Symmetric Total Knee System

K203779 · Signal Medical Corporation · MBH · Jun 10, 2021 · Orthopedic

Device Facts

Record IDK203779
Device NameSymmetric Total Knee System
ApplicantSignal Medical Corporation
Product CodeMBH · Orthopedic
Decision DateJun 10, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3565
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Symmetric™ Total Knee System consists of single use components intended for total knee arthroplasty with the following indications: 1. Rheumatoid arthritis, 2. Post-traumatic arthritis, 3. Osteoarthritis, 4. Degenerative arthritis in older patients whose age, weight and activity level are compatible with an adequate long-term result, 5. Failed osteotomies, unicompartmental replacement, or total knee replacement. All components are for cemented use only.

Device Story

Symmetric Total Knee System is a semi-constrained, cemented, patellofemorotibial prosthesis used in total knee arthroplasty. Device components include femoral components, tibial articular inserts, patellar components, tibial trays, stems, wedges, cones, and screws. System is implanted by orthopedic surgeons in a clinical/OR setting to replace damaged knee joints, aiming to restore function and reduce pain. This submission introduces an additional sterilization method using a flexible bag system; all other design and technological characteristics remain identical to the predicate device.

Clinical Evidence

No clinical data. Submission relies on bench testing regarding the qualification of the flexible bag sterilization vendor.

Technological Characteristics

Materials: Cobalt chromium alloy (ASTM F75), UHMWPE (ASTM F648), titanium (ASTM F1472). Coated components use plasma sprayed titanium (ASTM F1580). Semi-constrained, cemented, patellofemorotibial prosthesis. Sterilization: Flexible bag chamber method.

Indications for Use

Indicated for patients requiring total knee arthroplasty due to rheumatoid arthritis, post-traumatic arthritis, osteoarthritis, degenerative arthritis, or failed prior knee surgeries (osteotomies, unicompartmental, or total knee replacement).

Regulatory Classification

Identification

A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". June 10, 2021 Signal Medical Corporation Dora Culton Ouality Assurance and Regulator Engineer 400 Pvramid Dr. Marysville, Michigan 48040 Re: K203779 Trade/Device Name: Symmetric TM Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: June 4, 2021 Received: June 7, 2021 Dear Dora Culton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203779 Device Name Symmetric(TM) Total Knee System #### Indications for Use (Describe) The Symmetric(TM) Total Knee System consists of single use components intended for total knee arthroplasty with the following indications: - 1. Rheumatoid arthritis, - 2. Post-traumatic arthritis, - 3. Osteoarthritis, 4. Degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. - Failed osteotomies, unicompartmental replacement, or total knee replacement. న్. All components are for cemented use only. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K203779 Page 1 of 2 Image /page/3/Picture/1 description: The image contains a logo with a blue square containing a white wave-like design on the left. To the right of the logo, the words "SIGNAL MEDICAL" are written in blue. Below the logo and text, the words "05 - 510(k) Summary" are written in black. #### I. SUBMITTER Signal Medical Corporation 400 Pyramid Drive Marysville, MI 48040 Ph: (810) 364-7070 Fx: (810) 364-7072 Contact Person: Dora Culton Date Prepared: June 4, 2021 ### II. DEVICE | Name of Device: | Symmetric™ Total Knee System | |---------------------------------|----------------------------------------------------------------------------------------------| | Common or Usual Name: | Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, | | Cemented, Polymer/Metal/Polymer | | | Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis | | Regulatory Class: | II | | Product Codes: | JWH | | Regulation Number: | 888.3560 | ### III. PREDICATE DEVICE Symmetric Total Knee System, K080199 This predicate has not been subject to a design-related recall. No reference devices were used in this submission. ### IV. DEVICE DESCRIPTION The Femoral components are cobalt chromium alloy (ASTM F75). Tibial articular inserts and patellar components are UHMWPE (ASTM F648). Tibial trays, metaphyseal and revision stems, femoral and tibial wedges and cones, and screws (K961157), are titanium (ASTM F1472). Coated cobalt chromium and titanium components feature a coating of plasma sprayed titanium (ASTM F1580). The purpose of this 510(k) is to submit an additional sterilization method of a flexible bag system. Other descriptive details remain unchanged from the predicate. {4}------------------------------------------------ K203779 Page 2 of 2 Image /page/4/Picture/1 description: The image contains the logo for Signal Medical, which includes a blue square with a stylized "S" inside. To the right of the logo, the words "SIGNAL MEDICAL" are written in blue. Below the logo and text, the words "05 - 510(k) Summary" are written in black. ## V. INTENDED USE / INDICATIONS FOR USE The Symmetric™ Total Knee System consists of single use components intended for total knee arthroplasty with the following indications: - 1. Rheumatoid arthritis, - 2. Post-traumatic arthritis, - 3. Osteoarthritis, - 4. Degenerative arthritis in older patients whose age, weight and activity level are compatible with an adequate long-term result, - 5. Failed osteotomies, unicompartmental replacement, or total knee replacement. All components are for cemented use only. # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The same technological characteristics exist for both devices. ## VII. PERFORMANCE DATA The full sterilization qualification of the flexible bag sterilization vendor is included in the submission. ## VIII. CONCLUSIONS The technological characteristics/features and performance characteristics for the Symmetric Total Knee System are substantially equivalent to the legally marketed predicate device. This 510(k) is submitted for the addition of the flexible bag chamber sterilization method.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...