← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K190563

# 4CIS PEEK PLIF Cage, 4CIS Pebble Beach PEEK PLIF Cage, 4CIS Torrey Pines PEEK TLIF Cage, 4CIS Dunes PEEK DLIF Cage, 4CIS Augusta PEEK ALIF Cage (K190563)

_Solco Biomedical Co., Ltd. · MAX · Sep 18, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K190563

## Device Facts

- **Applicant:** Solco Biomedical Co., Ltd.
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Sep 18, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Spinal Fusion Cage System is an intervertebral body fusion devices intended for use to skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Spinal Fusion Cage System is indicated to be used with autologous bone graft to facilitate fusion and are intended to be used with supplemental fixation. The device is to be used in patients who have had six months of nonoperative treatment.

## Device Story

Spinal Fusion Cage System consists of PEEK intervertebral body fusion devices; available in various shapes/sizes for posterior, lateral, anterior, or oblique surgical approaches. Devices feature hollow centers for autologous bone graft packing; tantalum markers for radiographic visualization; vertical teeth on superior/inferior surfaces to prevent subsidence/migration. Implanted by surgeons in clinical settings to restore disc space height and maintain physiological lordotic angulation. Provides mechanical support during fusion process; used in conjunction with supplemental fixation hardware. Benefits include stabilization of lumbar spinal segments and facilitation of biological synostosis.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated via static axial compression, static compression-shear, static subsidence, and dynamic axial compression testing in accordance with ASTM F2077-14 and F2267-04. No clinical data provided.

## Technological Characteristics

Materials: Polyetheretherketone (ASTM F2026) and Tantalum (ASTM F560) markers. Design: Intervertebral body fusion cage with vertical teeth for anchoring. Energy source: None (mechanical). Connectivity: None. Sterilization: Provided non-sterile, intended for single use.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- 4CIS® PEEK PLIF and TLIF Cage System ([K092162](/device/K092162.md))
- PATRIOTTM Spacers (ContinentalTM ALIF Spacer, TransContinentalTM LLIF Spacer) ([K072970](/device/K072970.md))
- ARDIS SPACER ([K073202](/device/K073202.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.

Solco Biomedical Co., Ltd. % Hwi-geun Yu Manager 154 Seotan-ro, Seotan-myeon, Pyeongtaek, Gyeonggi-do, 17704, Republic of Korea

Re: K190563

Trade/Device Name: 4CIS® PEEK PLIF Cage, 4CIS® Pebble Beach PEEK PLIF Cage, 4CIS® Torrey Pines PEEK TLIF Cage, 4CIS® Dunes PEEK DLIF Cage, 4CIS® Augusta PEEK ALIF Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 14, 2019 Received: August 19, 2019

September 18, 2019

Dear Hwi-geun Yu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-re-ulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Melissa Hall Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

#### 510(k) Number (if known)

K190563

#### Device Name

Spinal Fusion Cage System (4CIS® PEEK PLIF Cage, 4CIS® Pebble Beach PEEK PLIF Cage, 4CIS® Torrey Pines PEEK TLIF Cage , 4CIS® Dunes PEEK DLIF Cage, 4CIS® Augusta PEEK ALIF Cage)

#### Indications for Use (Describe)

The Spinal Fusion Cage System is an intervertebral body fusion devices intended for use to skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Spinal Fusion Cage System is indicated to be used with autologous bone graft to facilitate fusion and are intended to be used with supplemental fixation. The device is to be used in patients who have had six months of nonoperative treatment.

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                                                                                 |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Submitter                                                                       | Solco Biomedical Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                 | 154 Seotan-ro, Seotan-myeon, Pyeongtaek, Gyeonggi-do, 17704<br>Republic of Korea                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                 | Phone. +82-31-664-4101                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                                 | Fax. +82-31-663-8983                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Contact Person                                                                  | Hwi-geun Yu                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                                 | 154 Seotan-ro, Seotan-myeon, Pyeongtaek, Gyeonggi-do,<br>17704 Republic of Korea                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                 | Phone: +82)31-610-4091                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                                 | Fax: +82)31-663-8983                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Submission Date                                                                 | Mar 01, 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trade / Proprietary name                                                        | Spinal Fusion Cage System (4CIS® Augusta PEEK ALIF Cage,<br>4CIS® Dunes PEEK DLIF Cage, 4CIS® Torrey Pines PEEK TLIF<br>Cage, 4CIS® Pebble Beach PEEK PLIF Cage, 4CIS® PEEK PLIF<br>Cage)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Common / Usual Name                                                             | Spinal Fusion Cage System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification Name                                                             | Intervertebral body fusion device with bone graft, lumbar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification Code                                                             | MAX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Regulatory Class                                                                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation Number                                                               | 888.3080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Predicate Device                                                                | 4CIS® PEEK PLIF and TLIF Cage System (K092162, SE<br>01/06/2010) [Solco Biomedical Co., Ltd.] – Primary Predicate<br>PATRIOTTM Spacers (ContinentalTM ALIF Spacer,<br>TransContinentalTM LLIF Spacer) (K072970, SE 01/18/2008)<br>[GLOBUS MEDICAL, INC.] – Additional Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                                                 | ARDIS SPACER (K073202, SE 01/30/2008) [ABBOTT SPINE,<br>INC.] – Additional Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Description of Device                                                           | The Spinal Fusion Cage System is single component devices used<br>to restore height of disc space via posterior, lateral, anterior, oblique<br>approach and to facilitate lumbar intervertebral body fusion with<br>maintaining physiological lordotic angulation of lumbar spine. To<br>allow maximum preservation and ensure ample contact surfaces<br>with bony endplate, a variety of shapes and sizes are available and<br>each device has tantalum (ASTM F560) markers for ease of<br>visualization on radiographs. Vertical square teeth on the top and<br>the bottom surface prevent subsidence of the cage into the vertebral<br>body while they increase the anchoring and prevent slipping or<br>expulsion. To make solid fusion of intervertebral body, hollow<br>space in the implant allows autologous bone graft material to be<br>filled. The implant has safety proven structure and material<br>(Polyetheretherketone, ASTM F2026) to promote biological<br>synostosis and assures mechanical safety against load. |
| Indication for Use                                                              | The Spinal Fusion Cage System is an intervertebral body fusion<br>devices intended for use to skeletally mature patients with<br>Degenerative Disk Disease (DDD) of the lumbar spine with up to<br>Grade 1 Spondylolisthesis at one or two contiguous levels from L2-<br>S1. DDD is defined as discogenic back pain with degeneration of the<br>disc confirmed by history and radiographic studies. Spinal Fusion<br>Cage System is indicated to be used with autologous bone graft to<br>facilitate fusion and are intended to be used with supplemental<br>fixation. The device is to be used in patients who have had six<br>months of non-operative treatment.                                                                                                                                                                                                                                                                                                                                                                   |
| Comparison of<br>Technological<br>Characteristics with the<br>Predicate Devices | The subject device and all the predicates have the same or similar<br>indications for use statements. The subject device is composed of the<br>same material as the predicate devices conforming to recognized<br>industry standards for permanent implants and surgical orthopedic<br>instruments. All they have similar basic design features and functions<br>as well as those dimensions. The subject device and cited predicate<br>devices are provided non-sterile for single use only. The subject<br>device demonstrated equivalent mechanical performance to the cited<br>predicate device under the same test conditions.                                                                                                                                                                                                                                                                                                                                                                                                  |
| Performance Data                                                                | Mechanical testing (static axial compression, static compression-<br>shear, static subsidence and dynamic axial compression) was<br>conducted in accordance with ASTM F2077-14 and F2267-04.<br>Above non-clinical performance data in the form of a comprehensive<br>literature review was provided in support of substantial equivalence<br>of the subject device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{4}------------------------------------------------

{5}------------------------------------------------

| Conclusion | The overall technology characteristics and mechanical performance<br>data lead to the conclusion that the subject device is substantially<br>equivalent to legally marketed predicate devices. |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

T

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K190563](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K190563)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
