BoneCam Suture Anchors

K152255 · Valeris Medical · MBI · Nov 16, 2015 · Orthopedic

Device Facts

Record IDK152255
Device NameBoneCam Suture Anchors
ApplicantValeris Medical
Product CodeMBI · Orthopedic
Decision DateNov 16, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

Valeris BoneCam Suture Anchors are intended to be used for suture (soft tissue) fixation to bone in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip.

Device Story

BoneCam Suture Anchors are orthopedic implants for soft tissue-to-bone fixation. Anchors feature integrated suture attachment and are pre-loaded on single-use inserters with UHMWPE sutures. Used by orthopedic surgeons in clinical/surgical settings for procedures including rotator cuff repair, ligament reconstruction, and tendon reattachment. Implants are secured into bone to provide mechanical fixation. Benefits include restoration of anatomical integrity for damaged soft tissues. Device is sterile and for single use only.

Clinical Evidence

Bench testing only. Axial pull-out testing per ASTM F543-13 performed in G20 and G40 polyurethane foam blocks using #2, 2-0, 3-0, and 4-0 sutures confirmed performance as intended.

Technological Characteristics

Materials: Titanium alloy (6Al4V ELI) per ASTM F136 or PEEK (Zeniva ZA-500) per ASTM F2026. Dimensions: 2.9mm-3.3mm diameter, 6mm-10mm length. Includes integrated UHMWPE sutures. Sterile, single-use. Mechanical fixation via threaded/smooth anchor design.

Indications for Use

Indicated for soft tissue-to-bone fixation in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip for patients requiring ligament, tendon, or capsular repair/reconstruction.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 16, 2015 Valeris Medical % Ms. Chervl Wagoner Principal Consultant Wagoner Consulting LLC P.O. Box 15729 Wilmington, North Carolina 28408 Re: K152255 Trade/Device Name: BoneCam Suture Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: August 18, 2015 Received: August 19, 2015 Dear Ms. Wagoner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152255 Device Name BoneCam Suture Anchor Indications for Use (Describe) Valeris BoneCam Suture Anchors are intended to be used for suture (soft tissue) fixation to bone in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip. Shoulder Rotator Cuff Repair Bankart Repair SLAP Lesion Repair Biceps Tenodesis Acormio-Clavicular Separation Repair Deltoid Repair Capisular Shift or Capsulolabral Reconstruction Foot/Ankle Lateral stabilization Medial stabilization Achilles tendon repair Metatarsal ligament repair Hallux Valgus reconstruction Digital tendon transfers Mid-foot reconstruction Knee Medial Collateral Ligament Repair Lateral Collateral Ligament Repair Posterior Oblique Ligament Repair Iliotibial Band Tenodesis Reconstruction Patellar Ligament/Tendon Repair Hand/Wrist Scapholunate Ligament Reconstruction Carpal Ligament Reconstruction Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital tendon transfers Elbow Biceps Tendon Reattachment Ulnar or Radial Collateral Ligament Reconstruction Hip Capsular Repair Acetabular Labral Repair {3}------------------------------------------------ Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary (as required by 21 CFR 807.92) | | (as required by 21 CFR 807.92) | |----------------|-------------------------------------------------------------------------------------| | Date Prepared | 11/12/2015 | | Manufacturer | Valeris Medical | | Address | 200 Cobb Pkwy N<br>Building 200, Suite 210<br>Marietta, GA 30062 | | Telephone | 888-404-3980 Ext 101 | | Fax | 678-669-2188 | | Contact Person | Daniel Lanois<br>President | | Address | Valeris Medical<br>200 Cobb Pkwy N<br>Building 200, Suite 210<br>Marietta, GA 30062 | | Telephone | 888-404-3980 Ext 101 | | Fax | 678-669-2188 | | email | daniel@valerismedical.com | | Trade Name | BoneCam Suture Anchor | |---------------------|----------------------------------------------------| | Common Name | Suture Anchor | | Panel Code | Orthopaedics/87 | | Classification Name | Smooth or threaded metallic bone fixation fastener | | Class | Class II | | Regulation Number | 21 CFR 888.3040 | | Product Code | MBI | | Name of Predicate Device | 510(k) # | Manufacturer | |---------------------------------------------------------|----------|--------------| | Arthrex SutureTak® Suture<br>Anchors | K140855 | Arthrex Inc. | | Valeris Apollo Suture Anchor<br>System and Titan Screws | K142230 | Valeris | | Description | The BoneCam Suture Anchor Delivery Systems are delivery systems for anchors for use in fixation of ligament, tendon, bone, or soft tissue to bone in knee, shoulder, foot/ankle, elbow, hip, and hand/wrist procedures. These anchors consist of anchors with integrated suture attachment. The Anchors are provided loaded on individual inserters with integrated UHMWPE sutures. Sterile, for single use only.<br>Anchor implants are made from either a titanium alloy (6AI4V ELI) per ASTM F136, or PEEK (Zeniva ZA-500) per ASTM F2026 from Solvay Advanced Polymers. Anchors range in size from 2.9mm to 3.3mm in diameter and 6mm to 10mm in length. | |-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications and<br>Intended Use | Valeris BoneCam Suture Anchors are intended to be used for suture<br>(soft tissue) fixation to bone in the foot, ankle, knee, hand, wrist, elbow,<br>shoulder, and hip. | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <b>Shoulder</b> | | | Rotator Cuff Repair | | | Bankart Repair | | | SLAP Lesion Repair | | | Biceps Tenodesis | {5}------------------------------------------------ | | Acormio-Clavicular Separation Repair | |--------------------------------------|------------------------------------------------------------------------------------------| | | Deltoid Repair | | | Capisular Shift or Capsulolabral Reconstruction | | | <b>Foot/Ankle</b> | | | Lateral stabilization | | | Medial stabilization | | | Achilles tendon repair | | | Metatarsal ligament repair | | | Hallux Valgus reconstruction | | | Digital tendon transfers | | | Mid-foot reconstruction | | | <b>Knee</b> | | | Medial Collateral Ligament Repair | | | Lateral Collateral Ligament Repair | | | Posterior Oblique Ligament Repair | | | Iliotibial Band Tenodesis Reconstruction | | | Patellar Ligament/Tendon Repair | | | <b>Hand/Wrist</b> | | | Scapholunate Ligament Reconstruction | | | Carpal Ligament Reconstruction | | | Repair/Reconstruction of collateral ligaments, | | | Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints<br>for all digits, | | | Digital tendon transfers | | | <b>Elbow</b> | | | Biceps Tendon Reattachment | | | Ulnar or Radial Collateral Ligament Reconstruction | | | <b>Hip</b> | | | Capsular Repair | | | Acetabular Labral Repair | | Technological<br>Characteristics and | Documentation was provided to demonstrate that the Subject device, | | Technological<br>Characteristics and<br>Substantial<br>Equivalence | Documentation was provided to demonstrate that the Subject device, Valeris<br>BoneCam Suture Anchor is substantially equivalent to the Predicate Arthrex<br>Suture Tak® Suture Anchors (K140855). The Subject device is substantially<br>equivalent to the predicate device in intended use, indications for use, materials,<br>technological characteristics, and labeling. | |--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Performance Data | Axial Pull-Out per ASTM F543-13 testing confirmed that the Subject<br>device performed as intended with #2, 2-0, 3-0, and 4-0 sutures in G20<br>PU foam and in G40 PU foam for #2 sutures. | |------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Conclusion | Based on the intended use, indications for use, technological characteristics,<br>materials, and comparison to predicate devices, the Subject Valeris BoneCa<br>Suture Anchor (Subject device) has been shown to be substantially equivalent to<br>legally marketed predicate devices |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...