Peridot Intervertebral body fusion System

K192026 · Gbs Commonwealth Co., Ltd. · MAX · Nov 26, 2019 · Orthopedic

Device Facts

Record IDK192026
Device NamePeridot Intervertebral body fusion System
ApplicantGbs Commonwealth Co., Ltd.
Product CodeMAX · Orthopedic
Decision DateNov 26, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

Peridot Intervertebral body fusion system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. Peridot Intervertebral body fusion system is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

Device Story

Intervertebral body fusion system; provides spinal fixation aid. Device consists of PEEK cages (ASTM F2026) with Tantalum markers (ASTM F560) for radiographic positioning; titanium alloy (ASTM F136) cage holders. Machined teeth on implant ends engage vertebral endplates. Used in spinal surgery; implanted by surgeons. Requires autogenous bone graft and supplemental fixation (anterior plate system for hyperlordotic cages >20 degrees). Supplied non-sterile with dedicated instruments. Benefits include stabilization of spinal segments in DDD patients.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F2077-14 (static/fatigue compression and compression-shear), ASTM F2267-04 (subsidence), and Draft ASTM F-04.25.02.02 (expulsion). No clinical data presented.

Technological Characteristics

Materials: PEEK (ASTM F2026), Tantalum markers (ASTM F560), Titanium alloy (ASTM F136). Form factor: Intervertebral cages with machined teeth and radiographic markers. Supplied non-sterile. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade 1 spondylolisthesis or retrolisthesis. Requires use with autogenous bone graft and supplemental fixation. Patients must have failed at least 6 months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" stacked vertically. November 26, 2019 GBS Commonwealth Co., Ltd. Jimmy Kim RA C-309, Woolim Lion's Valley, 168 Gasan Digital 1-ro Geumcheon-Gu, Seoul 08507 Korea Re: K192026 Trade/Device Name: Peridot Intervertebral body fusion system Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: October 22, 2019 Received: October 28, 2019 Dear Jimmy Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Brent Showalter, Ph.D. Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K192026 ### Device Name Peridot Intervertebral body fusion system ### Indications for Use (Describe) Peridot Intervertebral body fusion system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. Peridot Intervertebral body fusion system is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="unicode-bidi:isolate; direction:ltr;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="unicode-bidi:isolate; direction:ltr;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for GBS Commonwealth. The letters "GBS" are in large, bold, blue font. Below the letters, the word "COMMONWEALTH" is written in a smaller, blue font. The background of the logo is a light gray globe. # 510(k) SUMMARY The following 510(k) summary is being submitted as required by 21 CFR 807.92(a): ## 1. Device Identification | Submitter: | GBS Commonwealth Co., Ltd. | |-----------------|-----------------------------------------------------| | | #C-309, 168 Gasan Digital 1-ro, Geumcheon-Gu Seoul, | | | South Korea | | | Phone. 82-2-6925-4469 | | | e-mail: Jimmy.kim@gbscommonwealth.com | | Contact Person: | Jimmy Kim | | Date prepared | July, 25, 2019 | | Trade Name | Peridot Intervertebral body fusion system | |----------------|-------------------------------------------------------------------------------------| | Classification | 21 CFR 888.3080 Intervertebral body fusion device,<br>Class II<br>Product Code: MAX | #### Purpose of 510(k) 2. The GBS Commonwealth Co. Ltd., here by submits this traditional 510(k): for Initial product Introduction of Peridot Intervertebral fusion system. ## 3. Predicate or legally marketed devices which are substantially equivalent - 1) Primary Predicate Device: K151140 LnK Lumbar Interbody Fusion Cage System - 2) Additional Predicate device: K092193 Shurfit Interbody device system K153783 SpaceVision OLIF K110927 MectaLIF Oblique K161379 ELSA Spacers {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for GBS Commonwealth. The letters "GBS" are in large, bold, blue font. Below the letters is the word "COMMONWEALTH" in a smaller, bold, blue font. The logo is surrounded by a white background. K151214, K161230 NuVasive Lumbar Interbody Implants #### Description of the Device 4. The Peridot Intervertebral Body Fusion System is intervertebral body fixation devices intended for use as an aid in spinal fixation. The Peridot Intervertebral fusion system is made from PEEK as per ASTM F2026 and Tantalum marker as per ASTM F560. And some cage holders are made of titanium alloy as per ASTM F136. X-ray markers system on the cages permits the identification of cage position and allows post-operative assessment. The device is available in a variety of different sizes and heights to match more closely the patient's anatomy. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates. The device is supplied non-sterile. The device consists of cages differentiated by their approach, with varying dimensions and ancillary products for placement of the cages. The device is supplied with their specific instrument. The specific instrument is supplied dedicated tray and non-sterile. The device and instrument are supplied separately. The device must be used in combination with the dedicated instrument supplied. The hyperlordotic lumbar cages (>20 degree) the form of supplemental fixation should be an anterior plate system. #### 5. Indication for Use Peridot Intervertebral body fusion system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for GBS Commonwealth. The letters "GBS" are in large, bold, blue font. Below the letters, the word "COMMONWEALTH" is written in a smaller, blue font. The background of the logo is a light gray. bone graft. Peridot Intervertebral body fusion system is to be used supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage. ### Comparison of the technological characteristics of the subject and predicate 6. devices The Peridot Intervertebral body fusion system is considered substantially equivalent to the primary predicate LnK Lumbar Interbody Fusion Cage System (K151140) and additional predicate Shurfit interbody device system (K092193). They are similar in design, material, scientific technologies and indications for use. And the Peridot Intervertebral body fusion system is considered substantially equivalent to the additional predicates SpaceVision OLIF (K153783) and MectaLIF Oblique (K110927) and ELSA Spacers (K161379) and Nuvasive Lumbar Interbody Implants (K151214, K161230). They are similar in design, material, scientific technologies and indications for use. ## 7. Performance Testing The Peridot Intervertebral body fusion system conform to class II special controls quidance document: Intervertevbral body fusion device – Document issued on : June 12, 2007. Mechanical testing includes static compression, static compression - shear, fatigue compression, fatigue compression – shear performed according to ASTM F2077-14, subsidence testing performed according to ASTM F2267-04 and , expulsion testing performed according to Draft ASTM F-04.25.02.02. Results demonstrate comparable mechanical properties to the predicate device. No clinical date has been presented. ## 8. Conclusion The Peridot Intervertebral body fusion system is substantially equivalent to legally marketed predicates.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...