TWINFIX TI 3.5MM SUTURE ANCHOR WITH TWO ULTRABRAID SUTURE #2 WITH NEEDLES, SL

K123425 · Smith & Nephew, Inc. · MBI · Feb 20, 2013 · Orthopedic

Device Facts

Record IDK123425
Device NameTWINFIX TI 3.5MM SUTURE ANCHOR WITH TWO ULTRABRAID SUTURE #2 WITH NEEDLES, SL
ApplicantSmith & Nephew, Inc.
Product CodeMBI · Orthopedic
Decision DateFeb 20, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

Smith & Nephew TWINFIX Ti Preloaded Suture Anchor with Needles, SL are intended for use only for the fixation of non-absorbable, synthetic, surgical suture material for the following indications. Shoulder: Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsulolabral reconstructions, Biceps tenodesis, and Deltoid Repairs. Foot and Ankle: Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions. Elbow, Wrist and Hand: Scapholunate ligament reconstruction, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment. Knee: Extra-capsular repairs: Medial collateral ligament repair, Lateral collateral ligament repair, Posterior oblique ligament, Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement.

Device Story

Titanium alloy suture anchor with two non-absorbable sutures and stainless steel needles; used for soft tissue-to-bone reattachment. Surgeon-operated via preferred technique using provided insertion device (stainless steel shaft, ABS/polycarbonate handle). Provides secure fixation for various orthopedic repairs (shoulder, foot/ankle, elbow/wrist/hand, knee). Benefits patient by enabling stable tissue reattachment during surgical reconstruction.

Clinical Evidence

Bench testing only. Mechanical test data provided for insertion and pull-out strength to demonstrate performance equivalence.

Technological Characteristics

Titanium alloy suture anchor; non-absorbable sutures (braided, silicone/PTFE impregnated polyester, or UHMW polyethylene/polypropylene/nylon co-braid); stainless steel needles; insertion device with stainless steel shaft and ABS/polycarbonate handle.

Indications for Use

Indicated for fixation of non-absorbable, synthetic, surgical suture material in shoulder, foot, ankle, elbow, wrist, hand, and knee soft tissue-to-bone repairs.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### Smith & Nephew, Inc. TWINFIX Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL Traditional Premarket Notification | 510(k) Summary | Date Prepared: February 1, 2013 | | |------------------------------------------|--------------------------------------|--| | Submitter Information | Contact Information | | | Smith & Nephew, Inc., Endoscopy Division | Catherine Kilshaw | | | 150 Minuteman Road | Senior Regulatory Affairs Specialist | | | Andover, MA 01810 | Phone: (508) 337-4104 | | | | Fax: (978) 749-1443 | | | Device Name (Unmodified) | | |---------------------------|---------------------------------------------------------------------------------| | Trade or proprietary name | TWINFIX® Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 and<br>Needles, SL | | Common or usual name | Soft Tissue Fixation Device (MBI) | | Classification name | 21 CFR 888.3040 Fastener, fixation, nondegradable, soft tissue<br>(Class II) | ## Legally Marketed Predicate Device The Smith & Nephew TWINFIX® Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL is substantially equivalent in intended use and Fundamental Scientific Technology to the following legally marketed devices in commercial distribution: - K053344 Bioraptor Suture Anchor, TWINFIX Ti 2.8mm & TWINFIX 3.5mm (Cleared February 23, 2006) - K972326 PEBA Anchor/Suture Combination (Cleared February 6, 1998) #### Device Description . . 、「お気になる」ということです。 この時間のお気に The Smith & Nephew TWINFIX* Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL are designed to provide secure reattachment of soft tissue to bone. Attachment of soft tissue is performed by the surgeon's preferred technique. TWINFIX Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL consists of a titanium alloy suture anchor and non-absorbable sutures with attached stainless steel needles. The non-absorbable sutures are offered either braided, silicone or PTFE impregnated, polyester (USP) or braided, uncoated, UHMW polyethylene and UHMW polyethylene with monofilament polypropylene or nylon co-braid. An insertion device is also offered which contains a stainless steel shaft with a ABS and polycarbonate handle. The TWINFIX Ti is designed to provide secure reattachment of soft tissue to bone. Attachment of soft tissue is performed by the surgeon's preferred technique. {1}------------------------------------------------ ## K123425 #### Smith & Nephew, Inc. TWINFIX® Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL Traditional Premarket Notification ### Indications for Use - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Smith & Nephew TWINFIX® Ti Preloaded Suture Anchor with Needles, SL are intended for use only for the fixation of non-absorbable, synthetic, surgical suture material for the following indications. #### Shoulder Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsulolabral reconstructions, Biceps tenodesis, and Deltoid Repairs #### Foot and Ankle Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions #### Elbow. Wrist and Hand Scapholunate ligament reconstruction, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment #### Knee Extra-capsular repairs: Medial collateral ligament repair, Lateral ligament repair, Posterior oblique ligament, lliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement #### Technological Characteristics ------------------------------------------------------------------------------------------------------------------------------------------------11. 11. 11. 12. The Smith & Nephew TWINFIX® Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL is substantially equivalent in design and fundamental scientific technology to the defined predicate devices and raise no new issues of safety and efficacy. #### Performance Data Mechanical test data demonstrates the device has met the performance specifications for insertion and pull out strength and therefore, is considered substantially equivalent to the currently marketed predicate devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "HEALTH & HUMAN SERVICES - USA" is arranged in a semi-circle around the top of the caduceus. The word "DEPART" is at the bottom of the logo. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 20, 2013 Smith & Nephew, Inc., Endoscopy Division . % Ms. Catherine Kilshaw Senior Regulatory Affairs Specialist 150 Minuteman Road Andover, Massachusetts 01810 Re: K123425 Trade/Device Name: TWINFIX™ Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: November 6, 2012 Received: November 26, 2012 Dear Ms. Kilshaw: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ ## Page 2 -- Ms. Catherine Kilshaw device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Erin . Keith Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Smith & Nephew, Inc. TWINFIX Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL Traditional Premarket Notification #### Indications for Use #### 510(k) Number (if known): K123425 Device Name: TWINFIX* Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 with Needles, SL #### Indications for Use: Smith & Nephew TWINFIX Ti Preloaded Suture Anchor with Needles, SL are intended for use only for the fixation of non-absorbable, synthetic, surgical suture material for the following indications. #### Shoulder Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsulolabral reconstructions, Biceps tenodesis, and Deltoid Repairs #### Foot and Ankle Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions #### Elbow, Wrist and Hand Scapholunate ligament reconstruction, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment #### Knee Extra-capsular repairs: Medial collateral ligament repair, Lateral collateral ligament repair, Posterior oblique ligament, Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement Prescription Use _______ X AND/OR AND/OR (Per 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 807 Subpart C) IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anton E. Dmitriev, PhD Division of Orthopedic Device 2013.02.15 14:59:29-05'00'
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...