100KGY E-POLY ACETABULAR LINERS- ADDITIONAL PROFILES: +3 MAXROM AND +3 HI-WALL

K090103 · Biomet, Inc. · MAY · Feb 11, 2009 · Orthopedic

Device Facts

Record IDK090103
Device Name100KGY E-POLY ACETABULAR LINERS- ADDITIONAL PROFILES: +3 MAXROM AND +3 HI-WALL
ApplicantBiomet, Inc.
Product CodeMAY · Orthopedic
Decision DateFeb 11, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis. 3. Correction of functional deformity. 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. 5. Revision of previously failed total hip arthroplasty. Cemented and Uncemented Applications

Device Story

Acetabular liners manufactured from cross-linked polyethylene (100 kGy E-Poly); designed as components for hip joint prostheses. New +3 MaxRom and +3 Hi-Wall profiles added to existing product line to increase range of motion and joint stability in smaller patients. Used by orthopedic surgeons in total hip arthroplasty procedures; compatible with both cemented and uncemented applications depending on mating shell. Device provides bearing surface for hip joint; intended to restore joint function and stability.

Clinical Evidence

No clinical data provided as a basis for substantial equivalence; non-clinical laboratory testing performed.

Technological Characteristics

Material: 100 kGy cross-linked ultra-high molecular weight polyethylene (UHMWPE). Form factor: Acetabular liners with specific +3 MaxRom and +3 Hi-Wall profiles. Application: Semi-constrained hip prosthesis (cemented or uncemented).

Indications for Use

Indicated for patients requiring total hip arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or proximal femur fractures (non-union, femoral neck, trochanteric) unmanageable by other techniques, or revision of failed total hip arthroplasty.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K090103 (pg 1/2) Image /page/0/Picture/1 description: The image shows a logo with the word "BIOMET" in a stylized font. The letters are outlined with a rough, textured edge, giving them a three-dimensional appearance. The overall design is simple and bold, with a focus on readability and visual impact. The logo appears to be for a company named "BIOMET". FEB 11 2009 ## 510(k) Summary January 13, 2009 Preparation Date: Biomet Manufacturing Corp. Applicant/Sponsor: Contact Person: Becky Earl MANUFA Proprietary Name: 100 kGy E-Poly™Acetabular Liners- Additional Profiles: +3 MaxRom™ and +3 Hi-Wall Common Name: UHMWPE Liners #### Classification Name(s): - LPH- prosthesis, hip, semi-constrained, metal/polymer, porous uncemented (888.3358); � - JDI- prosthesis, hip, semi-constrained, metal/polymer, cemented (888.3350); � - LWJ- prosthesis, hip, semi-constrained, metal/polymer, uncemented (888.3360); g - MAY- prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous 0 cemented, osteophilic finish (888.3353) - LZO- prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, o uncemented (888.3353) Legally Marketed Devices To Which Substantial Equivalence Is Claimed: K070399, 100 kGy E-Poly™ Acetabular Liners-Additional Profiles Device Description: Biomet Manufacturing Corp. is adding new +3 MaxRom™ and +3 Hi-Wall profiles to bevice Description Liomet Handracumy to allow the surgeon an option for achieving more range of motion and joint stability in smaller patients. ### Intended Use: - 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - Rheumatoid arthritis. 2. - Correction of functional deformity. 3. - Treatment of non-union, femoral neck fracture, and trochanteric fractures of the র্থ proximal femur with head involvement, unmanageable using other techniques. - Revision of previously failed total hip arthroplasty. นว่ Cemented and Uncemented Applications Summary of Technologies: The intended use, indications and materials of the subject Summary of Treennoniges o its predicate counterpart, with the exception of the additional profiles. The components forflain now act it is cross-linked polyethylene in acetabular applications are adequately supported sarely and chectiveness of the erass innou paterials data, and testing results provided within this Premarket Notification. Mailing Address: P.O. Box 597 Warran, IN 46531-0587 Toll Free: 800,348,9500 stay bloadel com Shipping Aidress: 56 East Dell Driva Wasser 11 20552 {1}------------------------------------------------ .K090103 (pg 2/2) 510(K) Notification Biomet Manufacturing Corp. Blomet Manafaccaring of artistics of articles: +3 MaxRom™ and +3 Hi-Wall Page 2 of 2 Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use. Clinical Testing: None provided as a basis for substantial equivalence. All trademarks are property of Biomet, Inc. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes on its wing, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Biomet, Inc. % Ms. Becky Earl Regulatory Specialist 56 East Bell Drive P.O. Box 587 Warsaw, IN 46581 Re: K090103 Trade/Device Name: 100kGy E-Poly Acetabular Liners - Additional Profiles; +3 MaxRom™ and +3 Hi-Wall FEB 1 1 2009 Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prothesis Regulatory Class: Class II Product Code: MAY, LZO, LWJ, JDI, LPH Dated: January 13, 2009 Received: January 15, 2009 Dear Ms. Earl: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Becky Earl forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Wilkerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K090103 (ps 1/1) Device Name: 100kGy E-Poly™ Acetabular Liners-Additional Profiles: +3 MaxRom™ and +3 Hi-Wall Indications for Use: - 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - 2. Rheumatoid arthritis. - 3. Correction of functional deformity. - 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. - 5. Revision of previously falled total hip arthroplasty. Cemented and Uncemented Applications (as based on mating shell) YES Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use NO (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mirle H. Milken Page 1 of 1 (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...