It functions similarly to existing suture anchors, utilizing the material properties of PEEK for biocompatibility and structural integrity. The device is provided sterile and is intended for single use only.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing verified design specifications, mechanical performance (ASTM F543), material compliance (ASTM F2026, ASTM F2848), and sterilization/packaging validation (ISO 11135, ISO 11138, ISO 11140, ISO 11737, ISO 11607, ASTM F1980, ASTM F88, ASTM F1929). Bacterial endotoxin testing performed per USP <85> and <161>.

## Technological Characteristics

Materials: PEEK (ASTM F2026) anchor, non-absorbable suture, needle. Sizes: 2.3mm, 4.5mm, 4.8mm, 5.5mm. Sterilization: Ethylene oxide (EtO). Single-use, sterile. Mechanical performance tested per ASTM F543.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Fix2Lock (PEEK Self Punching) ([K202763](/device/K202763.md))

## Reference Devices

- 2.0 PK Suture anchor T and 2.0 PK Suture anchor S ([K081511](/device/K081511.md))
- Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching) ([K202806](/device/K202806.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an eagle and text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the text "U.S. FOOD & DRUG ADMINISTRATION" is written in blue.

June 7, 2023

Osteonic Co., Ltd. % Sanglok Lee Manager Wise Company Inc. #507, #508, 166 Gasandigital 2-ro, Geumcheon-gu Seoul, 08507 Korea, South

Re: K231322

Trade/Device Name: Fix2Lock(PEEK Self Punching) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 8, 2023 Received: May 8, 2023

Dear Sanglok Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sara S. Thompson -S

For

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231322

Device Name Fix2Lock (PEEK Self Punching)

Indications for Use (Describe)

Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; Orthopedic surgery for shoulders, knees, foot/ankle, hand/wrist and elbow.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### Date of Submission: 2023.06.07

#### 01. Applicant

OSTEONIC Co., Ltd.

405Ho, 505-2Ho, 505-3Ho, 508Ho, 902Ho, 1004Ho, 1005Ho, 1202Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea, 102Ho, 103Ho, 24, Digital-ro 27-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8456 FAX: +82-2-6902-8401 Email: dakham@osteonic.com

## 02. Submission Correspondent

Sanglok, Lee Wise COMPANY Inc. #507, #508, 166 Gasan digital 2-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org

## 03. Subject Device Identification

Trade Name: Fix2Lock (PEEK Self Punching) Common Name: fastener, fixation, non-biodegradable, soft tissue Classification Name: Smooth or threaded metallic bone fixation fastener Classification Product Code: MBI Panel: Orthopedic Regulation Number: 21 CFR 888.3040 Device Class: II

#### 04. Predicate Device

510(k) Number: K202763 Device Name: Fix2Lock (PEEK Self Punching) Manufacturer: Osteonic Co., Ltd.

#### 05. Reference Device

510(k) Number: K081511 Device Name: 2.0 PK Suture anchor T and 2.0 PK Suture anchor S Manufacturer: Smith &Nephew, Inc.

510(k) Number: K202806

Device Name: Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching) Manufacturer: Osteonic Co., Ltd.

#### 06. Device Description

This product is used for orthopedic surgery, which soft tissue fix such as ligaments, tendons, and capsules to bone. The implant part is made of PEEK (polyether ether ketone, ASTM F2026) with non-absorbable suture and consists of Driver Handle and Driver Shaft for anchor insertion. This product is sterilized product and single use only.

#### 07. Indication for use

Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; Orthopedic surgery for shoulders, knees, foot/ankle, hand/wrist and elbow.

{4}------------------------------------------------

# 08. Non-Clinical Test Conclusion

Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device with the following standards:

# - Material

- ASTM F2026: 2017 Standard specification for polyetheretherketone(PEEK) polymers for surgical implant applications

- ASTM F2848: 2017 Standard Specification for Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns

and Surgical Implants

# · Mechanical performance

- ASTM F543: 2013 Standard specification and test methods for metallic medical bone screws
- · Sterilization, shelf life and packaging for sterile product
- ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements for the development validation and routine control of a sterilization and routine control of a sterilization process for medical devices
- ISO 11138-1:2006, Sterilization of health care products Biological indicators Part 1: General requirements
- ISO 11138-2:2009, Sterilization of health care products -- Biological indicators -- Part 2: Biological indicators for ethylene oxide sterilization processes
- ISO 11140-1:2014, Sterilization of health care products Chemical indicators Part 1: General requirements
- ISO 11737-1:2018 Sterilization of medical devices Microbiological methods- Part 1: Estimation of population of microorganisms on products
- ISO 11737-2:2009 Sterilization of medical devices Microbiological methods- Part 2: Tests of sterility performed in the validation of a sterilization process
- ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system
- ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials.
- ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration

# ▪ Bacterial Endotoxin

- USP <85> Bacterial Endotoxin Test
- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests

{5}------------------------------------------------

# 09. Comparison of Technological Similarities and Differences Table 1: Substantial Equivalence Comparison

| Product<br>Name         | Modified(Subject) Device<br>Fix2Lock (PEEK Self<br>Punching), K231322                                                                                                                                   | Unmodified(Predicate)<br>Device<br>Fix2Lock (PEEK Self<br>Punching), K202763                                                                                       | REFERENCE Device<br>2.0 PK Suture anchor T and<br>2.0 PK Suture anchor S,<br>K081511                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Equivalence<br>Discussion |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Product<br>code         | MBI                                                                                                                                                                                                     | MBI                                                                                                                                                                | MBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                      |
| Regulatory<br>class     | Class II                                                                                                                                                                                                | Class II                                                                                                                                                           | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                      |
| Regulation<br>Number    | 21 CFR 888.3040                                                                                                                                                                                         | 21 CFR 888.3040                                                                                                                                                    | 21 CFR 888.3040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                      |
| Intended<br>use         | Fix2Lock is intended use<br>for fixation of soft tissue to<br>bone, using suture, in the<br>following procedure;<br>Orthopedic surgery for<br>shoulders, knees,<br>foot/ankle, hand/wrist and<br>elbow. | Fix2Lock is intended use for<br>fixation of soft tissue to<br>bone, using suture, in the<br>following procedure;<br>Orthopedic surgery for<br>shoulders and knees. | The 2.0 PK Suture Anchor T<br>and 2.0 PK Suture Anchor S<br>is intended for use for<br>reattachment of soft tissue<br>to bone for the following<br>indication:<br><b>Shoulder:</b> Capsular<br>stabilization - Bankart<br>repair, Anterior shoulder<br>instability, Slap lesion repair,<br>Capsular shift or<br>capsulolabral<br>reconstruction,<br>Acromilavicular separation<br>repairs, Deltoid repairs,<br>Rotator cuff repairs and<br>Biceps tenodesis<br><b>Foot/ankle:</b> Hallux valgus<br>repairs, Medial or lateral<br>instability<br>repairs/reconstruction,<br>Achilles tendon<br>repairs/reconstruction, Mid-<br>foot reconstruction,<br>Metatarsal ligament/tendon<br>repairs/reconstruction and<br>Bunionectomy.<br><b>Elbow, Wrist, and Hand:</b><br>Biceps tendon<br>reattachment, Ulnar or radial<br>collateral ligament<br>reconstruction, Lateral<br>epicondylitis repair, and<br>Scapholunate ligament<br>reconstruction.<br><b>Knee:</b> Extra-capsular<br>repairs - Medial collateral<br>ligament, Lateral collateral<br>ligament, Posterior oblique<br>ligament, Patellar<br>realignment and tendon<br>repairs - Vastus medialis<br>obliquus advancement and<br>Iliotibial band tenodesis. | Same                      |
| Operating<br>Principles | This product is used for<br>orthopedic surgery, which<br>soft tissue fix such as<br>ligaments, tendons, and<br>capsules to bone.                                                                        | This product is used for<br>orthopedic surgery, which<br>soft tissue fix such as<br>ligaments, tendons, and<br>capsules to bone.                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                           |
| Material                | PEEK anchor and non-<br>absorbable sutures,<br>needle                                                                                                                                                   | PEEK anchor and non-<br>absorbable sutures.                                                                                                                        | PEEK. Non-absorbable<br>suture, Stainless steel<br>needle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar                   |
| Structure               | This product consists of an<br>implanted anchor, non-<br>absorbable suture, driver<br>shaft and handle                                                                                                  | This product consists of an<br>implanted anchor, non-<br>absorbable suture, driver<br>shaft and handle                                                             | The 2.0 PK Suture Anchor T<br>and the 2.0 PK Suture<br>Anchor S consist of a non-<br>absorbable suture anchor<br>with attached non-<br>absorbable suture(s)<br>preassembles to a stainless<br>steel insertion device, and is<br>provided sterile, for single<br>use only. Both anchor<br>models achieve fixation via<br>ribbed/threaded design<br>characteristics that are<br>consistent with the repair of<br>soft tissue to bone.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                      |
| Product<br>Size         | 2.3mm, 4.5mm, 4.8mm,<br>5.5mm type                                                                                                                                                                      | 4.5mm, 5.5mm type                                                                                                                                                  | 2.03mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Similar                   |
| Sterilization           | Sterile (EtO sterilization)                                                                                                                                                                             | Sterile (EtO sterilization)                                                                                                                                        | Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                      |
| Single Use/<br>Reuse    | Single use                                                                                                                                                                                              | Single use                                                                                                                                                         | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                      |
| Packaging               | 1 EA / BOX                                                                                                                                                                                              | 1 EA / BOX                                                                                                                                                         | 1 EA / BOX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                      |
| Shelf life              | 3Years                                                                                                                                                                                                  | 5Years                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar                   |

{6}------------------------------------------------

# K231322

The Product Size of subject device is similar with the predicate device and the safety was evaluated as the performance bench test.

Reference device had been added to demonstrate the equivalency with the marketed device. This is to resolve that the needles are not applied for unmodified device.

# 10. Substantially Equivalent Conclusion

Based on above, the subject device, Fix2Lock (PEEK Self Punching), is determined to be Substantially Equivalent (SE) to the Unmodified(Predicate) device, Fix2Lock (PEEK Self Punching) (K202763) ,in respect of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K231322](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K231322)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
