Mitek TIGHT-N Anchor

K221364 · Medos International SARL · MBI · Jul 11, 2022 · Orthopedic

Device Facts

Record IDK221364
Device NameMitek TIGHT-N Anchor
ApplicantMedos International SARL
Product CodeMBI · Orthopedic
Decision DateJul 11, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The MITEK TIGHT-N™ Tendon Docking Anchor is indicated for use for reattachment of soft tissue to bone for the following procedures: Shoulder: Biceps tenodesis Knee: Medial patellofemoral ligament repair/reconstruction (MPFL), posterior oblique ligament repair (POL), medial collateral ligament (MCL) repair, lateral ligament (LCL) repair, anterolateral ligament (ALL) reconstruction, Iliotibial (IT) Band Tenodesis

Device Story

The MITEK TIGHT-N™ Tendon Docking Anchor is a sterile, single-use implant designed for reattaching soft tissue to bone. It is provided preloaded on a disposable inserter shaft with a handle, secured by a stay suture. The device is available in 5.5 mm, 7.0 mm, and 8.5 mm sizes. During surgery, the surgeon threads the suture tails of a pre-stitched soft tissue graft through the anchor; the anchor is then inserted into the bone to secure the graft. Used in orthopedic surgical settings, the device provides mechanical fixation to facilitate tissue healing. It benefits patients by restoring soft tissue-to-bone attachment in ligament and tendon repair procedures.

Clinical Evidence

Bench testing only. Verification activities included biocompatibility, shelf-life, pyrogenicity, and endotoxin testing. Performance testing assessed fixation strength and fatigue strength. Sterilization validated per ANSI/AAMI/ISO 11135:2014; EO residuals tested per AAMI/ANSI/ISO 10993-7:2008; non-pyrogenicity confirmed per ANSI/AAMI ST-72:2011.

Technological Characteristics

Material: PEEK (Polyetheretherketone). Design: Suspensory push-in anchor. Sizes: 5.5 mm, 7.0 mm, 8.5 mm. Sterilization: Ethylene Oxide (EO) per ANSI/AAMI/ISO 11135:2014. Single-use. No software or electronic components.

Indications for Use

Indicated for reattachment of soft tissue to bone in shoulder (biceps tenodesis) and knee (MPFL, POL, MCL, LCL, ALL reconstruction, IT Band tenodesis) procedures.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 11, 2022 Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Medos International SARL % Ashley Aromando (Goncalo) Regulatory Affairs Project Manager DePuy Synthes Mitek Sports Medicine 325 Paramount Drive Raynham, Massachusetts 02767 Re: K221364 Trade/Device Name: MITEK TIGHT-N™ Tendon Docking Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 5, 2022 Received: May 12, 2022 Dear Ashley Aromando (Goncalo): We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221364 Device Name MITEK TIGHT-N™ Tendon Docking Anchor Indications for Use (Describe) The MITEK TIGHT-N™ Tendon Docking Anchor is indicated for use for reattachment of soft tissue to bone for the following procedures: · Shoulder: Biceps tenodesis · Knee: Medial patellofemoral ligament repair/reconstruction (MPFL), posterior oblique ligament repair (POL), medial collateral ligament (MCL) repair, lateral ligament (LCL) repair, anterolateral ligament (ALL) reconstruction, Iliotibial (IT) Band Tenodesis Type of Use (Select one or both, as applicable) | <div> <span> <span style="text-decoration: overline;">☑</span> Remediation Under PA STR 301-Subject Site </span> </div> | <div> <span> ☐ One-Time Correction (No STR 301-Subject Site) </span> </div> | |-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) SUMMARY ## MITEK TIGHT-N™ Tendon Docking Anchor ### Date Prepared: 07/08/2022 | Submitter's<br>Name and<br>Address | DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br><br>On behalf of:<br>Medos International SARL<br>Chemin-Blanc 38, Le Locle Neuchatel<br>CH 2400, Switzerland | | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------| | Contact Person | Tatyana Korsunsky<br>Technical Manager, Regulatory Affairs<br><br>DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 | Telephone: 774-284-7014<br>Email: tkorsuns@its.jnj.com | | Name of<br>Medical Device | Proprietary Name: MITEK TIGHT-NT™ Tendon Docking Anchor (also<br>referred to as TIGHT-NT™ Anchor)<br><br>Classification Name:<br>Fastener, Fixation, Nondegradable, Soft Tissue<br><br>Product Code: MBI<br>Common Name: Suture Anchor | | | Substantial<br>Equivalence | The MITEK TIGHT-NT™ Tendon Docking Anchor is substantially<br>equivalent to:<br>• K201749 Arthrex Biocomposite 3.9mm SwiveLock Anchor | | | Device<br>Classification | The MITEK TIGHT-NTM Tendon Docking Anchor is classified as:<br>Fastener, Fixation, Nondegradable, Soft Tissue, classified as Class II,<br>product code MBI, regulated under 21 CFR 888.3040. | | | Device Panel | Orthopedic Devices | | | Device<br>Description | The MITEK TIGHT-NTM Tendon Docking Anchor is designed to<br>reattach soft tissue to bone when used in conjunction with suture<br>(provided separately). The anchor is preloaded on a disposable inserter<br>shaft with handle, held in place by a non-absorbable "stay suture". The<br>proposed anchor will be offered in three sizes: 5.5 mm, 7.0 mm, and 8.5<br>mm and is molded from PEEK (Polyetheretherketone) material. The<br>proposed anchor is provided sterile via Ethylene Oxide (EO)<br>sterilization and is for single use only. | | | Technological<br>Characteristics | The anchor design, principal of operation and intended use on MITEK<br>TIGHT-NTM Tendon Docking Anchor are similar when compared to the<br>predicate Arthrex 3.9mm SwiveLock Anchor (K201749). Both devices<br>are implantable anchors used to secure soft tissue to bone. Suture tails<br>of a pre-stitched soft tissue graft (suture provided separately) are<br>threaded through the anchors, thus securing the graft to the anchor.<br><br>The proposed MITEK TIGHT-NTM Tendon Docking Anchor is<br>designed as a suspensory push-in anchor, while the predicate<br>SwiveLock is designed as an interference thread-in anchor. While there<br>exists differences in the means of insertion style (thread-in vs. push in)<br>and mechanism of tissue retention, the proposed and predicate devices<br>are ultimately securing soft tissue to bone to achieve the device's<br>therapeutic benefit.<br><br>Both predicate and proposed devices are provided with sutures<br>facilitating placement of the anchor (stay sutures for both the proposed<br>and the predicate device and an additional axillary suture for the<br>proposed MITEK TIGHT-NTM Tendon Docking Anchor).<br><br>The proposed anchor is constructed from PEEK material, the predicate<br>Arthrex SwiveLock is composed of both PEEK and PLLA/βTCP<br>materials. The PEEK material, of the proposed anchor implant, is<br>identical to the PEEK material of existing Mitek implantable anchors –<br>Versalok Anchors (K100532)<br><br>Any differences between the proposed device and the predicate are<br>considered non-significant and do not raise different questions of safety<br>or effectiveness, supported by testing. | | | Indications for<br>Use | The MITEK TIGHT-NTM Tendon Docking Anchor is indicated for use<br>for reattachment of soft tissue to bone for the following procedures:<br>Shoulder: Biceps tenodesis Knee: Medial patellofemoral ligament repair/reconstruction<br>(MPFL), posterior oblique ligament repair (POL), medial collateral<br>ligament (MCL) repair, lateral collateral ligament (LCL) repair,<br>anterolateral ligament (ALL) reconstruction, Iliotibial (IT) Band<br>Tenodesis | | | Non-clinical<br>Testing | Verification activities were performed on the proposed device and / or<br>its predicates. Biocompatibility, shelf-life, pyrogenicity, endotoxin and<br>performance testing conducted on representative products was presented<br>to support determination of equivalence. Performance testing included<br>assessment of fixation strength and fatigue strength. | | | | Ethylene Oxide Sterilization was validated according to<br>ANSI/AAMI/ISO 11135: 2014 to a SAL of 1 x 10-6. | | | | EO residuals were tested per AAMI/ANSI/ISO 10993-7:2008 | | | | The proposed device has been determined to be non-pyrogenic per the<br>requirements set forth in ANSI/AAMI ST-72:2011, United States<br>Pharmacopeia (USP), and European Pharmacopeia (EP) using the<br>bacterial endotoxin testing (BET) method. | | | Safety and<br>Performance | Results of performance testing have demonstrated that the proposed<br>devices are suitable for their intended use. | | | | Based on similarities in the intended use, technological characteristics,<br>and performance in comparison to the predicate devices, the proposed<br>MITEK TIGHT-NTM Tendon Docking Anchor has shown to be<br>substantially equivalent to the predicate device under the Federal Food,<br>Drug and Cosmetic Act. | | {4}------------------------------------------------ {5}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...