Auxilock Titanium Screw-In Suture Anchor

K213104 · Auxein Medical Private Limited · MBI · Jul 29, 2022 · Orthopedic

Device Facts

Record IDK213104
Device NameAuxilock Titanium Screw-In Suture Anchor
ApplicantAuxein Medical Private Limited
Product CodeMBI · Orthopedic
Decision DateJul 29, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Auxilock Titanium Screw-In Suture Anchor is intended to be used with #2 Biobraid suture with white, blue, black colors for use for the reattachment of soft tissue to bone for the following indications: Shoulder: Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsutolabral reconstructions, Biceps tenodesis, and Deltoid Repairs, Anterior Shoulder Instability Repair. Foot and Ankle: Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy. Elbow, Wrist and Hand: Scapholunate ligament reconstruction, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment, Tennis elbow repair. Knee: Extra-capsular repairs: Medial collateral ligament repair. Lateral ligament repair. Posterior oblique ligament. Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement. Hip: Capsular repair, Acetabular labral repair.

Device Story

Threaded, tapered titanium alloy fastener; used for reattaching soft tissue to bone; self-tapping design; available in 2.8mm, 3.5mm, 5.0mm, and 6.5mm diameters; incorporates cancellous and cortical threads for pull-out strength. Used in orthopedic surgeries (shoulder, knee, hip, foot/ankle, hand/wrist); operated by surgeons; provided preassembled on an inserter with #2 Biobraid suture. Drill guide and bit included for accurate placement; minimizes anchor slippage/breakage. Device provides mechanical fixation to facilitate healing; sold sterile for single use.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included insertion tests, pull-out tensile static tests, and pull-out tensile fatigue tests compared against predicate devices. Biocompatibility evaluated per ISO 10993-3.

Technological Characteristics

Material: Titanium alloy (Ti-6Al-4V, ASTM F136); Suture: UHMWPE (ASTM F2848-17). Threaded, tapered fastener design. Sterilization: Ethylene Oxide (ISO 11135). Single-use, sterile. No software or electronic components.

Indications for Use

Indicated for adult patients requiring soft tissue-to-bone reattachment in shoulder, foot, ankle, elbow, wrist, hand, knee, and hip procedures, including ligament/tendon repairs and reconstructions.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 29, 2022 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized eagle. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Auxein Medical Private Limited Rahul Luthra Director Plot No. 168, 169, 170 Phase-IV, Sector 57, Kundli Industrial Area Sonipat, Haryana 131028 India ## Re: K213104 Trade/Device Name: Auxilock Titanium Screw-In Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: June 22, 2022 Received: June 27, 2022 ## Dear Rahul Luthra: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For. Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known) K213104 Device Name Auxilock Titanium Screw-In Suture Anchor #### Indications for Use (Describe) Auxilock Titanium Screw-In Suture Anchor: The Auxilock Titanium Screw-In Suture Anchor is intended to be used with #2 Biobraid suture with white, blue, black colors for use for the reattachment of soft tissue to bone for the following indications: Shoulder: Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsutolabral reconstructions, Biceps tenodesis, and Deltoid Repairs, Anterior Shoulder Instability Repair. Foot and Ankle: Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy. Elbow, Wrist and Hand: Scapholunate ligament reconstruction, Unar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment, Tennis elbow repair. #### Knee: Extra-capsular repairs: Medial collateral ligament repair. Lateral ligament repair. Posterior oblique ligament. Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement. #### Hip: Capsular repair, Acetabular labral repair. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. A registered trademark symbol is located to the right of the word. ## 510(k) Summary # General Company Information as required by 807.92 (a) | Submitter's Name: | Auxein Medical Private Limited | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Address: | Auxein Medical Private Limited<br>Plot No. 168-169-170, Phase-4, Kundli Industrial Area, HSIIDC,<br>Sector-57, Sonepat-131028, Haryana, India | | Contact Person Name: | Mr. Rahul Luthra | | Title: | Director | | Phone Number: | +91 9560557733 | | Dated: | 29.07.2022 | #### Person Responsible for Regulatory Compliance | Name: | Mr. Mohit Kumar | |-----------|-----------------------| | Title: | Sr. Research Engineer | | Email Id: | m.kumar@auxein.com | | Dated: | 29.07.2022 | # Proprietary Name: Auxilock Titanium Screw-In Suture Anchor #### Common or Usual Name: Soft Tissue Fixation Device # Classification Name: Fastener, fixation, non-degradable, soft tissue ## Product Code: MBI Device Class: II Review Panel: Orthopedic # Regulation Number: 21 CFR 888.3040 | Primary Product<br>Code | Classification Name | Common Name | Regulation Number | |-------------------------|-----------------------------------------------------|--------------------------------|-------------------| | MBI | Fastener, fixation, non-<br>degradable, soft tissue | Soft Tissue Fixation<br>Device | 21 CFR 888.3040 | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. A small "R" in a circle is located to the upper right of the word "Auxein". Following are the predicate device 510(k) with which we are declaring substantial equivalence: The following is the range of variants covered with their corresponding predicate devices. #### Primary Predicate: | 510(k) Number | K152566 | |---------------|--------------------------------| | Applicant | Smith & Nephew | | Common Name | Suture Anchor | | Device Name | PEBA Anchor/Suture Combination | #### Secondary Predicate: | 510(k) Number | K053344 | |---------------|---------------------------------------------------------| | Applicant | Smith & Nephew | | Common Name | Suture Anchor | | Device Name | TwinFix 2.8mm, 3.5mm and BioRaptor 2.9mm Suture Anchors | ## Device Description: The Titanium Screw-In Suture Anchor is a threaded, tapered fastener for use in attachment of soft tissue to bone. The self-tapping AUXILOCK® Titanium screw-in suture anchor is mostly used as a medial row anchor in rotator cuff repair surgery, Bankart or SLAP surgeries. It is recommended for use in small and large-joint repairs without the need for tapping or drilling. AUXILOCK® Titanium screw-in anchors are available in diameter of 2.8mm, 3.5mm, 5.0mm and 6.5mm. The anchor design incorporates a cancellous thread with a tiny core diameter to maximize pull-out strength in cancellous or osteoporotic bone. The AUXILOCK® Titanium screw-in anchor is designed for ultimate mechanical properties (pull-out strength, tensile strength, etc.) and ease-of-use. The anchor design incorporates a cortical thread with a small core diameter to maximize pull-out strength in a cortical or hard bone. The drill guide and drill bit are provided in the instrument set for the accurate placement of the anchor. It in turn minimizes the anchor slippage or breakage during the surgery. All of the Auxein's Non-Absorbable Suture Anchors are provided sterile, for single use only. All Non-Absorbable Suture Anchors are preassembled onto an inserter, sized appropriately to accommodate the indicated procedures. Auxein's Non-Absorbable Suture Anchors come in various configurations, including: with attached non-absorbable suture(s). In certain configurations, the Non-Absorbable Suture Anchors are packaged with a guide and awl. All configurations of the Non-Absorbable Suture Anchors are identical to the identified predicate devices. The device is made from a Titanium alloy, Ti-6Al-4V (ASTM F136) as per grade 5. Note: The #2 Biobraid Suture is not to be used as a stand-alone. It should be used only with the assembled implant. These implants are sold in sterile conditions (Ethylene Oxide Sterilization). The system is indicated for use in adult patients only. All implants are for single use only. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability. #### Indications for Use: #### Auxilock Titanium Screw-In Suture Anchor: The Auxilock Titanium Screw-In Suture Anchor is intended to be used with #2 Biobraid suture with white, blue, black colors for use for the reattachment of soft tissue to bone for the following indications: #### Shoulder Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsutolabral reconstructions, Biceps tenodesis, and Deltoid Repairs, Anterior Shoulder Instability Repair. #### Foot and Ankle Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy. ## Elbow, Wrist and Hand Scapholunate ligament reconstruction, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment, Tennis elbow repair. #### Knee Extra-capsular repairs: Medial collateral ligament repair, Lateral ligament repair, Posterior oblique ligament, Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement. #### Hip Capsular repair, Acetabular labral repair. ## Summary of Technological Characteristics as compared to the predicate devices A comparison between the Auxilock Titanium Screw-In Suture Anchor and predicate devices has been performed which has resulted in demonstration of similarities in dimensional and performance criteria. Following is the summary of parameters in which the comparison has been verified: | S.No. | Characterist<br>ics | Subject Device (Auxein) | Predicate Devices | Remarks | |-------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------| | 1. | Product Code | MBI | K152566, PEBA<br>Anchor/Suture Combination,<br>(Primary Predicate)<br><br>K053344, TwinFix 2.8mm,<br>3.5mm Suture Anchors<br>(Secondary Predicate) | Identical | | | | | | as<br>predicate<br>device. | | 2. | Regulation<br>Number | 21 CFR 888.3040 | 21 CFR 888.3040 | Identical<br>as<br>predicate<br>device. | | 3. | Regulatory<br>Class | Class II | Class II | Identical<br>as<br>predicate<br>device. | | 4. | Indications<br>for use | <b>Auxilock Titanium<br/>Screw-In Suture Anchor</b><br>The Auxilock Titanium Screw-In Suture Anchor is<br>intended to be used with #2 Biobraid suture with white,<br>blue, black colors for use for the reattachment of soft<br>tissue to bone for the following indications:<br><b>Shoulder:</b><br>Bankart lesion repairs, SLAP lesion repairs,<br>Acromiolavicular separation repairs, Rotator<br>cuff tear repairs, Capsular shift or Capsutolabral<br>reconstructions, Biceps tenodesis, and Deltoid<br>Repairs, Anterior Shoulder Instability Repair.<br><b>Foot and Ankle:</b><br>Hallux valgus reconstruction, Medial or<br>Lateral instability repairs/reconstructions,<br>Achilles tendon repair/reconstructions,<br>Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions | The PeBA Series<br>Anchor/Suture Combination<br>(Primary Predicate) is intended<br>for the fixation of surgical<br>suture material for the<br>following indications:<br><b>Shoulder:</b><br>Bankart lesion repairs, SLAP<br>lesion repairs, Acromio-<br>clavicular separation repairs,<br>Rotator cuff tear repairs,<br>Capsular shift or capsulolabral<br>reconstructions, Biceps<br>tenodesis, Deltoid repairs.<br><b>Foot and Ankle:</b><br>Hallux Valgus repairs, Medial<br>or lateral instability repairs,<br>Achilles tendon<br>repairs/reconstructions, Midfoot<br>reconstructions, Metatarsal<br>ligament/tendon<br>repairs/reconstructions<br><b>Elbow, Wrist, and Hand:</b><br>Scapholunate ligament<br>reconstructions, Ulnar or radial<br>collateral ligament<br>reconstructions, Tennis elbow<br>repair, Biceps tendon<br>reattachment. | Identical<br>as<br>predicate<br>devices. | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image features the logo for Auxein. The logo consists of a stylized letter "A" inside of a circle, followed by the word "Auxein" in a sans-serif font. The logo and text are both in a light blue color. {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. A small "R" in a circle is located to the right of the word "Auxein" indicating that the logo is a registered trademark. | | Bunionectomy. | <b>Knee:</b><br>Extra-capsular repairs, medial<br>collateral ligament, lateral<br>collateral ligament, posterior<br>oblique ligament, Iliotibial band<br>tenodesis, Patellar realignment<br>and tendon, repairs, including<br>vastus medialis, obliquous<br>advancement. | |--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <b>Elbow, Wrist and Hand:</b><br>Scapholunate ligament<br>reconstruction, Ulnar or<br>radial collateral ligament<br>reconstructions, Lateral<br>epicondylitis repair, Biceps<br>tendon reattachment,<br>Tennis elbow repair. | The K053344, TwinFix 2.8mm,<br>3.5mm Suture Anchors<br>(Secondary Predicate) is<br>intended for use for the<br>reattachment of soft tissue to<br>bone for the following<br>indications: | | | <b>Knee:</b><br>Extra-capsular repairs:<br>Medial collateral ligament<br>repair, Lateral collateral<br>ligament repair, Posterior<br>oblique ligament, Iliotibial<br>band tenodesis, Patellar<br>tendon repair - Vastus<br>medialis obliquous<br>advancement. | <b>Shoulder</b><br>Bankart lesion repairs, SLAP<br>lesion repairs,<br>Acromiolavicular separation<br>repairs, Rotator cuff tear<br>repairs, Capsular shift or<br>Capsutolabral reconstructions,<br>Biceps tenodesis, and Deltoid<br>Repairs, Anterior Shoulder<br>Instability Repair. | | | <b>Hip:</b><br>Capsular repair, Acetabular<br>labral repair. | <b>Foot and Ankle</b><br>Hallux valgus reconstruction,<br>Medial or Lateral instability<br>repairs/reconstructions,<br>Achilles tendon<br>repair/reconstructions, Mid-foot<br>reconstruction, Metatarsal<br>ligament/tendon<br>repairs/reconstructions,<br>Bunionectomy. | | | | <b>Elbow, Wrist and Hand</b><br>Scapholunate ligament<br>reconstruction, Ulnar or radial<br>collateral ligament | {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a blue triangle inside of it, followed by the word "Auxein" in blue font. There is a registered trademark symbol to the right of the word. | | | | reconstructions, Lateral<br>epicondylitis repair, Biceps<br>tendon reattachment, Tennis<br>elbow repair. | | |-----|-----------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | | | | Knee<br>Extra-capsular repairs: Medial<br>collateral ligament repair,<br>Lateral collateral ligament<br>repair, Posterior oblique<br>ligament, Iliotibial band<br>tenodesis, Patellar tendon repair<br>- Vastus medialis obliquous<br>advancement. | | | | | | Hip<br>Capsular repair, Acetabular<br>labral repair. | | | 5. | Material | Titanium Alloy as per<br>ASTM F136 and<br>UHMWPE Suture as per<br>ASTM F2848-17. | Titanium Alloy as per ASTM<br>F136 and and UHMWPE<br>Suture as per ASTM F2848-17. | Conforms<br>to Identical<br>standard as<br>predicate<br>device. | | 6. | Performance<br>Standards | Mechanical Testing<br>Protocol AMPL-MT-PRT-<br>002, Rev. 01. was used in<br>the scoped Device. | Mechanical Testing Protocol<br>AMPL-MT-PRT-002, Rev. 01.<br>was used in the predicate<br>Device. | Identical<br>as<br>predicate<br>device. | | 7. | Sterilization | Ethylene Oxide<br>Sterilization Method was<br>used in the subject device. | Ethylene Oxide Sterilization<br>Method was used in the<br>predicate devices. | Identical in<br>predicate<br>device. | | 8. | Dimensional<br>Verification | Same dimension as of Predicate device. | | Identical<br>as<br>predicate<br>device. | | 9. | Shelf-life | 5 Years | 5 Years | Identical<br>as<br>predicate<br>device. | | 10. | Single<br>Use/Reuse | Single Use | Single Use | Identical<br>as<br>predicate<br>device. | | 11. | Operating Principle | Secure tendon and ligament reconstructions with increased boney ingrowth at the anchor site with minimal invasive technique. | Secure tendon and ligament reconstructions with increased boney ingrowth at the anchor site with minimal invasive technique. | Identical as predicate device. | {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, with a registered trademark symbol next to it. The logo is simple and modern, and the blue color gives it a sense of trust and reliability. ## Discussion on the non-clinical testing performed Following are the applicable product standards considered for non-clinical standards: - Material Standards. ● - . Biocompatibility Standards - Performance Standards. - Sterilization, shelf-life and packaging for sterile product. - . Bacterial Endotoxin #### Non-Clinical Test Summary: Bench tests were conducted to verify that the subject device met all design specifications. The test results demonstrated that the subject device complies with the following standards: #### Material Standards: The material standards are the essential part to be complied with first, as it is the basis of manufacturing metallic surgical implants. We have complied with the following material standards: - ASTM F136/ISO 5832-3: Standard specification for wrought Titanium-6 Vanadium . ELI (Extra low interstitial) Alloy for surgical implant applications. - · ASTM F2848-17: Standard Specification for Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns. - ASTM F899-20: Standard Specification for Wrought Stainless Steels for surgical instruments. Note: We have used Grade 304 of Stainless steel (SS 304) Material for instruments as per ASTM F899-20, UHMWPE for Suture as per ASTM F2848-17 and Titanium Alloy (Ti- 6Al-4V) Grade 5 for Titanium Implants. #### Summary of Biocompatibility The device in its final finished form has been evaluated for biocompatibility according to ISO 10993-3. #### Conclusion of Mechanical performance The following are the mechanical tests that have been performed on the Subject device (i.e. Auxilock Titanium Screw-In Suture Anchor) and Predicate device (i.e. Smith & Nephew's Twinfix Suture Anchor): - 1. Insertion Test - 2. Pull-Out Tensile Static Test and Pull-Out Tensile Fatigue Test. {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a white "A" inside of it, followed by the word "Auxein" in blue. There is a registered trademark symbol next to the "n" in "Auxein". # Sterilization, shelf-life and packaging for sterile product ## Sterilization: ETO Sterilization The ETO sterilization has been performed to sterilize this medical device. EO penetrates the packaging, making contact with all accessible surfaces of the product to deliver the required sterility assurance level (SAL). Trace levels of EO and ethylene chlorohydrin (ECH) may remain on products after an EO sterilization process. To detect these traces, EO residual test was done. ISO 10993-7 outlines the specific limits of EO and ECH that must not be exceeded in order to ensure product and patient safety. - ISO 11135: 2014, Sterilization of health-care products Ethylene oxide Requirements for the ● development, validation and routine control of a sterilization process for medical devices. - ISO 17665-1:2006, Sterilization of health care products Moist heat Part 1: Requirements for ● the development, validation and routine control of a sterilization process for medical devices. - ISO/TS 17665-2:2009, Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1. - . ISO/TS 17665-3:2013 (en), Sterilization of health care products - Moist heat - Part 3: Guidance on the designation of a medical device to a product family and processing category for steam sterilization. - . ISO 11140-1:2014, Sterilization of health care products — Chemical indicators — Part 1: General requirements. - . ISO 11737-1:2018 Sterilization of medical devices - Microbiological methods- Part 1: Estimation of population of microorganisms on products. - ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods- Part 2: Tests of sterility performed in the validation of a sterilization process. ## Packaging of the Product: Tyvek Packaging The integrity of the final package is maintained at least for the claimed shelf-life of the medical device. The tyvek pouch is used for the packaging of Auxilock Titanium Screw-In Suture Anchor. The double sterile barrier is used for the packaging. We have followed the below standards for packaging of the device: - · ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system. - . ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes. ## Shelf-life: 5 years The stability study has been done to determine the shelf life. We have followed the below standards for performing shelf-life of the device: - ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. - ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials. - ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging ● {11}------------------------------------------------ Image /page/11/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a white triangle inside, followed by the word "Auxein" in blue, stylized font. There is a small "R" in a circle to the right of the word "Auxein". by Dye Penetration. #### Safety in MRI Not Evaluated The Auxilock Titanium Screw-In Suture Anchor has not been evaluated for safety in the MR environment. It has not been tested for heating or unwanted movement in the MR environment. The safety of Auxilock Titanium Screw-In Suture Anchor in the MR environment is unknown. Performing an MR exam on a person who has this medical device may result in injury or device malfunction. #### Bacterial Endotoxin - USP <85> Bacterial Endotoxin Test. . - USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests. Bacterial Endotoxin Test (BET): We perform Bacterial Endotoxin Test of every sterile batch products by using acceptance criteria as endotoxin testing limit must be <20EU/Device (as the products are general medical devices for implantation). The testing is done by using Limulus amebocyte lysate (LAL) test. The testing procedure is performed as per the standards USP 85 and ANSI/AAMI ST72:2019. ## Conclusion: There are no significant differences between the subject device and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, and operational principles as internal fixation components. From the data available we can justify that the Auxilock Titanium Screw-In Suture Anchor is as safe, and as effective and performs the same indications for use as that of already marketed predicate devices identified in 510(k) summary.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...