ADVANCE HA COATED SPIKED TIBIAL BASE AND ADVANCE HA COATED MODULAR KEEL

K043083 · Wrightmedicaltechnologyinc · MBH · Apr 18, 2005 · Orthopedic

Device Facts

Record IDK043083
Device NameADVANCE HA COATED SPIKED TIBIAL BASE AND ADVANCE HA COATED MODULAR KEEL
ApplicantWrightmedicaltechnologyinc
Product CodeMBH · Orthopedic
Decision DateApr 18, 2005
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3565
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ADVANCE® HA Coated Components are indicated for use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and 5) treatment of fractures that are unmanageable using other techniques. The ADVANCE® HA Coated Components are for use without bone cement and are single use devices.

Device Story

ADVANCE® HA Coated Spiked Tibial Base and Modular Keel are orthopedic implants for knee arthroplasty. Components are manufactured from titanium alloy with titanium sintered bead porous coating and hydroxyapatite (HA) coating. Tibial base features 4 distal spikes; modular keel features 2 anterior and 2 posterior fins. Devices are intended for cementless fixation. Used by orthopedic surgeons in clinical settings to replace damaged joint surfaces, reduce pain, and restore knee function. Output is the physical implantation of the device to provide structural support and joint stability.

Clinical Evidence

No clinical data provided; substantial equivalence supported by bench testing and materials data.

Technological Characteristics

Materials: Titanium alloy, titanium sintered bead porous coating, hydroxyapatite (HA) coating. Design: Spiked tibial base (4 spikes), modular keel (2 anterior/2 posterior fins). Fixation: Cementless. Sterilization: Not specified. Energy source: None (mechanical implant).

Indications for Use

Indicated for skeletally mature patients undergoing knee arthroplasty for noninflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, revision procedures, or fracture treatment unmanageable by other techniques.

Regulatory Classification

Identification

A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the text 'K043083 (pg 1 of)'. The text appears to be handwritten in black ink on a white background. The text likely represents a document identifier or page number. Image /page/0/Picture/1 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a stylized graphic consisting of three parallel lines that are angled downwards and slightly overlap each other, creating a sense of depth. The letters are large and prominent, and the graphic is positioned to complement the text. APR 1 8 2005 ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS In accordance with the Food and Drug Administration Rule to implement provisions of the Safe in accordance with the 1 000 and in conformance with 21 CRF 807, this information serves as a Moural Dovies Ave of 1998 and is a for the ADVANCE® HA Coated Spiked Tibial Base and the ADVANCE® HA Coated Modular Keel. | Submitted By: | Wright Medical Technology, Inc. | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------| | Date: | November 5, 2004 | | Contact Person: | Ehab M. Esmail<br>Sr. Manager, Regulatory Affairs | | Proprietary Name: | ADVANCE® HA Coated Spiked Tibial Base and<br>the ADVANCE® HA Coated Modular Keel | | Common Name: | HA COATED COMPONENTS | | Classification Name and Reference: | 21 CFR 888.3565 Prosthesis, Knee,<br>Patello/femorotibial, Semi-Constrained,<br>Uncemented, Porous Coated,<br>Polymer/Metal/Polymer - Class II | | Device Product Code and Panel Code: | Orthopedics/87/ MBH | #### DEVICE INFORMATION #### A. INTENDED USE The ADVANCE® HA Coated Components are indicated for use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions: - 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; - 2) inflammatory degenerative joint disease including rheumatoid arthritis; - 3) correction of functional deformity; - 4) revision procedures where other treatments or devices have failed; and - 5) treatment of fractures that are unmanageable using other techniques. #### headquarters Wright Medical Technology, Inc. Arlington, TN 38002 901.867.9971 phone 5677 Airline Road ww.wmt.com international subsidiaries 011.32.2.378.3905 Belgium 011.39.0250.678.227 Italy 05.826.1600 Canada 011.81.3.3538.0474 Japan 011.33.1.45.13.24.40 France 011.44.1483.721.404 UK 011.49.4161.745130 Germany {1}------------------------------------------------ K043083 (pg 2 of 2) The ADVANCE® HA Coated Components are for use without bone cement and are single use devices. ## B. DEVICE DESCRIPTION The ADVANCE® HA Coated Components are an HA Coated Spiked Tibial Base and an The ADVANCES In Feel. The design features of the components are described below. ADVANCE® HA Coated Spiked Tibial Base - 4 spikes on the distal surface . - Manufactured from titanium alloy . - Titanium sintered bead porous coated . - HA coated ● ### ADVANCE® HA Coated Modular Keel - 2 small anterior fins . - 2 posterior fins . - Manufactured from titanium alloy . - HA coated . The design features of ADVANCE® HA Coated Components are substantially equivalent to the design features of other devices previously cleared for market. # C. SUBSTANTIAL EQUIVALENCE INFORMATION The intended use, material, and design features of ADVANCE® HA Coated Components are substantially equivalent to the predicate devices previously cleared for market. The safety and effectiveness of ADVANCE® HA Coated Components are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/12 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or bird with three stylized lines forming its body and wings. APR 1 8 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Ehab M. Esmail Sr. Manager, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002 Re: K043083 K043083 Trade/Device Name: ADVANCE® HA Coated Spiked Tibial Base and the ADVANCE® HA Coated Modular Keel Regulation Number: 21 CFR 888.3565 Regulation Number: 21 CFK 866.5503 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prothesis Regulatory Class: II Product Code: MBH Dated: March 25, 2005 Received: March 28, 2005 Dear Mr. Esmail: We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Security (10). "I substantially equivalent (for the indications referenced above and have determined the device is substantial in interstate for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosule) to regary manced prought in the Medical Device Amendments, or to commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior in May 28, 1970, the chance with the provisions of the Federal Food. Drug, devices that have been reciassified in accordance while approval application (PMA). and Cosmetic Act (Act) that do not require approval of a provisions of the Act . The and Cosmetic Act (Act) that to hot require approvide on trols provisions of the Act. The You may, merelore, market the devree, becjocher to the one in annual registration, listing of general controls provisions of the Act include requirements for any province and general controls provisions of the Feet neralies requires against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (sec above) movement major regulations affecting your device can may be subject to such additional controls. Existing major regulations EDA may may be subject to such additional controller Entrological of 898. In addition, FDA may be found in the Code of Federal Regulations, Title 21, Parts 800 to Register be found in the Code of Pederal Registering your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualies of a subscription with other requirements of the Act that FDA has made a determination that your device complies . You must that FDA nas made a delerimiation mac your as roos be species. You must of any list in or any Federal statutes and regulations daministered of registration and listing (21 l eomply with all the Act's requirements, including, but not limited to set comply with all the Act S requirements, mercessed in the practice requirements as set CFR Part 807); labeling (21 CFR Party, good and 820); and if applicable, the electronic forth in the quality systems (QS) regulation (21 CFR Path, 21 CFR 1000 1050 forth in the quality systems (QD) regalantin (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000 {3}------------------------------------------------ Page 2 – Mr. Ehab M. Esmail This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin marketing your antial equivalence of your device to a legally premarket notification. The FDA finding of substantial equivalias and thus, premarket notification. The FDA Inding of substanced by July 19, 2017 11:55 marketed predicate device results in a classification for your device and thus, permits your devic to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 81), please an If you desire specific advice for your active of the can and the regulation entitled, the regulation entitled, the international contact the Office of Compliance at (2 to 21 to a market notification the Division of Small "Misbranding by relerence to premarkst nothleadon (1) be Act from the Division of Small other general information on your responsibilities under the Act from the (800) 638, 2 other general information on your responsion.com in toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance at its toll-free warming on html Manufacturers, International and Consumer Fissionalite aroved childring stry/support/index.html. Sincerely yours, Stupt Plurde S.E. Ph.D. Airiam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K043083 #### Indications for Use 510(k) Number (if known): Device Name: ADVANCE® HA Coated Spiked Tibial Base and ADVANCE® HA Coated Modular Keel Indications For Use: The ADVANCE® HA Coated Components are indicated for use in knee arthroplasty for The ADVANCE® HA Coated Components are mercion in skeletally mature patients with the following conditions: - 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; - - 3) correction of functional deformity; - 3) correction of functional deforming, 4) revision procedures where other treatments or devices have failed; and - 4) Tevision procedures where of the read of the rechniques. The ADVANCE® HA Coated Components are for use without bone cement and are single use devices. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) typt Purbe (Division Sign-Off) Page 1 of ____________________________________________________________________________________________________________________________________________________________________ Division of General, Restorative, and Neurological Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...