← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K202259

# Parcus GFS II and GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip (K202259)

_Parcus Medical, LLC · MBI · Oct 28, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K202259

## Device Facts

- **Applicant:** Parcus Medical, LLC
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Oct 28, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Parcus GFS II and GFS Mini are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. The Parcus GFS Naked devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. The GFS Ultimate devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. The GFS BTB devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. The Parcus ATLAS is indicated for adjunct fixation of the 1st and 2nd metacarpals in CMC arthroplasty. The Parcus Actiflip Naked, Actiflip Cinch, and Actiflip Whip are used for fixation of bone to bone or soft tissue to bone, and is intended as fixation post, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in the knee, shoulder, and elbow and may include the following indications: anterior cruciate ligament, posterior cruciate ligament, pectoralis repair (minor/major), biceps tendon repair and reattachment (distal/proximal), acromioclavicular repair, and ulnar collateral ligament reconstruction. The Parcus GFS Ultimate Hip is used for fixation of soft tissue to bone in the hip and is indicated for ligamentum teres reconstruction. The Parcus Synd-EZ Ti are intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. The Parcus Synd-EZ Ti are intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis (syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

## Device Story

Range of metallic bone fixation fasteners; used for ligament/tendon repair, CMC arthroplasty, and fracture fixation. Devices consist of titanium alloy (Ti-6AI4V ELI) implants, potentially with UHMWPE polyblend suture components. Operated by surgeons in clinical settings (OR). Function as fixation posts, distribution bridges, or adjuncts to stabilize bone/soft tissue during healing. Output is physical stabilization of tissue/bone. Benefit is secure fixation during healing process. Subject devices are identical to previously cleared predicates; update adds MR Conditional parameters to labeling.

## Clinical Evidence

Bench testing only. Evaluation performed for MR environment compatibility (magnetically induced force, torque, RF heating, image artifact) per FDA guidance and ASTM standards. No clinical data required.

## Technological Characteristics

Materials: Medical grade titanium alloy (Ti-6AI4V ELI) and UHMWPE polyblend suture. Form factor: Bone fixation fasteners/implants. Sterilization: Provided sterile. MR status: MR Conditional.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Parcus GFS II and GFS Mini ([K133757](/device/K133757.md))
- Parcus GFS Naked ([K152711](/device/K152711.md))
- Parcus GFS Ultimate ([K162198](/device/K162198.md))
- Parcus GFS BTB ([K183331](/device/K183331.md))
- Parcus ATLAS ([K190375](/device/K190375.md))
- Parcus Synd-EZ Ti ([K191783](/device/K191783.md))
- Parcus Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip ([K192750](/device/K192750.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

October 28, 2020

Parcus Medical, LLC Paul Vagts Director of Regulatory Affairs 6423 Parkland Drive Sarasota, Florida 34243

Re: K202259

Trade/Device Name: Parcus GFS II and GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: July 31, 2020 Received: August 10, 2020

Dear Mr. Vagts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K202259

#### Device Name

Parcus GFS II and GFS Mini. GFS Naked. GFS Ultimate. GFS BTB. ATLAS. Actiflip Cinch. Actiflip Whip. GFS Ultimate Hip and Synd-EZ Ti.

#### Indications for Use (Describe)

The Parcus GFS II and GFS Mini are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

The Parcus GFS Naked devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

The GFS Ultimate devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

The GFS BTB devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair

The Parcus ATLAS is indicated for adjunct fixation of the 1st and 2nd metacarpals in CMC arthroplasty.

The Parcus Actifiip Naked, Actiflip Cinch, and Actifip Whip are used for fixation of bone or soft tissue to bone, and is intended as fixation post, a distributing suture tension over areas of ligament or tendon repair in the knee, shoulder, and elbow and may include the following indications: anterior cruciate ligament, pectoralis repair (minor), biceps tendon repair and reattachment (distallproximal), acromioclavicular repair, and ulnar collateral ligament reconstruction.

The Parcus GFS Ultimate Hip is used for fixation of soft tissue to bone in the hip and is indicated for ligamentum teres reconstruction.

The Parcus Synd-EZ Ti are intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. The Parcus Synd-EZ Ti are intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

Type of Use (Select one or both, as applicable)

网 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Anika Therapeutics. The logo consists of a blue abstract symbol on the left, followed by the word "ANIKA" in blue, sans-serif capital letters. The abstract symbol appears to be a stylized letter "A" or a geometric shape, and it is positioned to the left of the company name.

Image /page/3/Picture/1 description: The image shows the logo for "Parcus Medical, LLC". The logo is in blue and features a stylized font. Below the logo, the text "Parcus Medical has joined Anika" is written in a smaller, gray font.

K202259

| 510(k) Summary            |                                                                                                                                                                                                               |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner & Submitter: | Parcus Medical, LLC<br>6423 Parkland Dr<br>Sarasota, FL 34243                                                                                                                                                 |
| Company Contact:          | Paul Vagts<br>Phone: (941)755-7965<br>Fax: (941)755-6543                                                                                                                                                      |
| Date Prepared:            | October 26th, 2020                                                                                                                                                                                            |
| Device Trade Name:        | Parcus GFS II and GFS Mini<br>Parcus GFS Naked<br>Parcus GFS Ultimate<br>Parcus GFS BTB<br>Parcus ATLAS<br>Parcus Synd-EZ Ti<br>Parcus Actiflip Naked, Actiflip Cinch, Actiflip Whip, and<br>GFS Ultimate Hip |
| Common Name:              | Fastener, Fixation, Non-Degradable, Soft Tissue                                                                                                                                                               |
| Device Class:             | Class II                                                                                                                                                                                                      |
| Classification Name:      | 21 CFR 888.3040 - Product Code MBI                                                                                                                                                                            |

## Predicate Device:

K133757 (cleared February 4, 2014) - Parcus GFS II and GFS Mini K152711 (cleared November 20, 2015) - Parcus GFS Naked K162198 (cleared November 18, 2016) - Parcus GFS Ultimate K183331 (cleared February 7, 2019) - Parcus GFS BTB K190375 (cleared April 24, 2019) - Parcus ATLAS K191783 (cleared October 9, 2019) - Parcus Synd-EZ Ti K192750 (cleared January 16th, 2020) - Parcus Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip

## Device Description:

The Parcus GFS II, GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip are a range of devices design for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. Each device is made from a medical grade titanium alloy, Ti-6AI4V ELI, and may also include an UHMWPE polyblend suture component. Each device is provided sterile.

PARCUS MEDICAL, LLC.

Page 1 of 3 1-941-755-7965

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Anika Therapeutics. The logo consists of a blue circular graphic on the left and the word "ANIKA" in blue capital letters on the right. The graphic appears to be a stylized letter "A" inside of a circle.

Image /page/4/Picture/1 description: The image shows the logo for Parcus Medical, LLC. The logo is in blue and features the company name in a stylized font. Below the logo, the text "Parcus Medical has joined Anika" is written in a smaller, gray font. The text indicates that Parcus Medical has been acquired by or merged with Anika.

K202259

# Intended Use:

The Parcus GFS II and GFS Mini are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

The Parcus GFS Naked devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

The GFS Ultimate devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

The GFS BTB devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

The Parcus ATLAS is indicated for adjunct fixation of the 1* and 200 metacarpals in CMC arthroplasty.

The Parcus Actiflip Naked, Actiflip Cinch, and Actiflip Whip are used for fixation of bone to bone or soft tissue to bone, and is intended as fixation post, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in the knee, shoulder, and elbow and may include the following indications: anterior cruciate ligament, posterior cruciate ligament, pectoralis repair (minor/major), biceps tendon repair and reattachment (distal/proximal), acromioclavicular repair, and ulnar collateral ligament reconstruction.

The Parcus GFS Ultimate Hip is used for fixation of soft tissue to bone in the hip and is indicated for ligamentum teres reconstruction.

The Parcus Synd-EZ Ti are intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. The Parcus Synd-EZ Ti are intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis (syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

# Substantial Equivalence Summary:

The Parcus GFS II, GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip are equivalent to the predicate Parcus devices because they are the same devices. No changes from the

PARCUS MEDICAL, LLC. 6423 Parkland Drive Sarasota, FL 34243 www.parcusmedical.com

Page 2 of 3 1-941-755-7965

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "ANIKA" in blue, with a stylized blue circle to the left of the word. The circle contains a white dot and a curved line. The font of the word "ANIKA" is sans-serif and the letters are evenly spaced.

Image /page/5/Picture/1 description: The image shows the Parcus Medical logo. The logo is blue and features the word "PARCUS" in a bold, sans-serif font. Below the word "PARCUS" is the word "MEDICAL, LLC" in a smaller font. Below the logo is the text "Parcus Medical has joined Anika".

K202259

existing Parcus Medical devices are proposed except for the inclusion of the MR Conditional parameters into the applicable Instructions for Use. Since the biocompatibility, packaging, shelf-life and sterilization process remain unchanged, no further testing or assessments were needed. LAL testing has been tested on representative samples and it was concluded that the Parcus GFS II, GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip do not raise any additional concern.

## Summary Performance Data:

The Parcus GFS II, GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip were evaluated for use in the MR Environment and were determined to fit the definition of MR Conditional. These titanium implants were evaluated based on the FDA Guidance Document - Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment and applicable ASTM standards, and worst-case devices were selected for testing. Devices were tested for magnetically induced force, magnetically induced torque, heating by RF fields, and image artifact. The results of this testing have been used to establish the MR system conditions suitable for safe use when these implants are present.

The Parcus GFS II, GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip have been determined to be substantially equivalent to the predicate devices stated above.

6423 Parkland Drive Sarasota, FL 34243 www.parcusmedical.com

Page 3 of 3 1-941-755-7965

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K202259](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K202259)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
