← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K112991

# MODIFIED ROG SUTURE ANCHOR (K112991)

_Rog Sports Medicine, Inc. · MBI · Dec 8, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K112991

## Device Facts

- **Applicant:** Rog Sports Medicine, Inc.
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** Dec 8, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The RoG™ 5.5 mm Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows: Shoulder indications :- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis. Wrist/hand indications:- Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction. Foot/ankle indications:- Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions. Elbow indications: Biceps tendon reconstruction, ulnar or radial collateral . ligament reconstruction, lateral epicondylitis repair. Knee indications: Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure. The RoG™ 2.9mm Suture Anchor is indicated for use in soft tissue reattachment procedures. Specific indications are as follows: Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis. . Wrist/hand indications: - Scapholunate ligament reconstruction. . Elbow indications: - Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction. Knee indications: Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure, extracapsular repair, vastus medialis obliquus (VMO) muscle advancement. The RoGTM 5.5 mm Knotless Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows: . Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis. Wrist/hand indications: Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction. Foot/ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions. . Elbow indications: - Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair. . Knee indications: - Lateral collateral ligament repair, medial collateral ligament repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure.

## Device Story

Screw-like suture anchor; composed of PEEK plastic; available in 5.5mm and 2.9mm diameters; standard (knotted) and knotless configurations; supplied with non-absorbable polyethylene suture; includes reusable taps and guides. Used by surgeons in orthopedic soft tissue reattachment procedures. Device provides mechanical fixation of soft tissue to bone; benefits patient through secure tissue reattachment during healing. Modification limited to expiration date extension.

## Clinical Evidence

No clinical data. Bench testing only; stability testing performed to support expiration date extension.

## Technological Characteristics

Material: PEEK (ISO 10993 compliant). Form factor: Screw-like, 5.5mm/2.9mm diameter, 10mm/17mm length. Sterilization: ISO 11135 compliant. Non-absorbable polyethylene suture meets USP requirements.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- RoGTM Suture Anchor ([K111590](/device/K111590.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY

#### Premarket Notification: K112991

December 8, 2011

(1/3)

DEVICE: RoGTM Sports Medicine Suture Anchor

#### SPONSOR/MANUFACTURER:

RoG Sports Medicine, Inc. 16450 S. 104th Ave. Orland Park, IL 60467

### SUMBITTER/REGULATORY CONTACT:

Curtis Raymond Orchid Design 80 Shelton Technology Center Shelton, CT 06484

#### FDA ESTABLISHMENT REGISTRATION NUMBER: (pending)

#### TRADE NAME, COMMON NAME, CLASSIFICATION:

TRADE NAME: Modified RoG™ Suture Anchor

COMMON NAME: Suture Anchor

CLASSIFICATION: Class II (ref.: 21 CFR 888.3040); Product Code MBI

#### PREDICATE DEVICE(S):

- K111590 RoGTM Suture Anchor .
#### DESCRIPTION OF SUBJECT DEVICE:

The subject device is screw-like in shape and composed exclusively of PEEK plastic. It is available in both standard ("knotted") and "knotless" configurations. It is also available in diameters of 5.5mm and 2.9mm and lengths of 10mm and 17mm. It is provided sterile and supplied with non-absorbable polyethylene suture. The anchor is supplied with reusable taps and guides of corresponding size.

#### INTENDED USE:

The RoG™ 5.5 mm Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:

- Shoulder indications :- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.

{1}------------------------------------------------

- Wrist/hand indications:- Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction.
- Foot/ankle indications:- Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- Elbow indications: Biceps tendon reconstruction, ulnar or radial collateral . ligament reconstruction, lateral epicondylitis repair.
- Knee indications: Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure.

The RoG™ 2.9mm Suture Anchor is indicated for use in soft tissue reattachment procedures. Specific indications are as follows:

- Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- . Wrist/hand indications: - Scapholunate ligament reconstruction.
- . Elbow indications: - Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction.
- Knee indications: Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure, extracapsular repair, vastus medialis obliquus (VMO) muscle advancement.

The RoGTM 5.5 mm Knotless Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:

- . Shoulder indications:- Bankart repair, rotator cuff repair, SLAP lesion repair, capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/hand indications: Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction.
- Foot/ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- . Elbow indications: - Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair.
- . Knee indications: - Lateral collateral ligament repair, medial collateral ligament repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure.

#### PERFORMANCE CHARACTERISTICS:

Performance characteristics of the anchors have not changed from those described in K111590. Sutures supplied with the device meet the requirements of the U.S. Pharmacopeia for non-absorbable suture as well as all requirements of Class II Special

{2}------------------------------------------------

Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA; June 3, 2003"

#### SAFETY CHARACTERISTICS:

The device is composed exclusively of polyetheretherketone (PEEK). A Master File demonstrating safety of the material has been supplied by the PEEK supplier and shows compliance to the requirements of ISO 10993. Sterilization of the device is in compliance with ISO 11135. Sterilant residues are within the limits of ISO 10993-7.

#### CONCLUSION(S):

The subject device has the same design considerations, assembly configurations, performance characteristics and indications for use as the predicate device. The proposed modification consists solely of extending the current expiration date. Testing has shown the anchor, suture and packaging to be stable for the proposed period.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

RoG Sports Medicine, Inc. % Orchid Design Mr. Curtis Raymond 80 Shelton Technology Center Shelton, Connecticut 06484

DEC- = 8 2011

Re: K112991

Trade/Device Name: RoGTM Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: November 7, 2011 Received: November 8, 2011

Dear Mr. Raymond:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Curtis Raymond

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

3

## Indications for Use

510(k) Number (if known): K

Device Name: RoGTM Suture Anchor

The RoG™ 5.5 mm Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:

- . Shoulder indications: - Bankart repair, rotator cuff repair, SLAP lesion repair, capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/hand indications: Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction.
- Eoot/Ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- . Elbow indications: - Biceps tendon reconstruction, ulnar or radial collateral ligament reconstruction, lateral epicondylitis repair.
- Knee indications: Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure.

The RoG 2.9mm Suture Anchor is indicated for use in soft tissue reattachment procedures. Specific indications are as follows:

- Shoulder indications: Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/hand indications: Scapholunate ligament reconstruction. .
- Elbow indications: Biceps tendon reconstruction, ulnar or radial collateral . ligament reconstruction.
- Knee indications: Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure, extracapsular repair, vastus medialis obliquus (VMO) muscle advancement.

The RoG 5.5 mm Knotless Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, elbow, wrist/hand, foot/ankle and knee. Specific indications are as follows:

- Shoulder indications: Bankart repair, rotator cuff repair, SLAP lesion repair, . capsule repair or capsulolabral reconstruction, acromio-clavicular separation, deltoid repair, biceps tenodesis.
- Wrist/Hand indications: Ulnar/Radial collateral ligament reconstruction, . scapholunate ligament reconstruction.

{6}------------------------------------------------

- Foot/Ankle indications: Achilles tendon repair/reconstruction, hallax valgus . reconstruction, lateral stabilization, medial stabilization, mid- and forefoot reconstructions.
- Elbow indications: Biceps tendon reconstruction, ulnar or radial collateral . ligament reconstruction, lateral epicondylitis repair.
- Knee indications: Lateral collateral ligament repair, medial collateral ligament . repair, posterior oblique ligament repair, patellar ligament/tendon repair, iliotibial band tenodesis, joint capsule closure.

Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Michael Owens to MXM

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K112991

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K112991](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K112991)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
