← Product Code [MBH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH) · K241597

# Freedom® Total Knee System - Porous Tibial Base Plate (K241597)

_Maxx Orthopedics, Inc. · MBH · Feb 13, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH/K241597

## Device Facts

- **Applicant:** Maxx Orthopedics, Inc.
- **Product Code:** [MBH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH.md)
- **Decision Date:** Feb 13, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3565
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Freedom® Total Knee System is indicated for the following: • Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. • Correction of functional deformities. • Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. • Moderate valgus, varus, or flexion trauma. • Knee fractures untreatable by other methods. • Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components and Primary PCK Components only). The Freedom® Porous Tibial Base Plate and Cementless Femoral Components are indicated for Cemented or Uncemented use. All other components are indicated for cemented use only.

## Device Story

Freedom® Porous Tibial Base Plate is a line extension of the Freedom® Total Knee System for total knee arthroplasty. Device is an additively manufactured (Ti-6Al-4V ELI) tibial base plate with a porous lattice structure on the distal face to facilitate fixation. Used in clinical settings by orthopedic surgeons; intended for cemented or uncemented implantation. Device integrates with existing Freedom® femoral and tibial liner components. Output is a mechanical interface for knee joint reconstruction; provides stability and restores joint function. Benefits include pain relief and improved mobility for patients with severe knee joint pathology.

## Clinical Evidence

Bench testing only. Evaluations included tibial tray fatigue (ASTM F1800), residual particle characterization (ASTM F1877), porous surface characterization (ASTM F1854, F1160, F1044, F1978, F1147), sterilization (ISO 11137-2), endotoxin (AAMI ST72), and biocompatibility (ISO 10993-1, ISO 10993-5). Modular disassembly testing leveraged data from reference device K090411.

## Technological Characteristics

Material: Ti-6Al-4V ELI (ASTM F3001). Manufacturing: Additive (EBM) and subtractive. Design: Asymmetric tibial base plate with porous lattice structure. Sterilization: Sterile (SAL 10^-6). Connectivity: None (mechanical implant).

## Regulatory Identification

A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

## Predicate Devices

- Exactech, Inc. TRULIANT Porous Tibial Tray ([K182346](/device/K182346.md))

## Reference Devices

- Smith & Nephew Profix TKA System ([K030623](/device/K030623.md))
- Maxx Orthopedics Inc. Freedom Metal Backed Tibial Component ([K090411](/device/K090411.md))
- DJO Surgical, Inc. EMPOWR Porous Knee System ([K171991](/device/K171991.md))
- Smith & Nephew, Inc. Porous Patella and Porous Tibia Baseplate ([K211221](/device/K211221.md))
- Zimmer, Inc. Persona™ Personalized Knee System ([K230321](/device/K230321.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 13, 2025

Maxx Orthopedics, Inc. % Justin Gracyalny Regulatory Affairs Manager Secure BioMed Evaluations 7828 Hickory Flat Hwy, Suite 120 Woodstock, Georgia 30188

Re: K241597

Trade/Device Name: Freedom® Total Knee System - Porous Tibial Base Plate Regulation Number: 21 CFR 888.3565 Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBH, JWH Dated: January 17, 2025 Received: January 17, 2025

Dear Justin Gracyalny:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/3 description: The image shows the name "Lixin Liu-S" in a large, sans-serif font. The text is black and appears to be the main focus of the image. The background is white, with a faint, light blue watermark-like design that is partially visible behind the text.

Lixin Liu, Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K241597

Device Name

Freedom® Total Knee System - Porous Tibial Base Plate

Indications for Use (Describe)

The Freedom® Total Knee System is indicated for the following:

· Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.

- · Correction of functional deformities.
- · Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
- · Moderate valqus, varus, or flexion trauma.
- · Knee fractures untreatable by other methods.

· Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components and Primary PCK Components only).

The Freedom® Porous Tibial Base Plate and Cementless Femoral Components are indicated for Cemented or Uncemented use. All other components are indicated for cemented use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# K241597

# Freedom® Total Knee System – Porous Tibial Base Plate

| Date Prepared                          | February 13, 2025                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor                                | Maxx Orthopedics Inc.<br>2460 General Armistead Ave, Suite 100<br>Norristown, PA 19403                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 510(k) Contact                         | Secure BioMed Evaluations<br>Justin Gracyalny, MSE<br>7828 Hickory Flat Highway, Suite 120<br>Woodstock, GA 30188<br>Phone: 770-837-2681<br>Email: Regulatory@SecureBME.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Trade Name                             | Freedom® Total Knee System – Porous Tibial Base Plate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Common Name                            | Knee Prosthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code -<br>Regulation<br>Number | MBH - Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis (21 CFR §888.3565) JWH – Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis (21 CFR §888.3560)                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Primary<br>Predicate                   | K182346 Exactech, Inc. TRULIANT Porous Tibial Tray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Reference<br>Device(s)                 | K030623 Smith & Nephew Profix TKA System<br>K090411 Maxx Orthopedics Inc. Freedom Metal Backed Tibial Component<br>K171991 DJO Surgical, Inc. EMPOWR Porous Knee System<br>K211221 Smith & Nephew, Inc. Porous Patella and Porous Tibia Baseplate<br>K230321 Zimmer, Inc. Persona™ Personalized Knee System                                                                                                                                                                                                                                                                                                                                                                                          |
| Device<br>Description                  | The Freedom® Porous Tibial Base Plate is a line extension of the Freedom® Total Knee System comprising of tibial base plate components for cemented or uncemented use in total knee arthroplasty. Freedom® Porous Tibial Base Plates are intended for use with existing, compatible Freedom® femoral and tibial liner components. Freedom® Porous Tibial Base Plates are additively manufactured from Ti-6Al-4V ELI Grade 23 and include a porous lattice structure on the distal face. Freedom® Porous Tibial Base Plates are available in eight asymmetric design offerings (Sizes 1 – 8, Left / Right configurations), based on anterior / posterior (A/P) and medial / lateral (M/L) dimensions. |

{5}------------------------------------------------

| Indications for<br>Use Statement | The Freedom® Total Knee System is indicated for the following:                                                                                                                        |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | • Severe knee joint pain, loss of mobility, and disability due to:<br>rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.                                       |
|                                  | • Correction of functional deformities.                                                                                                                                               |
|                                  | • Post-traumatic loss of knee joint contour, particularly when there is<br>patellofemoral erosion, dysfunction, or prior patellectomy.                                                |
|                                  | • Moderate valgus, varus, or flexion trauma.                                                                                                                                          |
|                                  | • Knee fractures untreatable by other methods.                                                                                                                                        |
|                                  | • Revision surgery where sufficient bone stock and soft tissue integrity<br>are present (For PCK Components and Primary PCK Components<br>only).                                      |
|                                  | The Freedom® Porous Tibial Base Plate and Cementless Femoral<br>Components are indicated for Cemented or Uncemented use. All other<br>components are indicated for cemented use only. |

### Comparison of Technological Characteristics

| Device<br>Characteristic | Subject Device<br>Maxx Orthopedics Inc.<br>Freedom® Total Knee System –<br>Porous Tibial Base Plate               | Primary Predicate Device<br>Exactech, Inc.<br>TRULIANT Porous Tibial Tray<br>K182346                              |
|--------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Intended Use             | Total Knee Arthroplasty                                                                                           | Total Knee Arthroplasty                                                                                           |
| Manufacturing            | Additive (EBM) and Subtractive                                                                                    | Additive (EBM)                                                                                                    |
| Material                 | Ti-6Al-4V ELI per ASTM F3001                                                                                      | Ti-6Al-4V per ASTM F2924 /<br>ASTM F136                                                                           |
| Porous Coating           | Yes                                                                                                               | Yes                                                                                                               |
| Base Plate<br>Symmetry   | Asymmetrical                                                                                                      | Symmetrical                                                                                                       |
| Sterilization            | Implants: Provided sterile, SAL 10-6<br>Instruments: Provided non-sterile<br>for end-user sterilization, SAL 10-6 | Implants: Provided sterile, SAL 10-6<br>Instruments: Provided non-sterile<br>for end-user sterilization, SAL 10-6 |
| Packaging                | Dual Tray Configuration                                                                                           | Dual Tray Configuration                                                                                           |
| Single Use Only          | Yes                                                                                                               | Yes                                                                                                               |
| Prescription Use<br>Only | Yes                                                                                                               | Yes                                                                                                               |

## Technological Characteristics

There are no significant technological differences between the subject and predicate device. The subject device uses similar materials, is a similar design, and achieves its intended use in an identical manner as the predicate and both devices are manufactured using similar additive manufacturing techniques. Minor differences in base plate geometry, symmetry, and sizing are addressed via performance testing and similarity to the reference devices (K171991, K211221, K230321). Minor differences in the indications for use do not introduce a new intended use when 

{6}------------------------------------------------

compared to the predicate device or impact the safety and effectiveness of the device when used as labeled.

### Non-Clinical Performance Testing Summary

All necessary testing has been performed for the subject device to assure substantial equivalence to the predicate device and to demonstrate the subject device performs as intended. All testing was performed on worst case implants or test coupons as dictated by the relevant performance standards. The following evaluations were conducted:

- Tibial Tray Fatigue Testing per ASTM F1800
- Residual Particle Characterization Testing per ASTM F1877
- Porous Surface Characterization Testing per ASTM F1854, ASTM F1160, ASTM F1044, ● ASTM F1978, ASTM F1147
- Sterilization per ISO 11137-2 ●
- Endotoxin per AAMI ST72 ●
- Biocompatibility per ISO 10993-1, ISO 10993-5

Results of the residual particle characterization were shown to be substantially equivalent to values presented in the literature for the reference device (K030623). Modular disassembly testing was leveraged from the reference device (K090411) as the subject device uses an identical tibial insert locking mechanism.

### Conclusions

Based on the similarities of the intended use / indications for use, technological and functional characteristics, and the results of the non-clinical performance testing, the subject device is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH/K241597](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH/K241597)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
