SpeedLock Knotless Fixation System

K153313 · ArthroCare Corporation · MBI · Dec 11, 2015 · Orthopedic

Device Facts

Record IDK153313
Device NameSpeedLock Knotless Fixation System
ApplicantArthroCare Corporation
Product CodeMBI · Orthopedic
Decision DateDec 11, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The SpeedLock Knotless Fixation Device is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon reattachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions

Device Story

Implantable bone anchor system for arthroscopic/orthopedic soft tissue-to-bone fixation; eliminates manual surgical knot tying. System comprises PEEK anchor preloaded on disposable inserter handle. Used by surgeons in OR settings. Surgeon drills hole in bone, inserts anchor, and deploys via plug/cylinder compression mechanism to lock suture. Provides secure fixation; facilitates tissue healing; reduces surgical time by removing knot-tying requirement.

Clinical Evidence

Bench testing only. In vitro testing performed per FDA Guidance Document 'Testing Bone Anchors' (April 1996), including insertion into simulated bone substrate, static and cyclic fatigue testing, and simulated use testing. No clinical or animal data included.

Technological Characteristics

Implantable bone anchor; material: Invibio PEEK Optima LT3; suture: #2 UHMWPE MagnumWire. Anchor features 9 radial barbs for interference fit; suture locking via plug/cylinder compression. Dimensions: 3.0-3.5 mm diameter, 12-13 mm deployed length. Sterile (EtO), single-use. Mechanical fixation principle.

Indications for Use

Indicated for fixation of soft tissue to bone in shoulder, ankle, foot, elbow, and knee procedures. Applicable to patients requiring soft tissue repair/reconstruction in these anatomical regions.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other, representing the department's focus on people and health. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 11, 2015 ArthroCare Corporation Ms. Laura Kasperowicz Principal RA Specialist 15285 Alton Parkway, Suite 200 Irvine, California 92618 Re: K153313 Trade/Device Name: SpeedLock Knotless Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: November 16, 2015 Received: November 17, 2015 Dear Ms. Kasperowicz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | |-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | | 510(k) Number (if known) | K153313 | | Device Name | SpeedLock Knotless Fixation System | | Indications for Use (Describe) | The SpeedLock Knotless Fixation Device is indicated for use in fixation of soft tissue to bone. | | Examples of such procedures include: | Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair | | | Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction | | | Foot: Hallux valgus reconstruction | | | Elbow: Tennis elbow repair, biceps tendon reattachment | | | Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions | | Type of Use (Select one or both, as applicable) | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label><br><label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | FORM FDA 3881 (8/14) | | Page 1 of 1 | |----------------------|--|-------------| |----------------------|--|-------------| PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a stylized, circular graphic on the left and the word "ArthroCare" in a sans-serif font on the right. The graphic is made up of several curved lines in a dark red color, arranged to create a sense of motion or rotation. The text "ArthroCare" is in a dark gray color, and there is a registered trademark symbol next to the word "Care". # 510(k) Summary ### ArthroCare® Corporation SpeedLock® Knotless Fixation System This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### General Information | Submitter Name: | ArthroCare Corporation, a Smith & Nephew Company | |-----------------|----------------------------------------------------------------------------------------------------------| | Address | 15285 Alton Parkway, Suite 200<br>Irvine, CA. 92618 | | Contact Person: | Laura Kasperowicz<br>Principle Regulatory Affairs Specialist<br>Phone: 949-585-2406<br>Fax: 949-585-2401 | | Date Prepared: | November 16, 2015 | #### Device Name SpeedLock® Knotless Fixation System Proprietary Name: | Common Name: | Bone Anchor | |----------------------|----------------------------------------------------| | Classification Name: | Smooth or threaded metallic bone fixation fastener | | Device Class: | Class II | | Product Code: | MBI | | CFR Section: | 21 CFR 888.3040 | #### Predicate Device SpeedLock® Knotless Fixation System: K130196 (cleared February 27, 2013) #### Description The SpeedLock Knotless Fixation System is an implantable bone anchor with inserter handle designed for use in arthroscopic and orthopedic procedures. The SpeedLock is a knotless fixation device, meaning that manually tying surgical knots is not necessary for the fixation of suture to tissue. The SpeedLock Knotless Fixation System consists of two primary parts: an implantable bone anchor and an anchor inserter, which is preloaded with the anchor. The anchor inserter is a disposable tool. The entire product is packaged in a tray with a Tyvek® Iid, and the finished product is sterilized by ethylene oxide. Both the anchor and inserter are designed for single use only. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for ArthroCare. The logo consists of a circular graphic on the left and the word "ArthroCare" on the right. The circular graphic is made up of several curved lines in a maroon color. The word "ArthroCare" is written in a sans-serif font in a dark gray color, with the registered trademark symbol next to it. #### Intended Use/Indications For Use The SpeedLock Knotless Fixation Device is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon reattachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions #### Non-Clinical Data Bench testing was performed on both the proposed and predicate devices in accordance with the FDA Guidance Document, Testing Bone Anchors, April 1996. This in vitro testing involved insertion of the anchors into a simulated human bone substrate followed by both static and cyclic fatigue testing. Design Verification testing was performed in order to verify the design outputs meet the design inputs. Simulated use testing was performed on the device and accessory devices which will be promoted for use with the SpeedLock system in order to affirm safety and effectiveness. The test results demonstrate that the proposed SpeedLock meets its design, performance, and safety specifications. Based on the test results, the proposed device performs as intended and mechanical properties are substantially equivalent to the predicate device when used in accordance with labeling. #### Clinical Data No clinical or animal data are included in this submission. #### Summary All testing demonstrates that the proposed SpeedLock performs as intended and mechanical properties are substantially equivalent when used in accordance with the labeling. As the intended use, operating principle, materials and technological characteristics are comparable to the predicate device, the proposed SpeedLock Knotless Fixation System is substantially equivalent. The minor differences between the proposed SpeedLock and predicate device do not raise any new questions of safety or effectiveness. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a circular design on the left, made up of several curved red lines. To the right of the design is the company name, "ArthroCare," in a dark gray, sans-serif font. A registered trademark symbol is present to the upper right of the word "Care." | Comparison of Technological Characteristics | | | |--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------| | Characteristics | Predicate Device<br>SpeedLock<br>(K130196) | Proposed Device<br>SpeedLock | | Intended Use | Fixation of soft tissue to bone | Same | | Delivery Method | Arthroscopic and Limited Access | Same | | How Supplied | Sterile (EtO) | Same | | Suture Material | #2 UHMWPE MagnumWire® Suture | Same | | Anchor Material | Invibio PEEK Optima® LT1 | Invibio PEEK Optima® LT3 | | Design Technology | Anchor with 9 radial barbs (interference fit) | Same | | Suture Locking Mechanism | Plug/Cylinder Compression | Same | | Diameter of Anchor Body | 3.0 - 3.5 mm | Same | | Surface of Cortical Lock | 1.25 cm² | Same | | Depth of Barb | 0.25 mm | Same | | Anchor Deployed Length | 12 - 13 mm | Same | | Method of Anchor Insertion | Inserted into a drilled hole | Same | | Accessory Devices<br>Promoted for use with the<br>SpeedLock System | 3.0mm Tapered Twist Drill, 3.0mm+<br>Tapered Twist Drill, 3.5mm High Visibility<br>Drill Guide, 3.5mm Low Profile Drill<br>Guide, Sharp Tipped Obturator, PathFinder®<br>Obturator | Same and SpeedStitch® Suture Passer |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...