K202883 · Osteonic Co., Ltd. · MBI · Nov 25, 2020 · Orthopedic
Device Facts
Record ID
K202883
Device Name
Sterile bone screw (PEEK ACL screw)
Applicant
Osteonic Co., Ltd.
Product Code
MBI · Orthopedic
Decision Date
Nov 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Sterile bone screw (PEEK ACL screw) is a fixation screw that fixes soft tissue such as ligaments, tendons, and the articular capsules to bone, and is used in orthopedic surgery. Knee • ACL repairs • PCL repairs • Extra-capsular repairs – Medial collateral ligament – Lateral collateral ligament – Posterior oblique ligament • Patellar realignment and tendon repairs – Vastus medialis obliquus advancement • Iliotibial band tenodesis
Device Story
Sterile bone screw (PEEK ACL screw) is a fixation device used in orthopedic surgery to secure soft tissues (ligaments, tendons, articular capsules) to bone. Device is manufactured from PEEK (poly-ether-ether-ketone) and supplied sterile via gamma irradiation in Tyvek pouches. It is intended for single-use. During surgery, the screw is inserted into bone to provide mechanical fixation of soft tissue. Clinical benefit is achieved through stable reattachment of tissues to facilitate healing. Device is operated by orthopedic surgeons in clinical settings.
Clinical Evidence
Bench testing only. Performance verified via ASTM F543-A3 (axial pullout), ASTM F543-A1 (torsional yield), ASTM F543-A2 (driving torque), and fatigue testing (70-220N tensile load, 1000 cycles at 1 Hz). Biocompatibility evaluated per ISO 10993. All tests passed.
Technological Characteristics
Material: PEEK (ASTM F2026). Dimensions: Diameters 7.0-10.0 mm, lengths 25-30 mm. Sterilization: Gamma irradiation. Single-use. Mechanical performance tested per ASTM F543. Packaging validated per ISO 11607 and ASTM F1980/F88/F1929.
Indications for Use
Indicated for fixation of soft tissue (ligaments, tendons, articular capsules) to bone in orthopedic surgery, specifically for knee procedures including ACL/PCL repairs, extra-capsular repairs (MCL, LCL, posterior oblique ligament), patellar realignment, tendon repairs (vastus medialis obliquus advancement), and iliotibial band tenodesis.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
November 25, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
Osteonic Co., Ltd. % Sanglok Lee Manager WISE COMPANY Inc. #507-508, 166 Gasan digital 2-ro Geumcheon-gu, Seoul, Republic of Korea 08503
#### Re: K202883
Trade/Device Name: Sterile bone screw (PEEK ACL screw) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: September 24, 2020 Received: September 28, 2020
#### Dear Sanglok Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
{3}------------------------------------------------
1206Ho. 38. Digital-ro 29 Tel: +82(2)-6902-8400 Fax: +82(2)-6902-8401 Home page: http://www.os
83 Page 1 of 4
## The assigned 510(k) Number: K202883
## 1. 510(k) Summarv Date: 11. 23. 2020
## 2. Applicant
### Sung-moo Hwang
OSTEONIC Co., Ltd. 303Ho, 405Ho, 505-3Ho, 505-3Ho, 902Ho, 1004Ho, 1201Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea, 102Ho, 103Ho, 24, Digital-ro 27-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8426 FAX: +82-2-6902-8401 Email: sungmhwang@osteonic.com
## 3. Submission Correspondent
Sang-lok, Lee Wise COMPANY Inc. #303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
# 4. Subject Device Identification
- 1) Trade Name: Sterile bone screw (PEEK ACL screw)
- 2) Common Name: Sterile bone screw (PEEK ACL screw)
- 3) Classification Name: fastener, fixation, non-biodegradable, soft tissue
- 4) Product Code: MBI
- 5) Panel: Orthopedic
- 6) Regulation Number: 21 CFR 888.3040
- 7) Device Class: II
## 5. Indication for use
The Sterile bone screw (PEEK ACL screw) is a fixation screw that fixes soft tissue such as ligaments, tendons, and the articular capsules to bone, and is used in orthopedic surgery.
Knee
- · ACL repairs
- · PCL repairs
- Extra-capsular repairs
- Medial collateral ligament
- Lateral collateral ligament
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Osteonic logo. The logo consists of a blue diamond shape with a white line running through it. To the right of the diamond is the word "OSTEONIC" in a dark gray sans-serif font. The logo is simple and modern.
K202883 Page 2 of 4
- Posterior oblique ligament
- Patellar realignment and tendon repairs - Vastus medialis obliquus advancement
- Iliotibial band tenodesis
## 6. Predicate devices
- 1) 510(k) Number: K083635
- 2) Device Name: Biosure PK Interference Screw
- 3) Manufacturer: Smith&Nephew PK Interference Screw
## 7. Device Description
The Sterile bone screw (PEEK ACL screw) is a bone fixation screw that fixes soft tissues such as ligament, tendon, and the articular capsules to bone, and is used in orthopedic surgery.
The implanted screw is made of PEEK (poly-ether-ether-ketone, ASTM F2026).
The Sterile bone screw (PEEK ACL screw) is supplied gamma irradiation sterile state and it is packed in Tyvek Pouch.
## 8. Non-Clinical Test Conclusion
#### 1) General Information
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
#### ▪ Material
- ASTM F2026:2017, Standard Specification for Poly-ether-ether-ketone(PEEK)Polymer for Surgical Implant Applications
#### · Mechanical Performance
- ASTM F543:17 Standard Specification and test method for metallic bone screws
- = Sterilization, Shelf-life and Packaging for Sterile Product
- ISO 11138-1:2006, Sterilization of health care products Biological indicators -- Part 1: General requirements
- ISO 11140-1:2014, Sterilization of health care products Chemical indicators Part 1: General requirements
- ISO 11137-1:2006. Sterilization of health care products Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11137-2:2013, Sterilization of health care products Radiation Part 2: Establishing the sterilization dose
- ISO 11137-3:2017, Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation and routine control
- ISO 11737-1:2018 Sterilization of medical devices Microbiological methods- Part 1: Estimation of population of microorganisms on products
- ISO 11737-2:2009 Sterilization of medical devices Microbiological methods- Part 2: Tests of
{5}------------------------------------------------
sterility performed in the validation of a sterilization process
- ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system
- ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980:2016 Standard guide for accelerated aging of sterile barrier systems for medical devices
- ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials.
- ASTM F1929:2015 Standard test method for detecting seal leaks in porous medical packaging by dye penetration
- Bacterial Endotoxin
- USP <85> Bacterial Endotoxin Test
- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests
#### 2) Summary of Biocompatibility
The Sterile bone screw (PEEK ACL screw) has been evaluated for biocompatibility according to ISO 10993.
#### 3) Summary of Performance
All specimens selected (worst-case) to verify the performance of the screw met the test criteria. Therefore, the performance of the entire model was proven.
| No. | Test Article | Test Method | Result |
|-----|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1 | Axial Pullout Strength | ASTM F543-A3.Test Method for Determining the Axial<br>Pullout Strength of Medical Bone Screws. | PASS |
| 2 | Torsional Yield Strength | ASTM F543-A1.Test Method for Determining the<br>Torsional Properties of Metallic Bone Screws. | PASS |
| 3 | Driving Torque<br>Insertion<br>Removal | ASTM F543-A2.Test Method for Driving Torque of Medical<br>Bone Screws. | PASS |
| 4 | Fatigue | After fully inserting the screw with the tendon into the fixed<br>artificial bone (20PCF), a fatigue tensile load of 70 to 220N<br>on the tendon is repeated 1000 cycles at 1 Hz. | PASS |
## 9. Substantially Equivalent Conclusion
#### Table 1: Substantial Equivalence Comparison
| Product<br>Name | SUBJECT Device | PREDICATE Device<br>Biosure PK Interference Screw (K083635) | Equivalence<br>Discussion |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Product<br>code | MBI | MBI | Same |
| Regulatory<br>class | Class II | Class II | Same |
| Regulation<br>Number | 21 CFR 888.3040 | 21 CFR 888.3040 | Same |
| Intended<br>Use | The Sterile bone screw (PEEK<br>ACL screw) is a fixation screw<br>that fixes soft tissue such as<br>ligaments, tendons, and the<br>articular capsules to bone, and is<br>used in orthopedic surgery | The BIOSURE PK Screw is indicated for<br>the reattachment of ligament, tendon, soft<br>tissue, or bone to bone during cruciate<br>ligament reconstruction surgeries of the<br>knee. All screws with a diameter of 9 mm<br>or less and a length of 25 mm or less are also | Same |
| | Knee<br>• ACL repairs<br>• PCL repairs<br>• Extra-capsular repairs<br>– Medial collateral ligament<br>– Lateral collateral ligament<br>– Posterior oblique ligament<br>• Patellar realignment and tendon<br>repairs<br>– Vastus medialis obliquus<br>advancement<br>• Iliotibial band tenodesis | intended for use in the following<br>procedures:<br>Knee<br>• ACL repairs<br>• PCL repairs<br>• Extra-capsular repairs<br>– Medial collateral ligament<br>– Lateral collateral ligament<br>– Posterior oblique ligament<br>• Patellar realignment and tendon repairs<br>– Vastus medialis obliquus advancement<br>• Iliotibial band tenodesis<br>Shoulder<br>• Acromioclavicular separation repairs<br>• Biceps tenodesis<br>Foot and Ankle<br>• Medial or lateral instability repairs/<br>reconstructions<br>• Achilles tendon repairs/reconstructions<br>• Metatarsal ligament/tendon repairs/<br>reconstructions<br>• Flexor hullucis longus (FHL)<br>• Tendon transfers<br>Elbow, Wrist, and Hand<br>• Biceps tendon reattachment<br>• Ulnar or radial collateral ligament<br>reconstructions<br>• Lateral epicondylitis repair<br>• Scapholunate ligament reconstruction<br>• Tendon transfers<br>• Carpomedicarpal joint arthroplasty | |
| Operating<br>Principles | The Sterile bone screw (PEEK<br>ACL screw) is a fixation screw<br>that fixes soft tissue such as<br>ligaments, tendons, and the<br>articular capsules to bone, and is<br>used in orthopedic surgery. | The Biosure PK Interference Screw is a<br>fixation screw that fixes soft tissue such as<br>ligaments, tendons, and the articular<br>capsules to bone, and is used in orthopedic<br>surgery. | Same |
| Material | PEEK 100% (Optima PEEK) | PEEK 100% (Optima PEEK) | Same |
| Product<br>Size | Ø 7.0 x (25,30) mm<br>Ø 8.0 x (25,30) mm<br>Ø 9.0 x (25,30) mm<br>Ø 10.0 x (25,30) mm | Ø 6.0 x (20,25,25(R)) mm<br>Ø 7.0 x (20,25,25(R),30) mm<br>Ø 8.0 x (20,25,25(R),30,35) mm<br>Ø 9.0 x (20,25,30,35) mm<br>Ø 10.0 x (20,25,30,35) mm<br>Ø 11.0 x (25,30,35) mm<br>Ø 12.0 x 35 mm | Similar |
| Sterilization | Sterile (Gamma sterilization) | Sterile (Gamma sterilization) | Same |
| Single Use/<br>Reuse | Single use | Single use | Same |
| Packaging | 1 EA / BOX | 1 EA / BOX | Same |
| Shelf life | 5 Years | 5 Years | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Osteonic logo. The logo consists of a blue geometric shape on the left and the word "OSTEONIC" in gray on the right. The geometric shape is a diamond with a stylized "O" inside.
K202883 Page 4 of 4
Based on above, the subject device, Sterile bone screw (PEEK ACL screw), is determined to be Substantially Equivalent (SE) to the predicate devices, Biosure PK Interference Screw (K083635) in respect of safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.