← Product Code [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI) · K070758

# STRYKER PEEK ZIP SUTURE ANCHOR (K070758)

_Stryker Endoscopy · MBI · May 16, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K070758

## Device Facts

- **Applicant:** Stryker Endoscopy
- **Product Code:** [MBI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI.md)
- **Decision Date:** May 16, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Stryker PEEK Zip Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Capsular Shift/Capsulolabral Reconstruction, Biceps Tenodesis, Deltoid Repair. Knee: Extra Capsular Repairs (Medial Collateral Ligament, Lateral Collateral Ligament, Posterior Oblique Ligament), Illiotibial Band Tenodesis, Patellar Tendon Repair. Elbow, Wrist, Hand: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Biceps Tendon Reattachment. Foot and Ankle: Medial Instability Repair/Reconstruction, Lateral Instability Repair/Reconstruction, Achilles Tendon Repair/Reconstruction, Midfoot Reconstruction, Hallux Valgus Reconstruction. Pelvis: Bladder Neck Suspension Procedures. The Stryker PEEK Zip Suture Anchor is intended for single-use only.

## Device Story

Single-use, screw-in soft tissue anchor; manufactured from PEEK-OPTIMA polymer; pre-threaded with non-absorbable braided polyester or polyethylene sutures; pre-assembled on disposable inserter. Used by surgeons in OR during reconstructive procedures to secure soft tissue to bone. Device provides mechanical fixation; surgeon manually inserts anchor into bone to reattach tendons or ligaments. Benefits include secure tissue-to-bone attachment to facilitate healing in orthopedic and pelvic reconstructive surgeries.

## Clinical Evidence

Bench testing only. Device validated to SAL of 10^-6 using ethylene oxide; EtO residuals tested per ISO 10993-7:1995. Conformance to voluntary safety and performance standards including ISO 10993-1, EN 550, EN 556-1, EN 11607-1/2, EN 980, EN 1041, and EN ISO 14971.

## Technological Characteristics

Material: PEEK-OPTIMA (polyetheretherketone). Form factor: Screw-in anchor with pre-threaded non-absorbable braided sutures on disposable inserter. Sterilization: Ethylene oxide (EtO) to SAL 10^-6. Standards: ISO 10993-1, ISO 10993-7, EN 550, EN 556-1, EN 11607, EN 980, EN 1041, EN ISO 14971.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Stryker 5.0mm BioZip Suture Anchor ([K023192](/device/K023192.md), [K041305](/device/K041305.md))
- Arthrex 5.0mm PEEK Corkscrew FT ([K061665](/device/K061665.md))

## Reference Devices

- USP braided polyester suture ([K953531](/device/K953531.md))
- USP braided polyethylene suture ([K033654](/device/K033654.md))
- USP braided polyethylene suture ([K040472](/device/K040472.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 16 2007

Image /page/0/Picture/3 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and there is a registered trademark symbol to the right of the "r". The letters are closely spaced, giving the word a compact appearance. The overall impression is a clean and modern logo.

Endoscopy

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Date: March 19, 2007

## Contact Person:

Erica A. Walters, RAC Sr. Regulatory Representative 408-754-2078(phone) 408-754-2521 (fax) erica.walters@stryker.com

## Device Name:

| Proprietary Name:      | Stryker PEEK Zip Suture Anchor                                                                                                            |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Common and Usual Name: | PEEK Suture Anchor                                                                                                                        |
| Classification Name:   | Screw, Fastener, Fixation, Nonabsorbable, Bone, Soft<br>Tissue (Class II, 21 CFR 888.3040, Product Code MBI,<br>Orthopedics Review Panel) |

## Predicate Devices:

Stryker 5.0mm BioZip Suture Anchor: # K023192, K041305 Arthrex 5.0mm PEEK Corkscrew FT: #K061665

## Device Description and Intended Use:

The Stryker PEEK Zip Suture Anchor is a single-use, soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as:

#### Shoulder:

- Rotator Cuff Repair
- Bankart Repair ■
- SLAP Lesion Repair
- Acromio-Clavicular Separation Repair
- 제 Capsular Shift/Capsulolabral Reconstruction
- Biceps Tenodesis
- Deltoid Repair.

#### Knee:

- Extra Capsular Repairs
	- Medial Collateral Ligament ಂ
	- Lateral Collateral Ligament o
	- Posterior Oblique Ligament ు
- ﻻ Illiotibial Band Tenodesis
- Patellar Tendon Repair. 피

#### Elbow, Wrist, Hand:

- Scapholunate Ligament Reconstruction
- Ulnar Collateral Ligament Reconstruction
- Radial Collateral Ligament Reconstruction
- . Biceps Tendon Reattachment.

#### Foot and Ankle:

- Medial Instability Repair/Reconstruction
- Lateral Instability Repair/Reconstruction
- Achilles Tendon Repair/Reconstruction
- Midfoot Reconstruction
- Hallux Valgus Reconstruction.

#### Pelvis:

- Bladder Neck Suspension Procedures.

{1}------------------------------------------------

The Stryker PEEK Zip Suture Anchor will be manufactured from PEEK-OPTIMA® (polyetheretherketone), a biocompatible polymer manufactured by Invibio Inc. The screw-in anchor is pre-threaded with non-absorbable braided sutures and pre-assembled on a The suture will be offered in both non-absorbable USP braided polyester disposable inserter. suture (#K953531) and non-absorbable USP braided polyethylene suture (#K033654 and #K040472.). The Stryker PEEK Zip Suture Anchor will be validated to a SAL of 10° using ethylene oxide. The EtO residuals will be tested according to ISO 10993-7:1995.

Prior to introducing the Stryker PEEK Zip Suture Anchor to market, the device will conform to the following voluntary safety and performance standards: ISO 10993-1, Blue Book Memorandum G95-1, EN 550, EN 556-1, EN 11607-1, EN 11607-2, EN 980, EN 1041, and EN ISO 14971.

The material differences between the Stryker PEEK Zip Suture Anchor and Stryker BioZip Suture Anchor (K023192, K041305) do not affect the safety and efficacy of the product; therefore, the Stryker PEEK Zip Suture Anchor is considered substantially equivalent in performance, intended use, safety, and efficacy to the Stryker BioZip Suture Anchor.

The Stryker PEEK Zip Suture Anchor is considered substantially equivalent in material composition, intended use, safety and efficacy to the Arthrex PEEK Corkscrew FT (K061665).

By: Chutine Kysar

Date: 19 March 2007

Christine Krueger Design Engineer Stryker Endoscopy

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved strokes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Stryker Endoscopy % Erica A. Walters, RAC Sr. Regulatory Representative 5900 Optical Court San Jose, California 95138

MAY 16 2007

Re: K070758

> Trade/Device Name: Stryker PEEK Zip Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: March 19, 2007 Received: April 3, 2007

Dear Ms. Walters:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

# Page 2 - Erica A. Walters, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours,

Karlaue Buehup
fell

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### INDICATIONS FOR USE

Device Name: Stryker PEEK Zip Suture Anchor

K070758 510(k) Number if known:

The Stryker PEEK Zip Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as:

#### Shoulder:

- Rotator Cuff Repair «
- Bankart Repair .
- . SLAP Lesion Repair
- Acromio-Clavicular Separation Repair .
- . Capsular Shift/Capsulolabral Reconstruction
- 프 Biceps Tenodesis
- I Deltoid Repair.

#### Knee:

- Extra Capsular Repairs .
	- Medial Collateral Ligament o
	- Lateral Collateral Ligament 0
	- Posterior Oblique Ligament 0
- . Illiotibial Band Tenodesis
- I Patellar Tendon Repair.

#### Elbow. Wrist. Hand:

- Scapholunate Ligament Reconstruction .
- Ulnar Collateral Ligament Reconstruction
- · Radial Collateral Ligament Reconstruction
- Biceps Tendon Reattachment. .

#### Foot and Ankle:

- Medial Instability Repair/Reconstruction I
- · Lateral Instability Repair/Reconstruction
- · Achilles Tendon Repair/Reconstruction
- · Midfoot Reconstruction
- · Hallux Valgus Reconstruction.

#### Pelvis:

- Bladder Neck Suspension Procedures. .
The Stryker PEEK Zip Suture Anchor is intended for single-use only.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbara Buchup

Division of General. Restorative. and Neurological Devices

**510(k) Number** K0

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K070758](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBI/K070758)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
