← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K152711

# Parcus GFS Naked (K152711)

_Parcus Medical, LLC · MBI · Nov 20, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K152711

## Device Facts

- **Applicant:** Parcus Medical, LLC
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Nov 20, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The GFS Naked are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

## Device Story

Parcus GFS Naked is a ligament retention device used for fixation of ligaments and tendons during repair procedures. Device consists of medical-grade titanium. It is provided sterile. Used by surgeons in clinical settings. Device functions as a bone fixation fastener to secure soft tissue. Performance evaluated via bench testing of worst-case configurations, including cyclic loading and ultimate failure strength testing, compared against predicate devices. Output is mechanical fixation of tissue to bone.

## Clinical Evidence

Bench testing only. Evaluation performed on worst-case configurations, including assembly with suture per instructions for use. Testing measured strength and elongation under cyclic loading and ultimate failure conditions. Results demonstrated substantial equivalence to predicate device performance data.

## Technological Characteristics

Material: medical-grade titanium. Form factor: bone fixation fastener. Sterilization: provided sterile. Principle: mechanical fixation of soft tissue to bone. No software or electronic components.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- GFS II Standard, GFS II Large and GFS Mini ([K133757](/device/K133757.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 20, 2015

Parcus Medical, LLC Mr. Paul Vagts RA/QA Manager 6423 Parkland Drive Sarasota, Florida 34243

Re: K152711

Trade/Device Name: Parcus GFS Naked Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class III Product Code: MBI Dated: October 21, 2015 Received: October 22, 2015

Dear Mr. Vagts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

| 510(k) Number (if known): | K152711          |
|---------------------------|------------------|
| Device Name:              | Parcus GFS Naked |

# Indications for Use:

The Parcus GFS Naked are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

AND/OR Prescription Use __ X __

Over the Counter Use _________________________________________________________________________________________________________________________________________________________

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

# K152711 Page 1 of 1

# 510(k) Summ

| Submitter:           | Parcus Medical, LLC<br>6423 Parkland Dr<br>Sarasota, FL 34243                                           |
|----------------------|---------------------------------------------------------------------------------------------------------|
| Company Contact:     | Paul Vagts<br>Phone: (941)755-7965<br>Fax: (941)755-6543                                                |
| Date Prepared:       | September 17, 2015                                                                                      |
| Device Trade Name:   | Parcus GFS Naked                                                                                        |
| Common Name:         | Ligament Retention Device                                                                               |
| Device Class:        | Class II                                                                                                |
| Classification Name: | Fastener, Fixation, Non-Degradable, Soft Tissue 21 CFR<br>888.3040 - Product Code MBI                   |
| Predicate Device:    | The predicate devices are the GFS II Standard, GFS II Large<br>and GFS Mini, K133757, February 4, 2014. |

# Device Description:

The Parcus GFS Naked devices are designed for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. The devices are made from medical grade titanium and are provided sterile.

# Intended Use:

The GFS Naked are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

# Substantial Equivalence Summary:

The GFS Naked are very similar to the predicate Parcus Medical GFS, GFS II and GFS Mini devices in that they are comprised of the same materials, are intended for the same indications and utilize similar designs. While the GFS Naked are offered in both smaller and larger sizes, testing has shown that this does not raise any concerns regarding the safety or efficacy of the device.

# Summary Performance Data:

The GFS Naked were evaluated and testing was conducted on the worst case configurations. Devices were assembled with suture per the DFU and placed in a test fixture. Devices were evaluated for strength and elongation under cycle loading and ultimate failure conditions. Results were compared with test data for the predicate device and demonstrated substantial equivalency.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K152711](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K152711)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
