← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K223688

# CONDUIT Lateral Lumbar Intervertebral Fusion Cage, CONDUIT Instruments (K223688)

_Medos International Sàrl · MAX · Mar 8, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K223688

## Device Facts

- **Applicant:** Medos International Sàrl
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Mar 8, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The CONDUIT™ LLIF Cage with a microscopic roughened surface and micro and nano-scale features is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. CONDUIT™ LLIF Cage is to be used with supplemental fixation. The CONDUIT™ LLIF Cages (>20 degrees) must be used with CONDUIT™ Lateral Switch Plate and supplemental fixation. When used with or without the CONDUIT™ Lateral Switch Plate, the system is indicated for use with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

## Device Story

CONDUIT™ LLIF Cage is an intervertebral body fusion device used to restore disc height and facilitate fusion in the lumbar spine (L2-S1). The device is implanted via a lateral approach by a surgeon. It features a hollow geometry for packing with autogenous/allogeneic bone graft or FDA-cleared bone void fillers. The cage is used in conjunction with supplemental spinal fixation (e.g., posterior pedicle screws). An optional CONDUIT™ Lateral Switch Plate may be attached to the cage and adjacent vertebral bodies to provide additional stability and migration resistance. The device benefits patients by stabilizing the spinal segment and promoting fusion in cases of degenerative disc disease.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including dynamic axial compression, dynamic compression shear, static axial compression, static compression shear (ASTM F2077-18), expulsion analysis (ASTM F1839-08), and subsidence analysis (ASTM F2267-04). MRI safety was evaluated per ASTM F2052-15, F2182-17, F2119-07, and F2182-19e2.

## Technological Characteristics

Material: Ti-6Al-4V alloy (ISO 5832-3). Manufacturing: Additive manufacturing. Design: Solid and porous structures with micro/nano-scale surface features. Dimensions: 14-18 mm heights, 22 and 30-degree lordosis angles. Connectivity: None. Sterilization: Not specified. Energy source: None (mechanical implant).

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- CONDUIT™ Cages and FIBERGRAFT™ BG Putty ([K222276](/device/K222276.md))
- EIT Cellular Titanium® LLIF Cage ([K201605](/device/K201605.md))
- EIT Cellular Titanium® Lumbar Cage LLIF ([K181644](/device/K181644.md))
- NuVasive Brigade System, Brigade Lateral System ([K203714](/device/K203714.md))
- SYNFIX Evolution System ([K162358](/device/K162358.md))
- CONDUIT Lateral Switch Plate ([K221325](/device/K221325.md))
- CONDUIT Instruments ([K210728](/device/K210728.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

March 8, 2023

Medos International SÀRL Mr. Delano Martins Senior Regulatory Affairs Specialist Chemin-Blanc 38 Le Locle 2400 Switzerland

Re: K223688

Trade/Device Name: CONDUIT™ Lateral Lumbar Intervertebral Fusion Cage: CONDUIT™ Instruments Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: December 8, 2023 Received: December 9, 2023

Dear Mr. Martins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K223688

#### Device Name

CONDUIT™ Lateral Lumbar Intervertebral Fusion Cage; CONDUIT™ Instruments

#### Indications for Use (Describe)

The CONDUIT™ LLIF Cages with a microscopic roughened surface and micro and nano-scale features is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. CONDUIT™ LLIF Cage is to be used with supplemental fixation. The CONDUIT™ LLIF Cages (>20 degrees) must be used with CONDUIT™ Lateral Switch Plate and supplemental fixation. When used with or without the CONDUIT™ Lateral Switch Plate, the system is indicated for use with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

| Type of Use (Select one or both, as applicable) |   |
|-------------------------------------------------|---|
| Prescription Use (Part 21 CFR 801 Subpart D)    | × |
| Over-The-Counter Use (21 CFR 801 Subpart C)     |   |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY - K223688

#### Submitter Information A.

|    | Manufacturer:                            | Medos International SÀRL<br>Chemin-Blanc 38<br>Le Locle 2400 Switzerland                                                                                                                                                                                                          |
|----|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | Submitter:                               | DePuy Synthes Spine<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                                                                                                                                                   |
|    | Contact Person:<br>Telephone<br>E-Mail   | Mr. Delano Martins<br>+1 (904) 699-5797<br>dmart152@its.jnj.com                                                                                                                                                                                                                   |
| B. | Date Prepared                            | February 17, 2023                                                                                                                                                                                                                                                                 |
| C. | Device Name                              |                                                                                                                                                                                                                                                                                   |
|    | Trade/Proprietary Name:                  | CONDUIT™ Lateral Lumbar Intervertebral Fusion<br>Cage; CONDUIT™ Instruments                                                                                                                                                                                                       |
|    | Device Classification<br>and Regulation: | Class II per 21 CFR §888.3080                                                                                                                                                                                                                                                     |
|    | Product Codes:                           | MAX                                                                                                                                                                                                                                                                               |
|    | D. Predicate Device Names                |                                                                                                                                                                                                                                                                                   |
|    | Primary Predicate:                       | CONDUIT™ Cages and FIBERGRAFT™ BG<br>Putty (K222276)                                                                                                                                                                                                                              |
|    | Additional Predicate:                    | EIT Cellular Titanium® LLIF Cage (K201605);<br>EIT Cellular Titanium® Lumbar Cage LLIF<br>(K181644); NuVasive Brigade System, Brigade<br>Lateral System (K203714); SYNFIX Evolution<br>System (K162358); CONDUIT Lateral Switch Plate<br>(K221325); CONDUIT Instruments (K210728) |

{4}------------------------------------------------

### E. Device Description Summarv

The purpose of this Traditional 510(k) is to seek marketing clearance for 22-degree and 30degree lordosis angled CONDUIT™ Lateral Lumbar Intervertebral Fusion (LLIF) Cages in the CONDUIT™ line of lumbar interbody cages with additional cage height of 14 to 18 mm. The CONDUIT™ line of lumbar interbody cages includes lordosis angles of 0, 8, and 16 degrees.

The CONDUIT™ LLIF Cage is used to restore intervertebral height and to facilitate intervertebral body fusion in the spine using an LLIF (Lateral Lumbar Intervertebral Fusion) approach with cancellous bone graft and/or corticocancellous bone graft materials, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The CONDUIT™ LLIF Cage is intended to be used from L2-S1 in patients with DDD and up to Grade I spondylolisthesis or retrolisthesis at one or two contiguous levels. The device is intended to be used alongside supplemental spinal fixation, either applied anteriorly or posteriorly (e.g., using posterior pedicle screws). The CONDUIT™ Lateral Switch Plate is an optional device that connects to the CONDUIT™ LLIF Cage and adjacent vertebral body(s) to provide additional migration resistance and stability via DePuy Synthes AEGIS Screws. When used with or without the CONDUIT™ Lateral Switch Plate, the system is indicated for use with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine.

The CONDUIT™ LLIF Cage is made from Ti-6A1-4V alloy (ISO 5832-3) by an additive manufacturing process. The design contains solid structures and porous structures. The hollow geometry of the implants allows the cage to be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.

### F. Indented Use/Indications for Use

The CONDUIT™ LLIF Cage with a microscopic roughened surface and micro and nano-scale features is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. CONDUIT™ LLIF Cage is to be used with supplemental fixation. The CONDUIT™ LLIF Cages (>20 degrees) must be used with CONDUIT™ Lateral Switch Plate and supplemental fixation. When used with or without the CONDUIT™ Lateral Switch Plate, the system is indicated for use with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

{5}------------------------------------------------

### G. Indications for Use Comparison

The subject device comprise an implant system which has indications for use similar to those of the predicate devices.

#### F. Non-Clinical Test Summary and Conclusions

The following tests and analyses were performed on the subject device system to demonstrate that the additional angles of the CONDUIT™ LLIF Cage are substantially equivalent to other predicate devices:

- · Dynamic Axial Compression (ASTM F2077-18)
- · Dynamic Compression Shear (ASTM F2077-18)
- · Static Axial Compression analysis (ASTM F2077-18)
- · Static Compression Shear (ASTM F2077-18)

Expulsion analysis (ASTMF1839-08) and Subsidence analysis (ASTM F2267-04) were evaluated to demonstrate substantial equivalence.

The additional angled CONDUIT™ LLIF Cages with CONDUIT™ Lateral Switch Plate Construct testing was evaluated and due to the high level of confidence in previous dynamic testing as well as the addition of a Switch Plate resulting in a non-worst-case condition in all Static type tests (ASTM F2077, ASTM F2267 and F1839) was evaluated to demonstrate substantial equivalence.

MRI Safety Testing per ASTM F2052-15, ASTM F2182-17, ASTM F2119-07 and ASTM F2182-19e2 was evaluated to demonstrate substantial equivalence.

The subject device is substantially equivalent to legally marketed predicate devices with respect to indications for use, design, function, material composition, and performance testing standards.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K223688](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K223688)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
