K223412 · Vy Spine, LLC · MAX · Jan 6, 2023 · Orthopedic
Device Facts
Record ID
K223412
Device Name
LumiVy Lumbar IBF System
Applicant
Vy Spine, LLC
Product Code
MAX · Orthopedic
Decision Date
Jan 6, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The LumiVy™ Lumbar IBF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. The implants are intended to be used with legally cleared supplemental spinal fixation cleared for the implanted level. The LumiVy™ Lumbar IBF System is also intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. The lumbar device is to be used in patients who have had six months of non-operative treatment. Additional supplemental fixation is not necessary for the LumiVy™ - AS implants if the integrated screws are implanted.
Device Story
LumiVy Lumbar IBF System consists of spinal interbody spacers and bone screws; implanted surgically via anterior, lateral, or posterior approach to facilitate vertebral fusion. Spacers available in various anatomical shapes (cylindrical, rectangular, elliptical, hemi-cylindrical) to accommodate vertebral bodies. Includes Tantalum wires for fluoroscopic visualization. Used by surgeons in clinical settings. Provides structural support to disc space; promotes fusion when used with autogenous bone graft. Benefits patients by stabilizing lumbar spine segments affected by degenerative disc disease.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Spacers made of PEEK Optima LT1 or PEEK Optima LT1-HA with Tantalum markers; NanoVy Ti components feature CP Titanium coating. Bone screws made of Titanium alloy. Mechanical interbody fusion device; no software or electrical components.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) and up to Grade 1 spondylolisthesis at one or two contiguous levels (L2-S1). Requires six months of failed non-operative treatment. Used with autogenous bone graft and supplemental fixation (unless using integrated screws).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K181118 — Reliance Lumbar IBF System · Reliance Medical Systems · Jan 9, 2019
K160699 — VIRTU Lumbar Spacer System · Met 1 Technologies, LLC · Aug 1, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 6, 2023
Vy Spine, LLC Jordan Hendrickson Operations Manager 2236 Capital Circle NE, Suite 103-1 Tallahassee, Florida 32308
Re: K223412
Trade/Device Name: LumiVyTM Lumbar IBF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: November 9, 2022 Received: November 9, 2022
Dear Jordan Hendrickson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223412
Device Name LumiVyTM Lumbar IBF System
#### Indications for Use (Describe)
The LumiVy™ Lumbar IBF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. The implants are intended to be used with legally cleared supplemental spinal fixation cleared for the implanted level.
The LumiVy™ Lumbar IBF System is also intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. The lumbar device is to be used in patients who have had six months of non-operative treatment.
Additional supplemental fixation is not necessary for the LumiVy™ - AS implants if the integrated screws are implanted.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
23 November 2022
Vy Spine, LLC 2236 Capital Circle NE Suite 103-1 Tallahassee, FL 32308 866-489-7746 Telephone: 850-597-8571 Fax:
Contact: Jordan Hendrickson Operations Manager
Common or Usual Name: Intervertebral Body Fusion Device Proposed Proprietary or Trade Name: LumiVyTM Lumbar IBF System Classification Name: Class II, Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 MAX Product Code:
## Substantial Equivalence
The LumiVy™ NanoVy™ Ti Lumbar IBF System is substantially equivalent to the primary predicate device LumiVy™ Lumbar IBF System (K212930) and the secondary predicate SeaSpine NanoMetalene Spacer System (K200879, K193636, K192132, K181079, K173606, K173260. K142488). The LumiVy™ NanoVy™ Ti Lumbar IBF System is equivalent to these commercially available devices in terms of material, intended use, levels of attachment, size range, and use with supplemental fixation.
#### Device Description
The LumiVy™ Lumbar IBF System device is intended to be used as an intervertebral body fusion device. The device is surgically implanted between vertebral bodies from an anterior. lateral, or posterior surgical approach. These devices also contain Tantalum wires to aid in fluoroscopic visualization. The LumiVy™ Lumbar IBF System device is available in multiple anatomical shapes and sizes, to accommodate various vertebral bodies. These include smaller cylindrical and rectangular shapes, elongated elliptical shapes, and larger hemi-cylindrical shapes. The spacer may be made of PEEK Optima LT1 or PEEK Optima LT1-HA with Tantalum markers. The LumiVy™ NanoVy™ Ti Lumbar IBF components are made of PEEK Optima LT1 with CP Titanium coating. The LumiVy™ Lumbar IBF System may also include bone screws manufactured from Titanium alloy to secure the device to the vertebral body.
#### Intended Use/Indications for Use
The LumiVyTM Lumbar IBF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. The implants are intended to be used with legally cleared supplemental spinal fixation cleared for the implanted level.
The LumiVy™ Lumbar IBF System is also intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with
{4}------------------------------------------------
degeneration of the disc confirmed by history and radiographic studies. The lumbar device is to be used in patients who have had six months of non-operative treatment. Additional supplemental fixation is not necessary for the LumiVyTM -AS implants if the integrated screws are implanted.
## Non-Clinical Testing
No additional testing was performed on the LumiVy™ Ti Lumbar IBF components.
#### Technological Modifications
The subject LumiVy™ NanoVy™ Ti Lumbar IBF System differs from the primary predicate devices in terms of CP titanium coating.
## Technological Characteristics
The subject LumiVy™ NanoVy™ Ti Lumbar IBF components are substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to have equivalent technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.