← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K200503

# ATL Anchors (K200503)

_International Life Sciences · MBI · Sep 2, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K200503

## Device Facts

- **Applicant:** International Life Sciences
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Sep 2, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The ATL Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications: Shoulder: Rotator Cuff Repair, Acromioclavicular Sepair, Bankart Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Oblique Ligament Repair, Extra Capsular Reconstruction, Illiotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair. Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment. Hand/Wrist: Scapholunate Ligament Reconstruction, Unar or Radial Collateral Ligament Reconstruction, TFCC. Hip: Acetabular labral repair.

## Device Story

ATL Anchors are press-fit suture anchors designed to secure soft tissue to bone; used in orthopedic surgeries including shoulder, knee, foot/ankle, elbow, hand/wrist, and hip repairs. Device consists of PEEK anchor pre-mounted on single-use driver, supplied with USP 0 UHMWPE suture. Surgeon creates pilot hole using cannulated or solid stainless-steel drill bits; anchor is tapped into place using standard mallet. Device provides mechanical fixation for sutures; enables reattachment of ligaments/tendons to bone. Benefits include stable tissue fixation during healing. Used in clinical/OR settings by orthopedic surgeons.

## Clinical Evidence

Bench testing only. Performance data confirmed specifications for insertion strength, pull-out strength, and cyclic loading. Side-by-side testing performed against Parcus Knotless PEEK (pull-out) and Smith and Nephew Bioraptor (cyclic loading). Biocompatibility evaluated per ISO 10993-1:2018; sterilization validated per ISO 11137-1:2006/11137-2:2011; shelf-life validated per ASTM D4169:2016 and ASTM F1980:2016; bacterial endotoxin testing met ANSI/AAMI/ST72:2011 limits.

## Technological Characteristics

Material: PEEK. Suture: USP 0 UHMWPE. Form factor: Press-fit anchor, pre-mounted on single-use driver. Accessories: Stainless-steel drill bits. Sterilization: Validated per ISO 11137. Standards: ISO 10993-1 (biocompatibility), ASTM D4169/F1980 (shelf-life), ANSI/AAMI/ST72 (endotoxins).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Parcus Knotless PEEK and PEEK CF Push-In Suture Anchors ([K193295](/device/K193295.md))
- Smith and Nephew BIORAPTOR Knotless Suture Anchor ([K093428](/device/K093428.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 2, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

International Life Sciences Tiffini Wittwer VP, Quality, Clinical, Regulatory 2252 Northwest Parkway SE Suite G Marietta, Georgia 30067

Re: K200503

Trade/Device Name: ATL Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: July 22, 2020 Received: July 29, 2020

Dear Tiffini Wittwer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200503

Device Name

ATL Anchors

Indications for Use (Describe)

The ATL Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:

• Shoulder: Rotator Cuff Repair, Acromioclavicular Sepair, Bankart Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.

• Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Oblique Ligament Repair, Extra Capsular Reconstruction, Illiotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.

· Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.

· Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment.

· Hand/Wrist: Scapholunate Ligament Reconstruction, Unar or Radial Collateral Ligament Reconstruction, TFCC.

· Hip: Acetabular labral repair.

Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary – K200503

| Submitter:                        | International Life Sciences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                   | Tiffini Wittwer<br>VP of QA, CA, RA<br>2252 Northwest Pkwy SE Suite G<br>Marietta GA 30067<br>Phone: 707.799.6732<br>E-mail: twittwer@mededge.io                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Trade Name:                       | ATL Anchors                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Common Name:                      | Smooth or threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification:                   | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code:                     | MBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation                        | 21 CFR 888.3040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Predicate Device(s):              | The subject device is equivalent to the following devices:<br>● Parcus Knotless PEEK and PEEK CF Push-In<br>Suture Anchors – K193295<br>● Smith and Nephew BIORAPTOR Knotless<br>Suture Anchor – K093428                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Description:               | The ATL Anchors are press fit suture anchors designed to<br>secure the implant in bone and provide suture fixation.<br>There are two configurations: with and without an eyelet.<br>The devices are made from PEEK, are provided sterile, and<br>pre-mounted on a single-use driver. The devices are<br>provided with USP 0 ultra-high molecular weight<br>polyethylene (UHMWPE) suture. ATL Anchors are used<br>with Artelon® provided cannulated or solid stainless-steel<br>drill bits. The devices are tapped into a pilot hole using a<br>standard mallet (not provided).                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indication for Use:               | The ATL Anchors are indicated for attachment of soft<br>tissue to bone. This product is intended for the following<br>indications:<br>Shoulder: Rotator Cuff Repair, Acromioclavicular<br>Separation Repair, Bankart Repair, Bankart Lesion<br>Repair, Biceps Tenodesis, Capsular Shift or<br>Capsulolabral Reconstruction, Deltoid Repair, SLAP<br>Lesion Repair. Knee: Medial Collateral Ligament Repair, Lateral<br>Collateral Ligament Repair, Posterior Oblique<br>Ligament Repair, Extra Capsular Reconstruction,<br>Illiotibial Band Tenodesis, Patellar Ligament and<br>Tendon Avulsion Repair. Foot/Ankle: Lateral Stabilization, Medial Stabilization,<br>Midfoot Reconstruction, Achilles Tendon Repair,<br>Hallux Valgus Reconstruction, Metatarsal Ligament<br>Repair. Elbow: Tennis Elbow Repair, Biceps Tendon<br>Reattachment. Hand/Wrist: Scapholunate Ligament Reconstruction,<br>Ulnar or Radial Collateral Ligament Reconstruction,<br>TFCC. Hip: Acetabular labral repair. |
| Technological<br>Characteristics: | The ATL Anchors are identical in intended use, principles<br>of operation (push / tap in), and material to both predicate<br>devices. Both the subject device and the predicates are<br>provided preloaded onto an insertion tool with suture. The<br>proposed device and predicates are similar in size,<br>configurations, and packaging.<br>The subject device differs from the predicates in retention<br>feature (helical versus radial grooves), driver mating<br>feature (hexalobe versus square), and the subject device<br>does not have a suture locking feature whereas the<br>predicate, Smith and Nephew Bioraptor, does.<br>Performance bench testing demonstrates that these<br>differences do not raise new questions of safety and<br>efficacy of the ATL Suture Anchor when compared to the<br>predicates.                                                                                                                                                                    |

{4}------------------------------------------------

{5}------------------------------------------------

| Summary Performance<br>Data: | Performance data demonstrated that the ATL Anchors met<br>performance specifications and are comparable to the<br>predicate. ATL Anchors met specification for insertion<br>strength, pull-out strength, and cyclic loading. Side by side<br>testing with the predicates was performed for pull-out<br>strength (Parcus Knotless PEEK) and cyclic loading<br>(Smith and Nephew Bioraptor).                                                                                     |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                              | A sterilization validation and routine sterilization has been<br>established in accordance with ISO 11137-1:2006 (R2015)<br>and ISO 11137-2:2011 (R2019). Shelf-life of the ATL<br>Anchors has been established in accordance with ASTM<br>D4169:2016 and ASTM F1980:2016.<br>The biocompatibility of the ATL Anchor was evaluated<br>against the requirements per ISO 10993-1:2018. All<br>acceptance criteria were met.<br>Bacterial endotoxin testing was completed and met |
|                              | acceptable endotoxin limits per ANSI/AAMI/ ST72:2011.<br>Based on the above testing and information, the ATL<br>Anchor is considered substantially equivalent to the<br>predicates and other currently marketed devices.                                                                                                                                                                                                                                                       |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K200503](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K200503)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
