5.5 / 6.5MM SPARTAN PEEK SURTURE IMPLANT WITH #2 MAGNUMWIRE / WITH NEEDLES, EXTRACTION TOOL, 5.5 / 6.5MM PUNCH TAP
K110781 · Arthrocare Corp. · MBI · May 27, 2011 · Orthopedic
Device Facts
Record ID
K110781
Device Name
5.5 / 6.5MM SPARTAN PEEK SURTURE IMPLANT WITH #2 MAGNUMWIRE / WITH NEEDLES, EXTRACTION TOOL, 5.5 / 6.5MM PUNCH TAP
Applicant
Arthrocare Corp.
Product Code
MBI · Orthopedic
Decision Date
May 27, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The ArthroCare Spartan PEEK Suture Implant is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon reattachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions
Device Story
ArthroCare Spartan PEEK Suture Implant System consists of fully-threaded, corkscrew-shaped PEEK suture anchor (5.5mm diameter) preconfigured with MagnumWire sutures; mounted on disposable delivery driver. Associated instruments (Punch Tap, Extraction Tool) used for implantation/removal. Device used by surgeons in clinical/OR settings for soft tissue-to-bone fixation. Implant provides mechanical anchor point for sutures to secure soft tissue during orthopedic repairs (e.g., rotator cuff, ligament reconstruction).
Clinical Evidence
Bench testing only. In vitro testing performed involving insertion of anchors in simulated human bone substrate, including ultimate strength and cyclic loading tests.
Technological Characteristics
Material: PEEK (polyether-etherketone). Form factor: Fully-threaded, corkscrew-shaped anchor, 5.5mm diameter. Delivery: Disposable driver, preconfigured with MagnumWire sutures. Sterile supply.
Indications for Use
Indicated for soft tissue to bone fixation in shoulder, ankle, foot, elbow, and knee orthopedic procedures requiring suture anchoring.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
ArthroCare 5.5mm Spartan PEEK Suture Implant System (K110164)
ArthroCare 5.5mm Spartan PEEK Suture Implant System (K102262)
Related Devices
K102262 — ARTHOCARE SPARTAN PEEK SUTURE IMPLANT SYSTEM; PUNCH TAP ; EXTRACTION TOOL · Arthrocare Corp. · Oct 21, 2010
K180464 — AEVUMED PHANTOM Suture Anchors · Aevumed, Inc. · May 31, 2018
K242895 — Aevumed PHANTOM Suture Anchors · Aevumed, Inc. · Aug 6, 2025
K260004 — Aevumed PROTEKT Suture Anchor · Aevumed, Inc. · Jan 28, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
K110781 (112)
# MAY 2 7 2011
#### 510(k) Summary
# ArthroCare Corporation Spartan™ PEEK Suture Implant System
### General Information
| Submitter Name/Address: | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-2936 |
|------------------------------------|---------------------------------------------------------------------------|
| Establishment Registration Number: | 2951580 |
| Contact Person: | Valerie Defiesta-Ng<br>Director, Regulatory Affairs |
| Date Prepared: | March 18, 2011 |
**Device Description**
| Trade Name: | ArthroCare Spartan PEEK Suture Implant System |
|------------------------|-------------------------------------------------------------------------|
| Generic/Common Name: | Screw, Fixation, Bone |
| Classification Name: | Smooth or Threaded Metallic Bone Fixation<br>Fastener (21 CFR 888.3040) |
| Device Classification: | Class II, 21 CFR 888.3040<br>Product Code MBI |
**Predicate Device**
| ArthroCare 5.5mm Spartan PEEK Suture<br>Implant System | K110164 (February 7, 2011) |
|--------------------------------------------------------|----------------------------|
| ArthroCare 5.5mm Spartan PEEK Suture<br>Implant System | K102262 (October 21, 2010) |
## Product Description
The ArthroCare Spartan PEEK (polyether-etherketone) Suture Implant is a fullythreaded, corkscrew shape anchor available in 5.5mm diameter size. The suture anchor comes preconfigured with MagnumWire® sutures for traditional knot tying and is mounted on a disposable delivery driver. The device is supplied sterile and is available with our without needles. Associated instruments (Punch Tap and Extraction Tool) for implantation and removal of the implant are available separately and altogether are referred to as the ArthroCare Spartan PEEK Suture Implant System.
{1}------------------------------------------------
## Intended Uses/Indications for Use
The ArthroCare Spartan PEEK Suture Implant is indicated for use in fixation of soft tissue to bone. Examples of such procedures include:
Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair
Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction
Foot: Hallux valgus reconstruction
Elbow: Tennis elbow repair, biceps tendon reattachment
Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions
### Substantial Equivalence
In establishing substantial equivalence to the predicate. ArthroCare compared the intended use, device design, technology, and device components of the subject device with the predicate device. Bench performance testing has been completed to demonstrate the substantial equivalence of the ArthroCare Spartan PEEK Suture Implant in accordance with FDA Guidance Document Testing for Bone Anchors. The in vitro testing performed involved insertion of the anchors in a simulated human bone substrate followed by both ultimate strength and cyclic loading tests. The performance testing and device comparison demonstrate that the subject device is substantially equivalent to the predicate device, and is safe and effective for its intended use.
#### Summary of Safety and Effectiveness
The proposed modifications to the Spartan PEEK Suture Implant System are not substantial changes, and do not significantly affect the safety or efficacy of the proposed device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure with three curved lines representing wings or snakes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
ArthoCare Corporation % Ms. Valerie Defiesta-Ng Director, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-2936
MAY 2 7 2011
Re: K110781
Trade/Device Name: ArthoCare® Spartan™ PEEK Suture Implant System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: May 13, 2011 Received: May 16, 2011
Dear Ms. Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Valerie Defiesta-Ng
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
E.Id Keith
for Mark
Director
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 110781 (111)
# Indications for Use
510(k) Number (if known):
Device Name:
ArthroCare® Spartan™ PEEK Suture Implant System
Indications For Use:
The ArthroCare Spartan PEEK Suture Implant is indicated for use in fixation of soft tissue to bone. Examples of such procedures include:
Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair
Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction
Foot: Hallux valgus reconstruction
Elbow: Tennis elbow repair, biceps tendon reattachment
Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Eiel Keitt
(Division Sign-Off) (Division Sign-On)
Division of Surgical, Orthopedic, Division of Son of Societive Devices
K110781
510(k) Number
Page 1 of
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.