The Ignite Anatomic Shoulder System is intended for use in total or hemi-shoulder replacement procedures to address the following: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - Rheumatoid arthritis. - Revision where other devices or treatments have failed. - Correction of functional deformity. - Fracture of the humeral head. - Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate.
Device Story
Total shoulder arthroplasty system; replaces damaged shoulder joint components. System includes humeral heads (38-57mm), offset taper adapters, stems (66-82mm), and glenoid components. Humeral stem implanted cementless; glenoid component implanted with cement. Used by orthopedic surgeons in clinical/OR settings. Provides mechanical replacement of joint surfaces to restore function and reduce pain. Benefits patients by addressing degenerative disease, fractures, and failed prior treatments.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via bench testing, including range of motion, biocompatibility, porous structure characterization, wear testing, taper pullout, construct fatigue, glenoid loosening (rocking horse, static/cyclic), glenoid pullout, shelf-life, and sterilization validation.
Indicated for patients requiring total or hemi-shoulder replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, humeral head fracture, or difficult clinical management cases like cuff arthropathy where other treatments are inadequate or failed.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Ignite Orthopedics LLC Russ Parrott VP of Research and Development 700 Park Avenue Suite F Winona Lake, Indiana 46590
Re: K202716
Trade/Device Name: Ignite Anatomic Shoulder System Regulation Number: 21 CFR 888.3670 Regulation Name: Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porouscoated uncemented prosthesis Regulatory Class: Class II Product Code: MBF, KWT, KWS, PAO, HSD Dated: April 19, 2021 Received: April 20, 2021
Dear Russ Parrott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
May 20, 2021
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael Owens Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202716
Device Name Ignite Anatomic Shoulder System
#### Indications for Use (Describe)
The Ignite Anatomic Shoulder System is intended for use in total or hemi-shoulder replacement procedures the following:
- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- · Rheumatoid arthritis.
- · Revision where other devices or treatments have failed.
- · Correction of functional deformity.
- · Fracture of the humeral head
· Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate.
Fixation Methods
The humeral stem is intended for cementless use. The glenoid is intended for cemented use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) Summary | |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared: | May 19, 2021 |
| Submitter: | Ignite Orthopedics LLC<br>700 Park Ave.<br>Suite F<br>Winona Lake, IN 46590 |
| Contact: | Russ Parrott<br>VP of Research and Development<br>Phone: 574.527.2864<br>russ.parrott@igniteorthopedics.com |
| Proprietary Name<br>of Subject Device: | Ignite Anatomic Shoulder System |
| Common Name: | Shoulder Arthroplasty System |
| Classification: | Shoulder Joint, Metal/Polymer/Metal, Non-Constrained or Semi-Constrained, Porous Coated, Uncemented Prosthesis (21 CFR §888.3670); Class II<br>Prosthesis, Shoulder, Non-Constrained, Metal/Polymer, Cemented (21 CFR §888.3650); Class II<br>Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Cemented (21 CER §888.3660); Class II<br>Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer + Additive, Cemented (21 CER §888.3660); Class II<br>Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented (21 CER §888.3690); Class II |
| Product Codes: | MBF, KWT, KWS, PAO, HSD |
| Substantially<br>Equivalent Predicate<br>Devices: | K060692 (Primary) - Comprehensive Primary Shoulder Stems, Biomet<br>Manufacturing Corp. |
{5}------------------------------------------------
K060694 - Modular Hybrid Glenoid, Biomet Manufacturing Corp.
K060716 - Versa-Dial Humeral Head Prosthesis, Biomet Manufacturing Corp.
K162024 - Altivate Anatomic Shoulder System, Encore Medical, L.P.
K111596 - Total Shoulder System, Shoulder Innovations, LLC
K091196 - Arthrosurface GRS™ Glenoid Resurfacing System, Arthrosurface, Inc.
K122698 - Aequalis Ascend Flex Shoulder System, Tornier SAS
K905786 - Global Total Shoulder, DePuy Inc.
K052472 – DePuy Global Shoulder Crosslink Glenoid, DePuy Orthopedics, Inc.
K121183 – Comprehensive Reverse Shoulder – E1 Polyethylene Claims, Biomet Manufacturing
#### Device Description:
The Ignite Anatomic Shoulder System is total shoulder arthroplasty system consisting of humeral heads (38 - 57mm diameters), offset taper adapters (multiple offsets), stems (66-82mm lengths), and glenoid components (multiple sizes).
The anatomic humeral consists of a wrought Cobalt-Chromium Humeral Head, a wrought Ti-6Al-4V Taper Adapter, and an additively manufactured Ti-6Al-4V Stem.
The All-Poly Anatomic Glenoid consists of Crosslinked Alpha-Tocopherol (Vitamin-E) infused Ultra-High-Molecular-Weight-Polyethylene. This device also has a 316L stainless steel pin pressed into it as an x-ray marker.
The system is provided with a set of instruments designed for preparation of the implant site and insertion of the implants into bone.
#### Intended Use / Indications:
The Ignite Anatomic Shoulder System is intended for use in total or hemi-shoulder replacement procedures to address the following:
- . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis. ●
- Revision where other devices or treatments have failed. ●
{6}------------------------------------------------
- Correction of functional deformity.
- Fracture of the humeral head.
- Difficult clinical management problems, including cuff arthropathy, where other methods . of treatment may not be suitable or may be inadequate.
## Fixation Methods
The humeral stem is intended for cementless use. The glenoid is intended for cemented use only.
### Summary of Technologies:
The Ignite Anatomic Shoulder System is substantially equivalent to the predicate devices in terms of its intended use and indications, material, size ranges, and design intent. While there are minor differences, the non-clinical testing described in the next section demonstrated that the differences do not raise new types of safety and effectiveness questions.
### Non-Clinical Testing:
The following performance data were provided in support of a substantial equivalence determination of the subject device:
- Range of Motion Evaluation (ASTM F1378)
- Biocompatibility Assessments (ISO 10993-1) ●
- Porous Structure Characterization (ASTM F1044, ASTM F1147, ASTM F1160, ASTM . F1978, ASTM F1854)
- Polyethylene Characterization ●
- Wear Testing (pristine and abrasive)
- Taper Pullout Testing (ASTM F2009)
- Construct Fatigue Testing ●
- Glenoid Loosening Testing (rocking horse - ASTM F2028)
- Glenoid Loosening Testing (static and cyclic torsional resistance) ●
- Glenoid Pullout Testing ●
- Shelf-Life Testing (ISO 11607-1 and ISO 11607-2) ●
- Sterilization Validation (ISO 11137-1 and ISO 11137-2)
# Clinical Testing:
Clinical testing was not necessary to demonstrate substantial equivalence of the Ignite Anatomic Shoulder System to the predicate devices.
# Conclusion:
A comparison of the subject and predicate devices, including comparison of the intended use, technological characteristics, and non-clinical testing results has demonstrated that the subject devices have a safety and effectiveness profile equivalent to that of the predicate devices. Thus, the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.