LEGION POROUS PLUS HA PRIMARY FEMORAL COMPONENTS

K091543 · Smith & Nephew, Inc. · MBH · Dec 21, 2009 · Orthopedic

Device Facts

Record IDK091543
Device NameLEGION POROUS PLUS HA PRIMARY FEMORAL COMPONENTS
ApplicantSmith & Nephew, Inc.
Product CodeMBH · Orthopedic
Decision DateDec 21, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3565
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact. The Smith & Nephew Legion Porous Plus HA Primary Femoral Components are indicated for use without bone cement and are single use devices.

Device Story

Legion Porous Plus HA Primary Femoral Components are orthopedic knee implants; geometrically identical to Legion Porous Primary femoral components (K073325); modified by addition of hydroxylapatite (HA) coating to porous-coated areas. Designed for uncemented fixation in total knee arthroplasty. Available in cruciate retaining (CR) and posterior stabilizing (PS) designs; sizes 1-8; right/left configurations. Utilizes existing Genesis II tibial baseplate and patellar components. Implanted by orthopedic surgeons in clinical/hospital settings. HA coating promotes bone apposition; device provides structural replacement for damaged knee joint surfaces; restores joint function; reduces pain. Mechanical testing confirms ability to withstand in vivo loading.

Clinical Evidence

Bench testing only. Mechanical testing results indicate the device is equivalent to clinically used devices and capable of withstanding expected in vivo loading without failure.

Technological Characteristics

Femoral knee prosthesis; metal/polymer construction; porous-coated for uncemented fixation; hydroxylapatite (HA) coating applied to porous areas. Available in CR and PS designs. Single-use device.

Indications for Use

Indicated for older patients with rheumatoid, post-traumatic, or degenerative arthritis requiring total knee replacement, or those with failed osteotomies or unicompartmental replacement. Posterior stabilized systems are for primary/revision surgery where cruciate ligaments are absent/incompetent and collateral ligaments are intact. For use without bone cement; single use only.

Regulatory Classification

Identification

A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K091543 # 510(k) Summary of Safety and Effectiveness LEGION® Porous Plus HA Primary Femoral Components Date of Summary: May 22, 2009 Contact Person and Address Jason Sells Manager, Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 E. Brooks Road Memphis, Tennessee 38116 (901) 399-5520 DEC 2 1 2009 Name of Device: Smith & Nephew Legion Porous Plus HA Primary Femoral Components Common Name: Femoral Knee Prosthesis Device Classification Name and Reference: 21 CFR 888.3565 Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis (Class II) Device Product Code: MBH ## Device Description Legion Porous Plus HA Primary Femoral Components are geometrically identical to the Legion Porous Primary femoral components cleared via K073325. This premarket notification seeks only to add a hydroxylapatite (HA) coating to the porous coated areas of the femoral implants. The subject devices will be initially offered in both cruciate retaining (CR) and posterior stabilizing (PS) designs in sizes 1-8 in right and left hand configurations. The subject devices will utilize the same articular inserts as the non-HA coated Legion Porous Primary femoral components cleared via K073325 and will use existing Genesis II tibial baseplate and patellar components cleared via K030612. ### Mechanical Testing A review of the mechanical testing results indicated that the Legion Porous Plus HA Primary Femoral Components are equivalent to devices currently used clinically and are capable of withstanding expected in vivo loading without failure. #### Intended Use Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, or degenerative anthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact. The Smith & Nephew Legion Porous Plus HA Primary Femoral Components are indicated for use without bone cement and are single use devices. #### Substantial Equivalence Information The Smith & Nephew Legion Porous Plus HA Primary Femoral Components are similar in overall design, indications, and materials to the Legion Porous Primary femoral components cleared via K073325. The subject devices feature the same HA coating as devices in the Genesis II Porous Plus HA Knee System cleared via K032683. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with flowing lines, representing the department's mission related to health and human welfare. The logo is simple and recognizable, often used in official documents and communications related to the department. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 Smith & Nephew, Inc. % Mr. Jason Sells Orthopaedic Division 1450 E Brooks Road Memphis, Tennessee 38116 DEC 2 1 2009 Re: K091543 Trade/Device Name: Legion Porous Plus HA Primary Femoral Components Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH Dated: December 8, 2009 Received: December 9, 2009 Dear Mr. Sells: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {2}------------------------------------------------ Page 2 - Mr. Jason Sells comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/dgfault.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Indications for Use 510(k) Number (if known): Device Name: LEGION Porous Plus HA Primary Femoral Components Indications for Use: Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact. The Smith & Nephew Legion Porous Plus HA Primary Femoral Components are indicated for use without bone cement and are single use devices. Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) FOR M. MELKERSON (Division Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 1 510(k) Number K091543
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...