K171640 · Medacta International S.A. · MBI · Oct 23, 2017 · Orthopedic
Device Facts
Record ID
K171640
Device Name
M-ARS ACL: Anatomic Ribbon Surgery System
Applicant
Medacta International S.A.
Product Code
MBI · Orthopedic
Decision Date
Oct 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
M-ARS ACL Tibial Pull Suture Plate (PSP): Reconstructive treatment of ruptured anterior and posterior cruciate ligaments by means of autologous grafts. M-ARS ACL: Anatomic Ribbon Surgery System: M-ARS ACL Extracortical Femoral Button: Reconstructive therapy of ruptures to the anterior cruciate ligament by means of autologous grafts.
Device Story
M-ARS ACL Anatomic Ribbon Surgery System consists of implantable extracortical suspension devices: Extracortical Femoral Button and Tibial Pull Suture Plate. Used in ACL/PCL reconstruction surgery to secure autologous grafts to bone. Operated by orthopedic surgeons in clinical/OR settings. Femoral button secures graft on femoral side; Tibial Pull Suture Plate secures graft on tibial side. Devices provide mechanical fixation of graft to bone, facilitating ligament healing. Benefits include stable graft fixation during the reconstruction process.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including geometrical design comparison, mechanical analysis, and surgeon evaluations via cadaver workshops.
Technological Characteristics
Materials: Titanium-6 Aluminum-4 Vanadium ELI (ISO 5832-3:1996, ASTM F136-13). Components: Extracortical Fixation Button (12.3mm x 4mm x 1.5mm) and Tibial Pull Suture Plate (20mm x 13mm x 1.2mm, C-shape). Fixation: Extracortical suspension. Sterilization: Sterile. Single-use. No software or electronic components.
Indications for Use
Indicated for reconstructive treatment of ruptured anterior and posterior cruciate ligaments using autologous grafts in patients requiring ACL or PCL reconstruction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
K133606 — N8TIVE ACL ANATOMIC RECONSTRUCTION SYSTEM, N8TIVE ACL FEMORAL IMPLANT, N8TIVE ACL TIBIAL IMPLANT, VARIOUS CLASS I ANCILL · ArthroCare Corporation · Apr 1, 2014
K250844 — PULLUP®; PULLUP® CLIP; BT LOOP®; PULLUP® TEX CLIP; RIGIDLOOP T; RIGIDLOOP Clip; RIGIDLOOP BTB; RIGIDLOOP Suture Loop · Science & Bio Materials (S.B.M.) Sas · Jun 2, 2025
K221389 — FairFix AM Adjustable Button FairFix PSP Adjustable Button · Medacta International S.A. · Jun 8, 2022
K221128 — Arthrex ACL TightRope®, PCL TightRope®, and TightRope® II · Arthrex, Inc. · Oct 5, 2022
K031666 — ARTHREX FIBERWIRE BUTTON REPAIR KIT, MODEL AR-8920DS/AR-8921DS · Arthrex, Inc. · Nov 18, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
Medacta International SA % Ms. Elizabeth Rose Manager, Regulatory Affairs Mapi USA, Inc. 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504
October 23, 2017
Re: K171640
Trade/Device Name: M-ARS ACL: Anatomic Ribbon Surgery System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: September 11, 2017 Received: September 12, 2017
Dear Ms. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171640
Device Name
M-ARS ACL: Anatomic Ribbon Surgery System
Indications for Use (Describe)
M-ARS ACL: Anatomic Ribbon Surgery System - M-ARS ACL Tibial Pull Suture Plate (PSP): Reconstructive treatment of ruptured anterior and posterior cruciate ligaments by means of autologous grafts.
M-ARS ACL: Anatomic Ribbon Surgery System: M-ARS ACL Extracortical Femoral Button: Reconstructive therapy of ruptures to the anterior cruciate ligament by means of autologous grafts.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <div></div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 3.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Manager Date Prepared: June 01, 2017 Date Revised: October 17, 2017
#### II. Device
| Device Proprietary Name: | M-ARS ACL: Anatomic Ribbon Surgery System |
|--------------------------|------------------------------------------------|
| Common or Usual Name: | Suture Anchor |
| Classification Name: | Fastener, Fixation, Nondegradable, Soft Tissue |
| Primary Product Code: | MBI |
| Regulation Number: | 21 CFR 888.3040 |
| Device Classification | 2 |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
- KSEA Flipptack, K982571, Karl Storz Endoscopy - America, Inc. (also referred to as Karl Storz's Flipptack)
- . Endotack, K022853, Karl Storz Endoscopy - America, Inc.
The following reference devices are cited within the submission:
- M.U.S.T. Extension, K132878, Medacta International SA ●
- M.U.S.T. Pedicle Screw System, K141988, Medacta International SA
- M.U.S.T. Pedicle Screw System, K153664, Medacta International SA
- M.U.S.T. Pedicle Screw System, K162061, Medacta International SA
#### IV. Device Description
The purpose of this submission is to gain clearance for the M-ARS ACL: Anatomic Ribbon Surgery System. The M-ARS ACL: Anatomic Ribbon Surgery System components are implantable devices used for the reconstructive treatment of ligament ruptures as well as the fixation of an implanted Anterior Cruciate Ligament graft by means of suspensory extracortical fixation.
{4}------------------------------------------------
The M-ARS ACL: Anatomic Ribbon Surgery System Extracortical Fixation Button component is an implantable extracortical suspension device used to secure the graft fixation on the femoral side during ACL reconstruction surgery. The M-ARS ACL: Anatomic Ribbon Surgery System Tibial Pull Suture Plate (PSP) component is an implantable extracortical suspension device used to secure the graft fixation on the tibial side during ACL reconstruction surgery.
The M-ARS ACL: Anatomic Ribbon Surgery System Extracortical Fixation Button and Tibial Pull Suture Plate are manufactured with Titanium-6 Vanadium Extra Low Interstitial (Ti-6Al-4V ELI).
#### V. Indications for Use
## M-ARS ACL Extracortical Femoral Button
Reconstructive therapy of ruptures to the anterior and posterior cruciate ligament by means of autologous grafts.
## M-ARS ACL Tibial Pull Suture Plate (PSP)
Reconstructive treatment of ruptured anterior and posterior cruciate ligaments by means of autologous grafts.
#### Comparison of Technological Characteristics VI.
The M-ARS ACL: Anatomic Ribbon Surgery System and the predicate devices share the following characteristics:
- Material;
- Sterile: ●
- device usage; and ●
- Extracortical Fixation Button; ●
- . length;
- . width:
- . thickness;
- number and purpose of holes; and .
- . critical section thickness.
The M-ARS ACL: Anatomic Ribbon Surgery System is technologically different from the predicate devices as follows:
- Tibial Pull Suture Plate, and
- length;
- . width:
- thickness; ●
- number and purpose of holes; ●
- . geometry; and
- . critical section dimensions
- Extracortical Fixation Button ●
- . critical section width.
{5}------------------------------------------------
The M-ARS ACL: Anatomic Ribbon Surgery System Extracortical Fixation Button and Tibial Pull Suture Plate are manufactured with Titanium-4 Vanadium Extra Low Interstitial Alloy according to ISO 5832-3:1996 Implants For Surgery - Metallic Materials - Part 3: Wrought Titanium 6-Aluminum 4-Vanadium Alloy and ASTM F136-13 Standard Specification For Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy For Surgical Implant Applications.
This material, used with Karl Storz's Flipptack (K982571) and Karl Storz's Endotack (K022853), has an extensive biocompatibility history and has been previously reviewed by FDA.
Due to the extensive history of use in currently marketed medical devices, additional biocompatibility testing was deemed unnecessary for the M-ARS ACL: Anatomic Ribbon Surgery System components. In addition, manufacturing processes for the subject devices are identical or similar to the reference devices the M.U.S.T. Pedicle Screw Systems (K162061, K153664, K141988, and K132878).
The tables below compare key technological features between the subject and predicate devices.
| Parameters | M-ARS ACL: Anatomic Ribbon<br>Surgery System Extracortical<br>Fixation Button<br>(Subject Device) | Karl Storz's Flipptack<br>K982571<br>(Predicate Device) |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Titanium Alloy (ELI) | Titanium Alloy |
| Length | 12.3 mm | 12.3 mm |
| Width | 4 mm | 4 mm |
| Thickness | 1.5 mm | 1.5 mm |
| Holes | 2 for suture loop for graft fixation<br>1 lateral hole for suture used to flip<br>implant<br>1 lateral hole for suture used to<br>pull construct | 2 for suture loop for graft fixation<br>1 lateral hole for suture used to flip<br>implant<br>1 lateral hole for suture used to pull<br>construct |
| Critical Section<br>Dimensions | Thickness: 1.5 mm<br>Width: 1.7 mm | Thickness: 1.5 mm<br>Width: 1.3 mm |
| Device Usage | Single Use | Single Use |
| Biocompatibility | Implant with permanent > 30 day | Implant with permanent > 30 day |
| Sterilization | Sterile | Sterile |
### Technological comparison
{6}------------------------------------------------
| Parameters | M-ARS ACL: Anatomic Ribbon<br>Surgery System Tibial Pull<br>Suture Plate<br>(Subject Device) | Karl Storz's Endotack<br>K022853<br>(Predicate Device) |
|------------------|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Material | Titanium Alloy (ELI) | Titanium Alloy |
| Length | 20 mm | 16 mm |
| Width | 13 mm | 10 mm |
| Thickness | 1.2 mm | 2 mm |
| Holes | 2 central holes for adjustable suture<br>loop<br>2 lateral holes for additional<br>sutures, graft fixation and tension | 2 holes used for graft fixation and<br>tension |
| Geometry | C-Shape | Round Shape |
| Critical Section | Thickness: 2 mm | Thickness: 3 mm |
| Dimensions | Width: 2.84 mm | Width: 2 mm |
| Device Usage | Single Use | Single Use |
| Biocompatibility | Implant with permanent > 30 day | Implant with permanent > 30 day |
| Sterilization | Sterile | Sterile |
### Discussion
As seen above, the technological differences between the subject and predicate devices do not raise new questions of safety and effectiveness. The M-ARS ACL: Anatomic Ribbon Surgery System is the same or similar to the predicate devices in terms of materials of construction, device usage, and sterility. Although there are some differences in the design attributes of the Tibial Pull Suture Plate, the intended use and functionality of the component are the same. Based on the comparison of technological characteristics and performance data provided within this submission, the data supports the substantial equivalence of the M-ARS ACL: Anatomic Ribbon Surgery System to the identified predicate devices.
#### VII. Performance Data
Risks were identified based on the proposed design and testing was conducted to mitigate those risks. Based on the risk analysis, the following studies were performed in support of a substantial equivalence determination:
Non-Clinical Studies
- . design comparison: geometrical and comparative analysis between the M-ARS ACL: Anatomic Ribbon Surgery System and the predicate devices to show mechanical equivalence:
- . cadaver workshop: surgeon evaluations to verify the design features and surgical technique: and
- substantial equivalence evaluation: a comparative analysis between the M-ARS ACL: Anatomic Ribbon Surgery System Tibial Pull Suture Plate and the predicate device.
{7}------------------------------------------------
Pyrogenicity
- Medacta uses both the Bacterial Endotoxin Test (LAL test) according to European . Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and the Pyrogen Test according to USP chapter <151> for pyrogenicity determination.
- Medacta does not intend to label the subject devices as non-pyrogenic or pyrogen free. ●
Clinical Studies
- No clinical studies were conducted ●
## VIII. Conclusion
Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics as well as performance evaluations. The M-ARS ACL: Anatomic Ribbon Surgery System is as safe and effective as the predicate devices, Karl Storz's Flipptack (K982571) and Karl Storz's Endotack (K022853).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.