← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K193377

# TopLock Anchor System (K193377)

_Fusion Orthopedics, LLC · MBI · Nov 25, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K193377

## Device Facts

- **Applicant:** Fusion Orthopedics, LLC
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Nov 25, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The TopLock Anchor System is intended to secure soft tissue to bone of: The Shoulder: Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation, Rotator Cuff Repair, Capsule Repair, Biceps Tenodesis. Deltoid Repair The Elbow: Ulnar or Radial Collateral Ligament Reconstruction, Bicep Tendon Reconstruction, Tennis Elbow Repair The Hand and Wrist: Scapholunate Ligament Reconstruction, Ulnar / Radial Collateral Ligament Reconstruction, Extensor Tendon Repair, Flexor Tendon Repair The Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Tendon Repair The Ankle and Foot: Lateral Ligament Stabilization Repair, Medial Ligament Stabilization Repair. Achilles Tendon Repair / Reconstruction, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Mid and Rear Foot Reconstruction

## Device Story

TopLock Anchor System comprises barbed suture anchors; used to secure soft tissue to bone in orthopedic procedures. Anchors manufactured from PEEK; diameters 3.5mm to 5.0mm; include eyelet options for suture/suture tape. Specialized instruments made of surgical grade stainless steel. Used by orthopedic surgeons in clinical/OR settings. Device functions as mechanical fastener; provides fixation for ligament/tendon repairs. Benefits patient by stabilizing soft tissue during healing process.

## Clinical Evidence

No clinical studies were performed. Evidence is based on bench testing, including cleaning, packaging, and sterilization validation, and mechanical testing (insertion, axial pull-out, cyclic fatigue, post-fatigue pullout). Pyrogenicity evaluated via LAL assay, meeting endotoxin limits of 20 EU/device.

## Technological Characteristics

Materials: PEEK (ASTM F2026), UHMWPE suture (ASTM F2848), stainless steel (ASTM F899), aluminum (ASTM B209), nylon. Barbed anchor design; 3.5mm-5.0mm diameters. Mechanical fixation principle. Non-active, non-implantable software.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Core Essence reNOVO Suture Anchors ([K071520](/device/K071520.md))
- Arthrex PEEK Mini PushLock Anchors ([K101679](/device/K101679.md))

## Reference Devices

- Core Essence Orthopaedics, Inc., TAC-tite™ PEEK ([K111716](/device/K111716.md))
- Parcus Medical Twist & Triple Twist PEEK Screw-In Suture Anchors ([K120942](/device/K120942.md))
- Parcus Medical PEEK CF Push-In Anchors ([K102162](/device/K102162.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 25, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Fusion Orthopedics, LLC Kolby Black 4135 S. Power Rd., Suite 110 Mesa, Arizona 85212

Re: K193377/S001

Trade/Device Name: TopLock Anchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: October 28, 2020 Received: October 30, 2020

Dear Mr. Black:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laura C. Rose. Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193377

#### Device Name TopLock Anchor System

Indications for Use (Describe)

The TopLock Anchor System is intended to secure soft tissue to bone of:

The Shoulder:

Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation, Rotator Cuff Repair, Capsule Repair, Biceps Tenodesis. Deltoid Repair

The Elbow:

Ulnar or Radial Collateral Ligament Reconstruction, Bicep Tendon Reconstruction, Tennis Elbow Repair The Hand and Wrist:

Scapholunate Ligament Reconstruction, Ulnar / Radial Collateral Ligament Reconstruction, Extensor Tendon Repair, Flexor Tendon Repair

#### The Knee:

Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Tendon Repair

#### The Ankle and Foot:

Lateral Ligament Stabilization Repair, Medial Ligament Stabilization Repair.

Achilles Tendon Repair / Reconstruction, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Mid and Rear Foot Reconstruction

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Date Prepared                                                 | November 24, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By                                                  | Fusion Orthopedics, LLC<br>4135 S. Power Rd., Suite 110<br>Mesa, AZ 85212<br>800-403-6876                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Primary Contact                                               | Kolby Black<br>4135 S. Power Rd., Suite 110<br>Mesa, AZ 85212<br>800-403-6876 Tele<br>e-mail: FDA@fusionorthopedics.com                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name                                                    | TopLock Anchor System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Submission<br>Number                                          | K193377                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification<br>Name                                        | Fastener, Fixation, Non-degradable, Soft Tissue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Class                                                         | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Product Code                                                  | MBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| CFR Section                                                   | 21 CFR Section 888.3040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Panel                                                  | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Primary Predicate<br>Device                                   | Core Essence reNOVO Suture Anchors (K071520)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Secondary<br>Predicate Device                                 | Arthrex PEEK Mini PushLock Anchors (K101679)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Reference<br>Devices                                          | Core Essence Orthopaedics, Inc., TAC-tite™ PEEK (K111716)<br>Parcus Medical Twist & Triple Twist PEEK Screw-In Suture Anchors (K120942)<br>Parcus Medical PEEK CF Push-In Anchors (K102162)                                                                                                                                                                                                                                                                                                                                                                                    |
| Device<br>Description                                         | The TopLock Anchor System consists of a collection of barbed suture anchors,<br>manufactured from medical grade Polyetheretherketone or PEEK (ASTM F2026). The<br>anchors are offered from diameters of 3.5mm to 5.0mm with each size offering two eyelet<br>size options to accommodate suture and suture tape. The size range and configurations are<br>offered for surgical treatment of ligament, tendon and soft tissue pathologies of the shoulder<br>and other joints. The specialized instruments are made primarily of surgical grade stainless<br>steel (ASTM F899). |
| Indications for<br>Use                                        | The TopLock Anchor System is intended to secure soft tissue to bone of:<br>The Shoulder:<br>Bankart Repair<br>SLAP Lesion Repair<br>Acromio-Clavicular Separation                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                               | Capsule Repair<br>Biceps Tenodesis<br>Deltoid Repair                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                               | The Elbow:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                               | Ulnar or Radial Collateral Ligament Reconstruction<br>Bicep Tendon Reconstruction<br>Tennis Elbow Repair                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                               | The Hand and Wrist:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                               | Scapholunate Ligament Reconstruction<br>Ulnar / Radial Collateral Ligament Reconstruction<br>Extensor Tendon Repair<br>Flexor Tendon Repair                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                               | The Knee:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                               | Medial Collateral Ligament Repair<br>Lateral Collateral Ligament Repair<br>Posterior Oblique Ligament Repair<br>Extra Capsular Reconstruction<br>Iliotibial Band Tenodesis<br>Patellar Tendon Repair                                                                                                                                                                                                                                                                                                                                                                           |
|                                                               | The Ankle and Foot:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                               | Lateral Ligament Stabilization Repair<br>Medial Ligament Stabilization Repair<br>Achilles Tendon Repair / Reconstruction<br>Metatarsal Ligament Repair<br>Hallux Valgus Reconstruction<br>Mid and Rear Foot Reconstruction                                                                                                                                                                                                                                                                                                                                                     |
| Materials                                                     | Polyetheretherketone (PEEK) (ASTM F2026)<br>Ultra High Molecular Weight Polyethylene Suture (UHMWPE) (ASTM F2848)<br>Stainless Steel (ASTM F899)<br>Aluminum (ASTM B209)<br>Nylon                                                                                                                                                                                                                                                                                                                                                                                              |
| Substantial<br>Equivalence<br>Claimed to<br>Predicate Devices | The TopLock Anchor System is substantially equivalent to the predicate devices in terms of<br>design, materials used, mechanical safety and performances.                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Non-clinical Test<br>Summary                                  | Validations were performed on the cleaning, packaging and sterilization of the implants and<br>associated surgical instruments. Mechanical Testing was performed to show design<br>equivalence including insertion testing, axial pull-out testing, cyclic fatigue testing, and post<br>fatigue pullout testing. Pyrogenicity was evaluated using the Limulus amebocyte lysate<br>(LAL) assay. The testing demonstrated that the subject device meets the recommended<br>maximum endotoxin levels of 20 (EU)/Device.                                                           |
| Clinical Test<br>Summary                                      | No clinical studies were performed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Conclusions:<br>Non clinical and                              | Fusion Orthopedics LLC considers the TopLock Anchor System to be equivalent to the<br>predicate devices listed above. This conclusion is based upon the devices' similarities in                                                                                                                                                                                                                                                                                                                                                                                               |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K193377](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K193377)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
