← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K230094

# Responsive Arthroscopy Stealth and Mini Stealth All-Suture Anchors (K230094)

_Responsive Arthroscopy, LLC · MBI · Aug 25, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K230094

## Device Facts

- **Applicant:** Responsive Arthroscopy, LLC
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Aug 25, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Responsive Arthroscopy Stealth and Mini Stealth All-Suture Anchors are intended to be used for fixation of soft tissue to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Repair. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair, Digital Tendon Transfers. Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. Hand/Wrist: Scapholunate Ligament Reconstruction of Collateral Ligaments, Carpal Ligament Reconstruction, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP Joints for all Digits, Digital Tendon Repairs. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair.

## Device Story

All-suture anchors for soft tissue-to-bone fixation; comprised of braided UHMWPE suture material. Push-in design; inserted through pilot hole (2.1mm or 3.0mm); deployed against inserter tip below cortical bone. Four longitudinally arranged suture bundles bunch and expand radially to achieve fixation. Delivered arthroscopically by surgeons using disposable inserters, drills, guide tubes, and probes. Provides secure fixation point for reattaching soft tissue; facilitates repair of tendons and ligaments. Benefits include minimally invasive soft tissue reattachment.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included insertion force testing, cyclic pullout force testing, shelf-life validation (ASTM F1980), extractable residue testing (ASTM F2459), and cytotoxicity testing (ISO 10993-5). All testing met acceptance criteria.

## Technological Characteristics

Materials: Braided nonabsorbable UHMWPE. Design: Soft anchor, push-in, four-bundle radial expansion. Sizes: 2.1mm and 3.0mm pilot hole diameters. Delivery: Pre-loaded on disposable inserters. Sterilization: Ethylene oxide (EO). Standards: ASTM F1980 (shelf-life), ASTM F2459 (extractables), ISO 10993-5 (cytotoxicity).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Stryker ICONIX All-Suture Anchor System ([K133671](/device/K133671.md))
- Stryker ICONIX TT All-Suture Anchor System ([K170098](/device/K170098.md))

## Reference Devices

- HS Fiber (Polyblend) ([K100006](/device/K100006.md))
- HS SutureTape ([K153307](/device/K153307.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 25, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Responsive Arthroscopy LLC Garrett Ahlborg Director of Regulatory, Quality and Compliance 701 N. 3rd Street, Suite 208 Minneapolis, Minnesota 55401

### Re: K230094

Trade/Device Name: Responsive Arthroscopy Stealth and Mini Stealth All-Suture Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: August 1, 2023 Received: August 1, 2023

### Dear Garrett Ahlborg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sara S. Thompson -S

For

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230094

#### Device Name

Responsive Arthroscopy Stealth and Mini Stealth All-Suture Anchors

#### Indications for Use (Describe)

The Responsive Arthroscopy Stealth and Mini Stealth All-Suture Anchors are intended to be used for fixation of soft tissue to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Repair.

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair, Digital Tendon Transfers.

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.

Hand/Wrist: Scapholunate Ligament Reconstruction of Collateral Ligaments, Carpal Ligament Reconstruction, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP Joints for all Digits, Digital Tendon Repairs.

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair.

| Type of Use (Select one or both, as applicable)                                                                          |                                                                                                                         |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

| DATE PREPARED:               | August 2, 2023                                                                                                       |
|------------------------------|----------------------------------------------------------------------------------------------------------------------|
| SUBMITTER INFORMATION:       | Responsive Arthroscopy LLC<br>701 N. 3rd Street, Suite 208<br>Minneapolis, MN 55401                                  |
| ESTABLISHMENT REGISTRATION:  | 3015200759                                                                                                           |
| CONTACT INFORMATION:         | Garrett Ahlborg<br>Director of Regulatory, Quality and Compliance<br>(612) 532-6800<br>Gahlborg@responsivesports.com |
| DEVICE INFORMATION:          |                                                                                                                      |
| Trade Name:                  | Responsive Arthroscopy Stealth and Mini Stealth All-Suture<br>Anchors                                                |
| Common Name:                 | Suture Anchor                                                                                                        |
| Classification Name:         | Smooth or threaded metallic bone fixation fastener                                                                   |
| Product Code:                | MBI                                                                                                                  |
| Classification:              | Class II                                                                                                             |
| Regulation Number:           | 21 CFR 888.3040                                                                                                      |
| Primary Predicate Device:    | Stryker ICONIX All-Suture Anchor System (K133671)                                                                    |
| Additional Predicate Device: | Stryker ICONIX TT All-Suture Anchor System (K170098)                                                                 |
| Reference Devices:           | HS Fiber (Polyblend) (K100006)<br>HS SutureTape (K153307)                                                            |

The predicate and reference devices have not been subject to any design-related recalls.

### DEVICE DESCRIPTION:

The Stealth and Mini Stealth All-Suture Anchors intended for the fixation of soft tissue to bone. The All-Suture Anchors feature a push-in design and are comprised entirely of suture material configured to provide an anchor in bone. The anchors create a secure fixation point for the reattachment of soft tissue to bone when they are inserted through a pilot hole and deployed against the inserter tip below cortical bone in the desired anatomy. The All-Suture Anchors feature four longitudinally arranged round bundles that bunch together and expand radially when deployed to achieve fixation. The All-Suture Anchors may be delivered arthroscopically using inserters and surgical instruments such as drills, guide tubes, and probes.

The subject device systems include two anchor sizes, the Stealth All-Suture Anchor which is designed for a 3.0mm diameter pilot hole and the Mini Stealth All-Suture Anchor which is designed for a 2.1mm diameter pilot hole. Both anchor systems are comprised of braided nonabsorbable ultra-high molecular weight polyethylene (UHMWPE) material and are available in a variety of configurations containing one or more working USP #2 sutures or 1.5mm suture tapes to facilitate a repair.

{4}------------------------------------------------

The Stealth and Mini Stealth All-Suture Anchors are pre-loaded on disposable inserters and provided sterile via ethylene oxide (EO), while the reusable instruments are non-sterile and are intended to be sterilized by the end user.

### INDICATIONS FOR USE:

The Responsive Arthroscopy Stealth and Mini Stealth All-Suture Anchors are intended to be used for fixation of soft tissue to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Repair.

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair, Digital Tendon Transfers.

Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.

Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Carpal Ligament Reconstruction, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP Joints for all Digits, Digital Tendon Repair.

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair.

### TECHNOLOGICAL CHARACTERISTICS:

The subject Stealth and Mini Stealth All-Suture Anchors have intended use, indications for use, and fundamental scientific technology as the predicate devices cleared under K133671. Both the subject devices and the predicate devices feature similar technological characteristics, including a soft anchor design, principles of operation and insertion method, materials, repair suture offerings, packaging and shelf life, and sterilization method. In addition, both the subject devices and predicate devices are provided sterile and single use only pre-loaded on an inserter.

The subject devices feature slight differences in technology as compared to the predicate devices, including implant construct design, pre-deployment anchor dimensions, pilot hole dimensions, and implant deployment method. However, these technological characteristics are deemed equivalent to the predicate devices and have no impact on the ability of the subject devices to fulfill their intended use.

### SUBSTANTIAL EQUIVALENCE:

The subject Stealth and Mini Stealth All-Suture Anchors have intended use, indications for use and fundamental scientific technology as the predicate devices, while having similar technological characteristics. The differences in technology do not raise different questions of safety or efficacy. Therefore, the Stealth and Mini Stealth All-Suture Anchors are substantially equivalent to the predicate devices.

{5}------------------------------------------------

### PERFORMANCE TESTING:

Nonclinical performance testing was completed to demonstrate that the Stealth and Mini Stealth All-Suture Anchors met the established performance characteristics and design requirements. Performance testing consisted of design verification testing) that included side-by-side comparative testing with the predicate devices, shelf-life validation, extractable residue testing, and cytotoxicity testing. All testing met acceptance criteria and demonstrated that the devices met design specifications and performed as intended.

The following testing was performed on the subject devices:

- . Insertion Force Testing
- . Cyclic Pullout Force Testing
- Shelf-life Testing per ASTM F1980
- Extractable Residue Testing per ASTM F2459
- MEM Elution Testing per ISO 10993-5

In summary, performance testing of the Stealth and Mini Stealth All-Suture Anchors indicated no new risks and demonstrated substantial equivalence in performance compared to the legally marketed predicate devices.

### CONCLUSION:

In conclusion, the subject devices have the same indications for use, intended use, and fundamental scientific technology as the predicate devices. The differences in technological characteristics raise no new or different issues of safety and effectiveness, and performance testing has demonstrated that the subject devices are at least as safe and effective as the predicate devices. Therefore, the subject devices are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K230094](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K230094)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
