← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K251335

# Tera Lumbar Interbody Fusion System (Various PNs) (K251335)

_Acuity Surgical Devices, LLC · MAX · Jun 2, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K251335

## Device Facts

- **Applicant:** Acuity Surgical Devices, LLC
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Jun 2, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Tera Lumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. The Tera Lumbar Interbody Fusion System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Tera Lumbar Interbody Fusion System is intended to be used with supplemental spinal fixation system cleared for use in the lumbosacral spine.

## Device Story

Intervertebral body fusion device system; used for TLIF, Oblique TLIF, or PLIF surgical approaches. Implants provide spinal stability and support fusion; used with allogenic bone graft or autograft. Requires supplemental spinal fixation. Cages manufactured via laser powder bed fusion (L-PBF) additive manufacturing; titanium alloy (Ti-6Al-4V ELI) construction; optional hydroxyapatite (HA) nano-coating for osseointegration. Provided sterile; used in clinical settings by surgeons. Output is physical structural support for vertebral bodies; aids fusion process; benefits patients by stabilizing degenerated spinal segments.

## Clinical Evidence

No clinical data. Bench testing only. Mechanical performance validated via static/dynamic axial compression and shear (ASTM F2077) and subsidence testing (ASTM F2267). Additional testing included tensile, microstructure, and chemical composition assessments (ASTM F3001).

## Technological Characteristics

Titanium alloy (Ti-6Al-4V ELI) per ASTM F3001; L-PBF additive manufacturing; layered porous lattice structure; optional 20nm hydroxyapatite (HA) coating. Sterile, single-use implants; reusable surgical instruments provided for steam sterilization. Standalone mechanical device; no software or electronics.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Align Lumbar Interbody Fusion Device ([K221535](/device/K221535.md))
- Ventris ([K243386](/device/K243386.md))
- Align HA ([K222561](/device/K222561.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

June 2, 2025

Acuity Surgical Devices, LLC
Chuck Forton
VP of New Product Development
8710 N. Royal Lane
Irving, Texas 75063

Re: K251335
Trade/Device Name: Tera Lumbar Interbody Fusion System (Various PNs)
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: MAX
Dated: April 29, 2025
Received: April 30, 2025

Dear Chuck Forton:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251335 - Chuck Forton
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K251335 - Chuck Forton
Page 3

Sincerely,

Brent Showalter -S

Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251335  |   |
|  Device Name Tera Lumbar Interbody Fusion System (Various PNs)  |   |
|  Indications for Use (Describe) The Tera Lumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. The Tera Lumbar Interbody Fusion System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Tera Lumbar Interbody Fusion System is intended to be used with supplemental spinal fixation system cleared for use in the lumbosacral spine.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

A C U I T Y S U R G I C A L K251335

# 510(k) Summary

## Contact Details

Applicant Name: Acuity Surgical Devices LLC
Applicant Address: 8710 N. Royal Lane
Irving TX, 75063 United States
Applicant Contact No.: (512) 585-3537
Applicant Contact: Mr. Chuck Forton
Applicant Contact email: cforton@acuitysurgical.com

Date Prepared: May 16, 2025

## Device Name

Device Trade Name: Tera Lumbar Interbody Fusion System (Various PNs)
Common Name: Intervertebral body fusion device
Classification Name: Intervertebral Fusion Device With Bone Graft, Lumbar
Regulation Number: 888.3080
Product Code(s): MAX

## Legally Marketed Predicate Devices

|  510(k) Number | Product Code | Trade Name | Manufacturer  |
| --- | --- | --- | --- |
|  Primary Predicate Device  |   |   |   |
|  K221535 | MAX | Align Lumbar Interbody Fusion Device | Acuity Surgical Devices LLC  |
|  Additional Predicate Devices  |   |   |   |
|  K243386 | OVD | Ventris | Acuity Surgical Devices LLC  |
|  K222561 | OVD | Align HA | Acuity Surgical Devices LLC  |

## Device Description

The Tera Lumbar Interbody Fusion System consists of intervertebral body fusion devices intended for lumbar interbody fusion using a transforaminal lumbar interbody fusion surgical approach (TLIF), an oblique transforaminal interbody fusion surgical approach (Oblique TLIF), or a posterior lumbar interbody fusion surgical approach (PLIF). The devices are intended to improve stability of the spine while supporting fusion. The Tera Lumbar Interbody Fusion System cages are intended for use at one or two contiguous levels in the lumbar spine (L2-S1). The components are offered in different shapes and sizes to meet the requirements of the individual patient's anatomy and are provided sterile.

{5}

ACUITY SURGICAL K251335

Tera Lumbar Interbody Fusion System cages are made of titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) with Powder Bed Fusion. All cages are manufactured using the L-PBF (laser powder bed fusion) additive manufacturing method. L-PBF allows for the formation of solid, non-porous cages with a layered porous lattice structure on the surfaces of the components, including the surfaces of the interior graft window. This intricate structure facilitates bone in-growth by providing a larger surface of implant/bone contact than a buffed surface. The cages are also titanium anodized to allow for identification of various heights by color.

Tera-T Ti Interbody cages, Tera-O Ti Interbody cages, and Tera-P Ti Interbody cages are also available with a Promimic HA nano Surface (hydroxyapatite (HA) coating). Cages are coated with a 20nm HA layer composed of crystalline hydroxyapatite particles that mimic human bone tissue through shape, composition, and structure. This surface treatment increases implant anchoring by facilitating osseointegration and enhancing early bone growth. All Tera Lumbar Interbody Fusion System implants are only available sterile packaged.

Non-sterile, reusable surgical instruments to support implantation of the system are provided for use with Tera Lumbar Interbody Fusion System devices are provided in steam sterilization trays.

## Intended Use / Indications for Use

The Tera Lumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion.

The Tera Lumbar Interbody Fusion System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The Tera Lumbar Interbody Fusion System is intended to be used with supplemental spinal fixation system cleared for use in the lumbosacral spine.

Page 2 of 4

{6}

ACUITY SURGICAL K251335

# Indications for Use Comparison

The Tera Lumbar Interbody Fusion System has identical Indications for Use as the primary predicate device, Align Lumbar Interbody Fusion System (K221535) and similar Indications for Use to the additional predicate devices cleared in K243386 and K222561. The subject device is indicated for the same patient population, condition, and mechanism of action as the primary predicate device.

# Technological Comparison

The Tera Lumbar Interbody Fusion System has the same indications, design, similar dimensions, and uses identical materials as the primary predicate device, K221535. The subject device also has identical or similar technological characteristics as the primary predicate device.

Differences between the subject device and primary predicate device include a change in the device name, updated implant cage inserter interface designs, additional implant heights and lordosis offerings, an update to instrument designs, additional instrument options, and an additional HA coating option. The additive manufacturing process was also changed to a new supplier, which is identical to the additional predicate device K243386.

The subject device has identical or similar characteristics to additional predicate devices K243386 and K222561 with similar size range and lordosis angle to K243386. The subject device comes with an optional HA coating, which is the same material as the additional predicate device cleared in K222561.

To ensure these technological characteristics do not affect the safety and effectiveness of the subject device, a worst-case analysis and resulting worst-case configuration mechanical testing was conducted on the new implant sizes and configurations. Based on the testing conducted including mechanical performance testing, design validation analysis, biocompatibility testing, and sterilization and packaging validation, it can be concluded that the subject device does not raise any new issues of safety and effectiveness as compared to the primary predicate K221535, and additional predicate devices K243386 and K222561.

The Tera Lumbar Interbody Fusion System is substantially equivalent to the legally marketed predicate device Align Lumbar Interbody Fusion System (K221535) under regulation 21 CFR 888.3080, product code MAX.

Page 3 of 4

{7}

A C U I T Y S U R G I C A L K251335

# Non-Clinical and/or Clinical Tests Summary &amp; Conclusions

The subject device underwent mechanical performance testing to validate that the performance of the worst-case subject device is substantially equivalent to the previously cleared predicate devices. Test methods including static axial compression, static axial compression shear, dynamic axial compression, and dynamic axial compression shear per ASTM F2077 Test Methods for Intervertebral Body Fusion Devices, and subsidence testing per ASTM F2267 Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Devices Under Static Axial Compression ensured that the design features of the subject device met the required mechanical strength criteria for their intended use. Additional testing included tensile testing, microstructure assessment, and chemical composition assessment per ASTM F3001. The performance testing demonstrated substantial equivalence between the subject and predicate devices. The results of the non-clinical testing did not identify any new or increased risks associated with the change in additive manufacturing supplier. Performance equivalence demonstrated that the subject device met the acceptance criteria of the standards and is substantially equivalent to the predicate devices.

Clinical Testing was Not Applicable.

The performance testing demonstrated substantial equivalence between the subject and predicate devices. The results of the non-clinical testing did not identify any new or increased risks associated with the change in additive manufacturing supplier. Performance equivalence demonstrated that the subject device met the acceptance criteria of the standards and is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K251335](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K251335)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
