K113294 · Smith & Nephew, Inc. · MBI · Jan 20, 2012 · Orthopedic
Device Facts
Record ID
K113294
Device Name
SMITH & NEPHEW HEALICOIL PK SUTURE ANCHOR
Applicant
Smith & Nephew, Inc.
Product Code
MBI · Orthopedic
Decision Date
Jan 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Smith & Nephew HEALICOIL PK Suture Anchor is intended for use for the reattachment of soft tissue to bone for the following indications: Shoulder: Bankart lesion repairs, Slap lesion repairs, Capsular shift or capsulolabral reconstructions, Acromioclavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs, Biceps tenodesis; Knee: Extra-capsular repairs: Medial collateral ligament, Lateral collateral ligament, Posterior oblique ligament, Patellar realignment and tendon repairs: Vastus medialis obliquous advancement, Iliotibial band tenodesis; Foot & Ankle: Hallux valgus repairs, Medial or lateral instability repairs/reconstructions, Achilles tendon repairs/reconstruction, Midfoot reconstructions, Metatarsal ligament/tendon repairs/reconstructions; Elbow: Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment; Hip: Gluteal tendon repairs - Gluteus medius and gluteus minimus repair.
Device Story
HEALICOIL PK Suture Anchor is a screw-in orthopedic fixation device; manufactured from PEEK; preassembled on stainless steel inserter; pre-loaded with suture. Used by surgeons in clinical settings for soft tissue-to-bone reattachment. Device features open architecture with fenestrations between threads and central channel to facilitate bone ingrowth. Clinical benefit derived from secure fixation of soft tissue to bone during reconstructive or repair surgeries. Device is provided sterile.
Clinical Evidence
No human clinical data. Bench testing only. Animal study (n=12 sheep) evaluated bone ingrowth into the anchor at 12 weeks post-implantation using micro-CT and histology. Results showed new bone formation within fenestrations and central channel, with 21.9% of implant volume consisting of new bone compared to 34.5% in control bone.
Technological Characteristics
Material: PEEK (polyetheretherketone). Form factor: Screw-in anchor, 4.5mm and 5.5mm diameters. Features: Open architecture with fenestrations and central channel. Delivery: Preassembled on stainless steel inserter, pre-loaded with suture. Sterilization: Sterile.
Indications for Use
Indicated for reattachment of soft tissue to bone in shoulder, knee, foot/ankle, elbow, and hip procedures. Includes repairs for Bankart, SLAP, rotator cuff, biceps, collateral ligaments, patellar realignment, hallux valgus, Achilles tendon, and gluteal tendons.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Smith & Nephew Next Generation Fully Threaded PK Suture Anchor (K110545)
Submission Summary (Full Text)
{0}------------------------------------------------
K113294
ם מוס הנחמו המועד שנים בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בין בי שיר מתוך מסוג מסוג מוסטו NI 10000 MODITA MUNITA 10000000 ជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជាជ 来宾申博官网 AFFECT FOR FOR THE FOR FOR THE FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FILE FOR FILE FOR FILE FOR FILE FOR FIL ENDING WATER BEAUTHERS BETHER FREE THE WARD COMMENT CONSTITUTION CONTENTION CONTENTION CONTENTION CONTENTION CONTENTIONAL 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 משפט מתחק מתמועד מתמוני מתנועד ש
## SECTION IV
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
As required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based.
# Smith & Nephew HEALICOIL PK Suture Anchor
Date Prepared: November 7, 2011
### A. Submitter's Name:
Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road, Andover, MA 01810
## B. Company Contact
Elizabeth Lavelle Senior Regulatory Affairs Specialist Phone: (508) 261-3607
### C. Device Name
| Trade Name: | HEALICOIL PK Suture Anchor |
|----------------------|-------------------------------------------------|
| Common Name: | Suture Anchor |
| Classification Name: | Fastener. fixation, non-degradable, soft tissue |
## D. Predicate Devices
The Smith & Nephew HEALICOIL PK Suture Anchor is substantially equivalent in intended use and fundamental scientific technology to the following legally marketed device in commercial distribution: Smith & Nephew Next Generation Fully Threaded PK Suture Anchor (K110545).
#### Description of Device E.
The HEALICOIL PK Suture Anchor is manufactured from PEEK (polyetheretherketone) and is offered in diameters of 4.5mm and 5.5mm sizes. The screw-in anchor is preassembled onto a stainless steel inserter and pre-loaded with up to three strands of suture.
#### F. Intended Use
The Smith & Nephew HEALICOIL PK Suture Anchor is intended for use for the reattachment of soft tissue to bone for the following indications:
{1}------------------------------------------------
Kff3294
| Shoulder: | Knee: |
|----------------------------------------------------------|-----------------------------------------------------------------------|
| Bankart lesion repairs | Extra-capsular repairs: |
| Slap lesion repairs | Medial collateral ligament |
| Capsular shift or capsulolabral<br>reconstructions | Lateral collateral ligament |
| Acromioclavicular separation repairs | Posterior oblique ligament |
| Deltoid repairs | Patellar realignment and tendon repairs: |
| Rotator cuff tear repairs | Vastus medialis obliquous advancement |
| Biceps tenodesis | Iliotibial band tenodesis. |
| Foot & Ankle: | Elbow: |
| Hallux valgus repairs | Ulnar or radial collateral ligament reconstructions |
| Medial or lateral instability<br>repairs/reconstructions | Lateral epicondylitis repair |
| Achilles tendon repairs/reconstruction | Biceps tendon reattachment |
| Midfoot reconstructions | Hip: |
| Metatarsal ligament/tendon<br>repairs/reconstructions | Gluteal tendon repairs<br>- Gluteus medius and gluteus minimus repair |
## G. Comparison of Technological Characteristics
The Smith & Nephew HEALICOIL PK Suture Anchor is substantially equivalent in design, materials, technological characteristics, intended use, and indications for use to its currently marketed predicate device, the Smith & Nephew Next Generation Fully Threaded PEEK Suture Anchor.
## H. Summary Performance Data
There have been no design modifications that required additional mechanical testing. Performance characteristics of the anchors have not changed from those described in K110545.
The purpose of this submission was to update the labeling to include results from an animal study. The animal study evaluated bone growth into the HEALICOIL PK Suture Anchor and a control suture anchor via micro computed tomography (micro CT) and histology. The labeling updates are as follows:
# (1) The open architecture of the HEALICOIL PK Suture Anchor allows for new bone to fill the fenestrations between threads and into the central channel.
HEALICOIL PK suture anchors (5.5 x 20mm) were implanted into 4.5 x 20mm sites created in ovine cancellous bone (n=12) of the medial distal femur and evaluated by micro-CT and histology analysis 12 weeks post implantation. Micro-CT and histology analysis demonstrated that new bone formed within the HEALICOIL PK suture anchor in the fenestrations between the threads and into the central channel in all specimens.
Note: Animal data is not necessarily indicative of human clinical outcomes. These results have not been demonstrated in humans having a variety of bone quality based on specific disease states such as osteoporosis. The effect of formation of new bone on pullout strength was not shown.
{2}------------------------------------------------
K113294
(2) The Smith and Nephew HEALICOIL PK Suture Anchor contained 63% of the bone volume of control bone by 12 weeks post-implantation in sheep.
HEALICOIL PK suture anchors (5.5 x 20mm) were implanted into 4.5 x 20mm sites created in ovine cancellous bone of the medial distal femur and compared to control intact bone of the same anatomical site at 12 weeks post implantation. Micro-CT analysis of new bone formed within the HEALICOIL PK suture anchor demonstrated that 21.9% of the implant volume was new bone by 12 weeks post-implantation (n=12). Micro-CT analysis of a 4.5 x 20mm volume of control bone from the same anatomical site demonstrates that intact bone has an average bone volume of 34.5% (n=12).
Note: Animal data is not necessarily indicative of human clinical outcomes. These results have not been demonstrated in humans having a variety of bone quality based on specific disease states such as osteoporosis. The effect of formation of new bone on pullout strength was not shown.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, depicted with stylized lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Smith & Nephew Incorporated % Ms. Elizabeth Lavelle Senior Regulatory Affairs Specialist 150 Minuteman Road Andover. Massachusetts 01810
JAN 2 0 2012
K113294 Re:
Trade/Device Name: Smith & Nephew HealiCoil PK Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: January 5, 2012 Received: January 6, 2012
Dear Ms. Lavelle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Elizabeth Lavelle
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/defaulthtm.
Sincerely yours,
Mark N. Melkerson
Director
Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
K113294
510(k) Number (if known):
Device Name: HEALICOIL PK Suture Anchor (formerly Next Generation Fully Threaded Suture Anchor)
Indications For Use:
.
The HEALICOIL PK Suture Anchor (formerly Smith & Nephew Next Generation Fully Threaded PEEK Suture Anchor) is intended for use for the reattachment of soft tissue to bone for the following indications:
| Shoulder: | Knee: |
|-------------------------------------------------|------------------------------------------|
| Bankart lesion repairs | Extra-capsular repairs: |
| Slap lesion repairs | Medial collateral ligament |
| Capsular shift or capsulolabral reconstructions | Lateral collateral ligament |
| Acromioclavicular separation repairs | Posterior oblique ligament |
| Deltoid repairs | Patellar realignment and tendon repairs: |
| Rotator cuff tear repairs | Vastus medialis obliquous advancement |
| Biceps tenodesis | Iliotibial band tenodesis. |
Foot & Ankle: Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstruction Midfoot reconstructions · Metatarsal ligament/tendon repairs/reconstructions
Elbow: Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Biceps tendon reattachment
Hip: Glutea! tendon repairs - Gluteus medius and gluteus minimus repair
Prescription Use x (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113294
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.