Coronet Soft Tissue Fixation System

K200028 · Conextions Medical · MBI · Mar 4, 2020 · Orthopedic

Device Facts

Record IDK200028
Device NameCoronet Soft Tissue Fixation System
ApplicantConextions Medical
Product CodeMBI · Orthopedic
Decision DateMar 4, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

Coronet Soft Tissue Fixation System is intended for fixation of tissue to bone and tissue to tissue. This product is intended for the following indications: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valqus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar and Radial Ligament Reconstruction. Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar and Radial Ligament Reconstruction.

Device Story

Coronet Soft Tissue Fixation System is a single-use, sterile orthopedic implant for tissue-to-bone and tissue-to-tissue fixation. System comprises an anchor and soft tissue washer connected by a continuous suture loop, pre-assembled on an inserter. Surgeon uses provided drill bit to create pilot hole in bone; inserter facilitates concurrent placement of anchor and washer. Device used in clinical settings by orthopedic surgeons. Output is mechanical fixation of soft tissue to bone or other tissue, aiding healing and structural repair. Benefits include secure tissue attachment during reconstructive surgeries.

Clinical Evidence

Bench testing only. Performance data includes static pullout, cyclic, insertion, and fatigue limit testing. Biocompatibility testing (cytotoxicity, sensitization, irritation, pyrogen, MEM elution, acute systemic injection, intramuscular implantation), sterilization validation, MR compatibility (displacement, torque, heating, image artifact), and packaging integrity/accelerated aging tests were performed. Human factors end-user validation was also conducted.

Technological Characteristics

System consists of an anchor, soft tissue washer, and continuous suture loop. Inserter and drill bit included. Materials are non-degradable. Mechanical fixation principle. Sterile, single-use device.

Indications for Use

Indicated for patients requiring soft tissue-to-bone or tissue-to-tissue fixation in orthopedic procedures including shoulder (rotator cuff, Bankart, SLAP, biceps, AC, deltoid, capsular repairs), foot/ankle (stabilization, Achilles, hallux valgus, midfoot, metatarsal repairs), knee (collateral/patellar ligament repairs), hand/wrist (ligament reconstruction), and elbow (biceps, tennis elbow, ligament reconstruction).

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 4, 2020 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. CoNextions Medical % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103 Re: K200028 Trade/Device Name: Coronet Soft Tissue Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: January 6, 2020 Received: January 6, 2020 Dear Ms. Hogan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Laura Rose. Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K200028 Device Name #### Coronet Soft Tissue Fixation System Indications for Use (Describe) Coronet Soft Tissue Fixation System is intended for fixation of tissue to tissue to tissue. This product is intended for the following indications: - Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-clavicular ● Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction - . Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valqus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair - Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, ● Posterior Oblique Ligament Repair - Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar and Radial Ligament Reconstruction ● - Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar and Radial Ligament Reconstruction ● Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. " {3}------------------------------------------------ #### K200028 510(k) SUMMARY CoNextions Medical's Coronet Soft Tissue Fixation System #### Applicant Information | Manufacturer: | CoNextions Medical, Inc<br>150 North Wright Brothers Drive, Suite 560<br>Salt Lake City, Utah 84116<br>Ph: 201.419.0118<br>F: 801.436.5369 | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Matthew Swift, Director of Marketing | | Date Prepared: | March 2, 2020 | | Device Information | | | Proprietary Name: | Coronet Soft Tissue Fixation System | | Common Name: | Soft Tissue Fixation Device | | Classification Name: | MBI (888.3040) Fastener, fixation, nondegradable, soft tissue | | Predicate Devices: | K162429 ZipE Knotless Tissue Repair and Attachment Devices,<br>Ziptek LLC. (Primary)<br>K070882 TwinLoopTM-PEEK Anchor, Stryker, Inc.<br>K013806 Synthes ® Metallic Spiked Washer, Depuy Synthes, Inc.<br>K130033 ToggleLocTM System, Zimmer Biomet, Inc. | ## Device Description The Coronet Soft Tissue Fixation System consists of a single-use, sterile implant used to fixate tissue to bone and tissue to tissue. The implant consists of an anchor and soft tissue washer interconnected via a continuous loop of suture. The implant comes pre-assembled to an inserter, which facilitates concurrent placement of the anchor and soft tissue washer. A drill bit is provided to allow for the creation of a properly sized pilot hole in the bone implantation site. #### Intended Use/Indications for Use Coronet Soft Tissue Fixation System is intended for fixation of tissue to bone and tissue to tissue. This product is intended for the following indications: - Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps O Tenodesis, Acromio-clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction {4}------------------------------------------------ - Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux o Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair - Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar o Tendon Repair, Posterior Oblique Ligament Repair - Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar and Radial Ligament o Reconstruction - Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar and Radial о Ligament Reconstruction ## Technological Characteristics Coronet Soft Tissue Fixation System is substantially equivalent to other legally marketed bone anchor devices. Specifically, use of an anchor for bone attachment with a washer for soft tissue attachment is substantially equivalent to the ZipE® Knotless Tissue Repair and Attachment Device from Ziptek LLC (K162429). Coronet and the ZipE device are similar in design, and have same intended use, same specific indications, similar technological characteristics and similar principles of operation. The anchor component of Coronet is equivalent to the TwinLoop™-PEEK Anchor from Stryker, Inc. (K070882), which has the same intended use, similar specific indications, similar technological characteristics and similar principles of operation. The soft tissue washer component of Coronet is equivalent to the Synthes® Spiked Metallic Washer from Depuy Synthes, Inc (K013806) which has a similar intended use and similar technological characteristics and principles of operation. Benchtop testing comparing Coronet to the ZipE and TwinLoop predicates demonstrates the Coronet system is substantially equivalent and introduces no new for different questions of safety or effectiveness. ## Performance Data | Testing Category | Specific Tests Performed | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Bench Testing | Static pullout testing<br>Cyclic testing<br>Insertion testing<br>Fatigue limit testing | | Human Factors | End-user validation | | Biocompatibility Testing | Cytotoxicity<br>Sensitization<br>Irritation<br>Pyrogen testing<br>MEM elution testing<br>Acute system injection<br>Intramuscular implantation | | Sterilization validation | Bioburden<br>Relative resistance<br>EO residual testing<br>Endotoxin (LAL) validation | | MR Compatibility | Induced displacement testing<br>Induced torque testing<br>Induced heating testing | The following non-clinical testing was conducted on Coronet Soft Tissue Fixation System. {5}------------------------------------------------ | | Image artifact testing | |-----------|-------------------------| | Packaging | Package integrity tests | | | Accelerated aging | #### Conclusions The Coronet Soft Tissue Fixation System is substantially equivalent to predicate devices with respect to its indications for use, design, and function. The minor technological differences between Coronet and the predicate devices raise no new or different questions of safety or effectiveness and do not adversely impact the performance or function of the device as confirmed by performance testing. Thus, Coronet is substantially equivalent to the predicates.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...