K140879 · Smith & Nephew, Inc. · MBI · Jul 21, 2014 · Orthopedic
Device Facts
Record ID
K140879
Device Name
BIOSURE HEALICOIL PK INTERFERENCE SCREW
Applicant
Smith & Nephew, Inc.
Product Code
MBI · Orthopedic
Decision Date
Jul 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BIOSURE HEALICOL PK Interference Screw is indicated for the reattachment of ligament, tendon, soft tissue, or bone-tendon-bone for the following indications: Knee ACL repairs PCL repairs Extra-capsular repairs Medial collateral ligament Lateral collateral ligament Posterior oblique ligament Patellar realignment and tendon repairs Vastus medialis obliquus advancement Iliotibial band tenodesis Shoulder Acromioclavicular separation repairs Biceps tenodesis Foot and Ankle Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstructions Metatarsal ligament/tendon repairs/reconstructions Flexor hullucis longus (FHL) Tendon transfers Elbow, Wrist, and Hand Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Scapholunate ligament reconstruction Tendon transfers Carpomedicarpal ioint arthroplasty
Device Story
BIOSURE HEALICOIL PK Interference Screw is a sterile, single-use, non-absorbable PEEK bone fixation device. Features fenestrated open-lattice design to facilitate bone ingrowth; central channel allows packing with autologous bone. Used by orthopedic surgeons in clinical/OR settings for soft tissue-to-bone fixation (e.g., ACL/PCL repair, biceps tenodesis, ligament reconstruction). Device provides mechanical fixation strength comparable to standard interference screws while minimizing implanted material volume. Surgeon inserts screw into bone tunnel to secure graft; fenestrations allow bone integration over time, potentially improving long-term fixation stability.
Clinical Evidence
Bench testing (pull-out and insertion) demonstrated mechanical equivalence to predicates. Animal study (n=11) evaluated bone ingrowth using micro-computed tomography (µCT) and histology at 12 weeks; results showed new bone formation within fenestrations and central cannulations. No human clinical data provided.
Technological Characteristics
Material: Polyetheretherketone (PEEK). Design: Fenestrated, open-lattice interference screw with central cannulation. Dimensions: 6mm to 10mm diameter. Sterilization: Sterile, single-use. Principle: Mechanical bone fixation fastener.
Indications for Use
Indicated for reattachment of ligament, tendon, soft tissue, or bone-tendon-bone in knee, shoulder, foot, ankle, elbow, wrist, and hand procedures. Applicable to patients requiring surgical repair or reconstruction of specified ligaments and tendons.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{1}------------------------------------------------
# JUL 2 1 2014
#### 510(k) Summary
## Date Prepared: May 23'd, 2014
| SUBMITTERINFORMATION | CONTACT INFORMATION |
|----------------------|----------------------------------|
| Smith & Nephew, Inc. | Vivek Mukhatyar |
| 150 Minuteman Road | Regulatory Affairs Specialist II |
| Andover, MA 01810 | ' Phone: (978)749-1043 |
| | Fax: (978)749-1443 |
| DEVICE NAME (UNMODIFIED) | |
|---------------------------|--------------------------------------------------------------------------|
| Trade or proprietary name | BIOSURE HEALICOIL PK Interference Screw |
| Common or usual name | Soft Tissue Fixation Device |
| Classification name | 21 CFR §888.3040 - Smooth or threaded metallic bone fixation<br>fastener |
| Device Class | Class II |
| Product Code | MBI |
#### LEGALUY MARKETED PREDICATE DEVICE
The Smith & Nephew BIOSURE HEALICOIL PK interference screw is substantially equivalent in intended use and Fundamental Scientific Technology to the following legally marketed devices in commercial distribution:
| K083635 | Smith & Nephew BIOSURE PK<br>(Cleared on January 30, 2009) |
|---------|------------------------------------------------------------|
| K083226 | |
K042552 Biomet Bio-Core (Cleared on December 14, 2004)
# DEVICE DESCRIPTION
Smith & Nephew BIOSURE HEALICOIL PK interference screw is a non-absorbable, sterile, single-use, interference screw composed of Polyetheretherketone (PEEK). It has a fenestrated open design which allows for bone ingrowth. It is provided in size ranging from 6mm to 10 mm diameter. BIOSURE HEALICOIL interference screw minimizes the volume of material implanted into the body while providing equal fixation strength of a standard interference screw. The central channel of the screw can be packed with autologous bone.
{2}------------------------------------------------
# INTEXPED Clase
The BIOSURE HEALICOL PK Interference Screw is indicated for the reattachment of ligament, tendon, soft tissue, or bone-tendon-bone for the following indications:
#### Knee
ACL repairs PCL repairs Extra-capsular repairs Medial collateral ligament Lateral collateral ligament Posterior oblique ligament Patellar realignment and tendon repairs Vastus medialis obliquus advancement Iliotibial band tenodesis
#### Shoulder
Acromioclavicular separation repairs Biceps tenodesis
#### Foot and Ankle
Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstructions Metatarsal ligament/tendon repairs/reconstructions Flexor hullucis longus (FHL) Tendon transfers
#### Elbow, Wrist, and Hand
Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Scapholunate ligament reconstruction Tendon transfers Carpomedicarpal ioint arthroplasty
### TECHNOID ACA CHARACTERESTICS *
Smith & Nephew BIOSURE HEALICOIL interference screw is substantially equivalent in intended use and fundamental scientific technology to the legally marketed predicate devices (Smith and Nephew BIOSURE PK - K083226 & K083635 and BIOMET Bio-Core K042552) and raises no new issues of safety and efficacy. Smith & Nephew BlOSURE HEALICOIL interference screw and the predicate Smith & Nephew BIOSURE PK Interference screws (K083226 & K083635) are manufactured from the same PEEK material. Smith & Nephew BIOSURE HEALICOIL interference screw and the predicate Biomet Bio-core (K042552) have a similar open lattice construction to allow for bone ingrowth through the fenestrations to the central channel of the screws or to be packed with autologous bone.
### SUMMARK OF PERFORMANCE DATA
Mechanical testing data for pull out and insertion testing demonstrates the BIOSURE HEALICOIL PK interference screw is substantially equivalent to the currently marketed predicate devices.
The animal study evaluated bone ingrowth into the BIOSURE HEALICOIL PK interference screw (n=11) and control (n=11) via micro computed tomography (µCT) and histology. µCT demonstrated at twelve weeks that there was bone between the fenestrations and within the central cannulations. Histology andwsis
{3}------------------------------------------------
demonstrated that new bone formed within the BIOSURE HEALICOIL PK interference screw and control in the fenestration between the threads into the central channel in all specimens.
Note: Animal data is not necessarily indicative of human clinical outcomes. These results have not been demonstrated in humans having variety of bone quality based on specific disease states such as osteoporosis.
| SUBSTANTIAL EQUIVALENCE INFORMATION | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| and the control control control control control controller and the controlled to the contribution of the contribution of the contribution of the contribution of the contribut | | | |
The substantial equivalence of the BIOSURE HEALICOIL PK interference screw is based on similarities in indications for use, design features, operational principles, material composition, and performance to the predicate devices listed above. Based on the similarities to the BIOSURE HEALICOIL PK interference screw is substantially equivalent to its predicates.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread, with three lines representing the feathers.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2014
Smith & Nephew, Incorporated Vivek Mukhatyar, Ph.D. Regulatory Affairs Specialist II 150 Minuteman Road Andover, Massachusetts 01810
Re: K140879
Trade/Device Name: BIOSURE HEALICOIL PK Interference Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: May 23, 2014 Received: May 28, 2014
Dear Dr. Mukhatyar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading:
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Vivek Mukhatyar, Ph.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
#### 510(k) Number (if known)
# 140879
Device Name The BIOSURE HEALICOIL PK Interference Screw
Indications for Use (Describe)
The BIOSURE HEALICOL PK Interference Screw is indicated for the reattachment of ligament, tendon, soft tissue, or bone-tendon-bone for the following indications:
#### Knee
ACL repairs, PCL repairs, Extra-capsular repairs, Medial collateral ligament, Lateral ligament, Posterior oblique ligament, Patellar realignment and tendon repairs, Vastus medialis obliquus advancement, Iliotibial band tenodesis
#### Shoulder
Acromioclavicular separation repairs, Biceps tenodesis
#### Foot and Ankle
Medial or lateral instability repairs/reconstructions, Achilles tendon repairs/reconstructions, Metatarsal ligament/tendon repairs/reconstructions, Flexor hullucis longus (FHL), Tendon transfers
#### Elbow, Wrist, and Hand
Bicens tendon reattachment, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Scapholunate ligament reconstruction, Tendon transfers, Carpomedicarpal joint arthroplasty
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Casey L. Hanley; Ph.D.
# Division of Orthopedic Devices
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.