← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K210583

# SeaSpine WaveForm A Interbody System (K210583)

_SeaSpine Orthopedics Corporation · MAX · Jun 2, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K210583

## Device Facts

- **Applicant:** SeaSpine Orthopedics Corporation
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Jun 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Interbody Device (IBD) Implants (i.e., interbody implants used alone): The SeaSpine WaveForm A System Interbody is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the device is to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine WaveForm A Interbody is intended for use with supplemental fixation. Degenerative disc disease is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be sketally mature and have had at least six (6) months of non-operative treatment. No-Profile Implants w/ Screws: The SeaSpine WaveForm A System No-Profile Interbody, when used with Screws and a No-Profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation. No-Profile Implants w/ Inline Fixation Anchors: The SeaSpine WaveForm A System No-Profile Interbody, when used with Inline Fixation Anchors and a No Profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The SeaSpine WaveForm A No-Profile Implants w/ Inline Fixation Anchors is intended for use with supplemental fixation. TruProfile Interbody Implants: The SeaSpine WaveForm A Interbody assembled with the Anterior Plate, when used with Screws and an Anterior Plate Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.

## Device Story

Additively manufactured intervertebral fusion device; titanium alloy (Ti-6Al-4V ELI). Features large central graft windows for autogenous/allogeneic bone graft packing. System includes interbody spacers, optional anterior faceplates, bone screws, inline fixation anchors, and locking covers. Used in spinal surgery to facilitate fusion; implanted by surgeons. Provides structural support to disc space; promotes bone ingrowth via graft. Instruments provided for placement, adjustment, and removal. Sterile-packed; steam sterilized in facility. Benefits patients by stabilizing spinal segments and addressing discogenic pain.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated via ASTM F2077 (compression/compression-shear), ASTM F2267 (subsidence), F1877 (wear), expulsion, and static screw push-out. Sterility, packaging, and bacterial endotoxin (ANSI/AAMI ST-72:2011) testing performed.

## Technological Characteristics

Material: Ti-6Al-4V ELI titanium alloy (ASTM F3001). Additively manufactured. Sterile-packed. Mechanical testing per ASTM F2077, F2267, F1877. No software or electronic components.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- SeaSpine Meridian System ([K201193](/device/K201193.md))
- SeaSpine Vu a•POD Intervertebral Body Fusion Device ([K101310](/device/K101310.md))
- SeaSpine WaveForm TO System ([K201755](/device/K201755.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 2, 2021

SeaSpine Orthopedics Corporation Aly Alvarez Associate Manager, Regulatory Affairs 5770 Armada Drive Carlsbad, California 92008

Re: K210583

Trade/Device Name: SeaSpine WaveForm™ A Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OVD Dated: May 3, 2021 Received: May 5, 2021

Dear Aly Alvarez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K210583

Device Name SeaSpine WaveForm™ A Interbody System

#### Indications for Use (Describe)

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The SeaSpine WaveForm A System Interbody is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the device is to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine WaveForm A Interbody is intended for use with supplemental fixation. Degenerative disc disease is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be sketally mature and have had at least six (6) months of non-operative treatment.

#### No-Profile Implants w/ Screws:

The SeaSpine WaveForm A System No-Profile Interbody, when used with Screws and a No-Profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.

#### No-Profile Implants w/ Inline Fixation Anchors:

The SeaSpine WaveForm A System No-Profile Interbody, when used with Inline Fixation Anchors and a No Profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The SeaSpine WaveForm A No-Profile Implants w/ Inline Fixation Anchors is intended for use with supplemental fixation.

#### TruProfile Interbody Implants:

The SeaSpine WaveForm A Interbody assembled with the Anterior Plate, when used with Screws and an Anterior Plate Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

{3}------------------------------------------------

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

# Contact Details

| Applicant Name:     | SeaSpine Orthopedics Corporation                                   |
|---------------------|--------------------------------------------------------------------|
| Address:            | 5770 Armada Drive, Carlsbad CA                                     |
| Phone number:       | (619) 884-4342 (Aly) or 760-271-6804 (Alicia)                      |
| Fax number:         | (760) 683-6874                                                     |
| Contact person:     | Aly Alvarez, Associate Manager, Regulatory Affairs                 |
| Additional contact: | Alicia McArthur, Regulatory Affairs Specialist, Regulatory Affairs |
| Date Prepared:      | May 4, 2021                                                        |

Device Name

| Trade Name:                       | SeaSpine WaveForm™ A Interbody System                                                                                         |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Device Classification Regulation: | 888.3080                                                                                                                      |
| Common Name:                      | Intervertebral Body Fusion Device                                                                                             |
| Classification Name:              | Intervertebral body fusion device with bone graft, lumbar; Intervertebral body fusion device with integrated fixation, lumbar |
| Class:                            | II                                                                                                                            |
| Product Code:                     | MAX, OVD                                                                                                                      |

{5}------------------------------------------------

#### Legally Marketed Predicate Devices

| 510(k) Number               | Product Code | Trade Name                                                | Manufacturer                                                           |
|-----------------------------|--------------|-----------------------------------------------------------|------------------------------------------------------------------------|
| Primary Predicate Device    |              |                                                           |                                                                        |
| K201193                     | MAX, OVD     | SeaSpine Meridian System                                  | SeaSpine<br>Orthopedics<br>Corporation                                 |
| Additional Predicate Device |              |                                                           |                                                                        |
| K101310                     | OVD          | SeaSpine Vu a•POD<br>Intervertebral Body Fusion<br>Device | SeaSpine<br>Orthopedics<br>Corporation<br>(previously Theken<br>Spine) |
| Additional Predicate Device |              |                                                           |                                                                        |
| K201755                     | MAX          | SeaSpine WaveForm TO<br>System                            | SeaSpine<br>Orthopedics<br>Corporation                                 |

### Device Description

The Seaspine WaveForm A System are additively manufactured intervertebral fusion devices with large central graft windows which are packed with autogenous bone graft and/or allogenic bone graft, composed of cancellous, cortical, and/or corticocancellous bone prior to implantation. The implants are manufactured from Ti-6Al-4V ELI titanium alloy per ASTM F3001 and are provided sterile-packed. The WaveForm A interbodies have the ability to accept an anterior faceplate and offer various configurations that can be used with bone screws or inline fixation anchors, locking covers and an optional spin plate. The instruments included with the system facilitate the placement and adjustment of the interbody spacers, and removal if necessary. The instruments are placed in system-specific tray components for storage, protection, and organization prior to and during the steam sterilization process.

### Intended Use/Indications for Use

### Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The SeaSpine WaveForm A System Interbody is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the device is to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The SeaSpine WaveForm A Interbody is intended for use with supplemental fixation. Degenerative disc disease is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment.

{6}------------------------------------------------

#### No-Profile Implants w/ Screws:

The SeaSpine WaveForm A System No-Profile Interbody, when used with Screws and a No-Profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.

### No-Profile Implants w/ Inline Fixation Anchors:

The SeaSpine WaveForm A System No-Profile Interbody, when used with Inline Fixation Anchors and a No Profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The SeaSpine WaveForm A No-Profile Implants w/ Inline Fixation Anchors is intended for use with supplemental fixation.

### TruProfile Interbody Implants:

The SeaSpine WaveForm A Interbody assembled with the Anterior Plate, when used with Screws and an Anterior Plate Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. Hyperlordotic sizes (25 and 30 degrees) are intended for use with supplemental fixation.

### Summary of Technological Characteristics

The SeaSpine WaveForm A System is identical or similar to the cited predicate devices in regard components, device description, intended use/indications for use, technological to characteristics (operating principle, design, materials, etc.) and performance (mechanical safety).

{7}------------------------------------------------

The implants are used to treat the same conditions, have essentially the same precautions and contraindications for use, and represent a basic design concept in terms of safety and effectiveness, and differ only in design details and not functionality.

### Non-Clinical Testing

Mechanical Performance in compression and compression-shear (ASTM F2077), subsidence (ASTM F2267), wear testing (F1877), expulsion, and static screw push out was performed. Packaging, shipping and sterilization tests were also conducted to validate a Sterility Assurance Level (SAL) of 10 -6 and ensure maintenance of a sterile barrier. Bacterial Endotoxin (BET) was conducted in accordance with ANSI/AAMI ST- 72:2011. The SeaSpine WaveForm A System has demonstrated equivalent mechanical performance to the predicate system in accordance with these requirements.

### Conclusions

The submitted data demonstrates that the SeaSpine WaveForm A System performs at least as safely and effectively as the cited legally marketed predicate.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K210583](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K210583)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
