Nexxt Matrixx System

K193370 · Nexxt Spine, LLC · MAX · Jan 21, 2020 · Orthopedic

Device Facts

Record IDK193370
Device NameNexxt Matrixx System
ApplicantNexxt Spine, LLC
Product CodeMAX · Orthopedic
Decision DateJan 21, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

When used as a lumbar intervertebral fusion device, the Nexxt Matrixx® open devices are indicated for use at one or two contiguous levels in the lumbar spine, from L2-S1, in skeletally mature patients who have had six months of nonoperative treatment for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. Additionally, the Nexxt Matrixx® lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. The device is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and with supplemental fixation.

Device Story

Nexxt Matrixx® System consists of additively manufactured titanium alloy spinal spacers; designed to provide structural stabilization of spine. Implants feature external structural frame with roughened surface (~7µm) and internal geometric lattices with 300-700µm pores. Open devices include large vertical cavity for packing autograft/allograft bone graft; solid devices are closed for partial vertebral body replacement. Used in lumbar spine surgery; implanted by surgeons. Available in various heights, lengths, widths, and lordotic angles to match patient anatomy. Provides structural support to facilitate fusion; used with supplemental fixation. Benefits include stabilization of spinal segments in patients with degenerative conditions.

Clinical Evidence

No clinical data. Performance evaluated via bench testing, specifically dimensional analyses, to confirm substantial equivalence to predicate devices.

Technological Characteristics

Additively manufactured Ti-6Al-4V ELI titanium alloy (ASTM F3001). Features structural frame with ~7µm surface roughness and 300-700µm lattice pores. Available in various dimensions and lordotic angles. Standalone implant; non-electronic.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbar levels (L2-S1) with up to Grade 1 spondylolisthesis, or as an adjunct to fusion for degenerative scoliosis, following six months of failed nonoperative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 21, 2020 Nexxt Spine, LLC % Karen E. Warden, PhD President BackRoads Consulting Inc. PO Box 566 Chesterland, Ohio 44026 Re: K193370 Trade/Device Name: Nexxt Matrixx® System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: December 23, 2019 Received: December 26, 2019 Dear Dr. Warden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Brent Showalter. PhD Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K193370 Device Name Nexxt Matrixx® System Indications for Use (Describe) When used as a lumbar intervertebral fusion device, the Nexxt Matrixx® open devices are indicated for use at one or two contiguous levels in the lumbar spine, from L2-S1, in skeletally mature patients who have had six months of nonoperative treatment for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. Additionally, the Nexxt Matrixx® lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. The device is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and with supplemental fixation. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------| | <span style="padding-right: 5px;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="padding-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor and a person is not required to respond to a collection of "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary | Date: | 4 December 2019 | |---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | Nexxt Spine, LLC<br>14425 Bergen Blvd, Suite B<br>Noblesville, IN 46060<br>Office: 317.436.7801<br>Fax: 317.245.2518 | | Sponsor Contact: | Andy Elsbury, President | | 510(k) Contact: | Karen E. Warden, PhD<br>BackRoads Consulting Inc.<br>PO Box 566<br>Chesterland, OH 44026<br>Office: 440.729.8457 | | Proposed Trade Name: | NEXXT MATRIXX® System | | Common Name: | Interbody fusion system | | Device Classification: | Class II | | Regulation Name,<br>Regulation Number,<br>Product Code: | Intervertebral fusion device with bone graft, lumbar, 888.3080, MAX | | Submission Purpose: | The subject 510(k) adds an anterior lumbar interbody fusion (ALIF) device<br>to the NEXXT MATRIXX® System. | | Device Description: | The NEXXT MATRIXX® System is a collection of additively manufactured<br>spacers for cervical, lumbar/lumbosacral and thoracolumbar implantation.<br>The basic shape of these implants is a structural column to provide surgical<br>stabilization of the spine. Each device comprises an external structural<br>frame having a roughened surface (~7µm). The intervening geometric<br>lattices have pores 300-700µm.<br>The inferior/superior aspects of the NEXXT MATRIXX® open devices<br>incorporate a large vertical cavity which can be packed with bone graft<br>material. The inferior/superior aspects of the NEXXT MATRIXX® solid<br>devices are closed and do not permit the packing of bone graft within the<br>implant. The solid devices are only to be used for partial vertebral body<br>replacement. The open and solid devices are available in an assortment of<br>height, length, width and lordotic angulation combinations to accommodate<br>the individual anatomic and clinical circumstances of each patient. | | Indications for Use: | When used as a lumbar intervertebral fusion device, the NEXXT MATRIXX®<br>open devices are indicated for use at one or two contiguous levels in the<br>lumbar spine, from L2-S1, in skeletally mature patients who have had six<br>months of non-operative treatment for the treatment of degenerative disc<br>disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as<br>back pain of discogenic origin with degeneration of the disc confirmed by<br>history and radiographic studies. Additionally, the NEXXT MATRIXX®<br>lumbar implants can be used as an adjunct to fusion in patients diagnosed<br>with degenerative scoliosis. The device is intended for use with autograft<br>and/or allograft comprised of cancellous and/or corticocancellous bone graft<br>and with supplemental fixation. | | Materials: | NEXXT MATRIXX® implants are manufactured from Ti-6AI-4V ELI titanium<br>alloy (ASTM F3001). | | Primary Predicate: | NEXXT MATRIXX® System (Nexxt Spine, LLC - K171140) | | Additional Predicates: | Cascadia Interbody System (K2M Inc. - K172009) | {4}------------------------------------------------ | Performance Data: | The modified ALIF device was evaluated via dimensional analyses. The results demonstrated the performance of the modified ALIF is substantially equivalent to the predicate. | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological<br>Characteristics: | The modified NEXXT MATRIXX® System ALIF possesses the same technological characteristics as one or more of the predicate devices. These include: | | | performance (as described above), basic design (additively manufactured structural interbody), material (titanium alloy) and size (dimensions are comparable to those offered by the cleared devices). | | | Therefore the fundamental scientific technology of the modified NEXXT MATRIXX® System ALIF is the same as previously cleared devices. | | Conclusion: | The modified NEXXT MATRIXX® System ALIF possesses the same intended use and technological characteristics as the predicate devices. Therefore the modified NEXXT MATRIXX® System ALIF is substantially equivalent for its intended use. |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...