← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K181837

# Spinal Elements Ti-Bond coated devices (K181837)

_Spinal Elements, Inc. · MAX · Nov 2, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181837

## Device Facts

- **Applicant:** Spinal Elements, Inc.
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Nov 2, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Lucent Lucent intervertebral body fusion devices are intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device. Lucent Lucent Magnum+ is an intervertebral body fusion device intended for spinal fusion procedure at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device. Crystal Crystal devices are intended for spinal fusion at one or two contiguous levels in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach from the C2-C3 disc space to the C7-T1 disc space and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The device is intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems). Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices. Vertu Vertu devices are stand-alone interbody fusion devices intended for spinal fusion procedures at one or two contiguous levels from the C2/C3 disc space to the C7/T1 disc space in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices. The implant is designed to accommodate two screws should be used to ensure adequate fixation of the implant.

## Device Story

Intervertebral body fusion devices (Lucent, Lucent Magnum+, Crystal, Vertu) featuring Ti-Bond additive plasma-sprayed coating; commercially pure titanium (ASTM F1580). Devices implanted by surgeons via open anterior approach to facilitate spinal fusion; packed with bone graft; used with supplemental fixation systems. Ti-Bond coating provides macro-, micro-, and nano-surface topography to support fusion. Devices serve as structural spacers to maintain disc height and promote arthrodesis in patients with degenerative disc disease.

## Clinical Evidence

Bench testing only. Nonclinical testing demonstrated macro, micro, and nano-scale surface topography of the Ti-Bond coating.

## Technological Characteristics

Intervertebral body fusion devices with additive plasma-sprayed commercially pure titanium coating (ASTM F1580). Surface features macro, micro, and nano-topography. No changes to predicate device materials, dimensions, or operating principles.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Spinal Elements Lucent devices ([K110632](/device/K110632.md))
- Spinal Elements Lucent devices ([K150061](/device/K150061.md))
- Spinal Elements Crystal/Vertu devices ([K133218](/device/K133218.md))
- Spinal Elements Lucent XP devices ([K152011](/device/K152011.md))
- Spinal Elements Lucent Lateral devices ([K122967](/device/K122967.md))
- Spinal Elements Lucent Lateral devices ([K170235](/device/K170235.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food & Drug Administration. The Department of Health and Human Services logo is on the left and consists of a stylized caduceus. The U.S. Food & Drug Administration logo is on the right and consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 2, 2018

Spinal Elements Inc. Julie Lamothe Sr. Director Regulatory Affairs 3115 Melrose Dr., Suite 200 Carlsbad, California 92010

Re: K181837

Trade/Device Name: Spinal Elements Ti-Bond coated devices Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OVD, MQP, ODP, OVE Dated: August 3, 2018 Received: August 6, 2018

Dear Ms. Lamothe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Melissa Hall -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181837

#### Device Name

Spinal Elements Ti-Bond coated devices

#### Indications for Use (Describe)

Lucent

Lucent intervertebral body fusion devices are intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).

This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, and anterior screw and rod systems).

This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device.

#### Lucent

Lucent Magnum+ is an intervertebral body fusion device intended for spinal fusion procedure at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous nonfusion spinal surgery at the involved spinal level(s).

This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device.

## Crystal

Crystal devices are intended for spinal fusion at one or two contiguous levels in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach from the C2-C3 disc space to the C7-T1 disc space and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

The device is intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).

Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.

#### Vertu

Vertu devices are stand-alone interbody fusion devices intended for spinal fusion procedures at one or two contiguous levels from the C2/C3 disc space to the C7/T1 disc space in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.

The implant is designed to accommodate two screws. Two screws should be used to ensure adequate fixation of the implant.

{3}------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(k) Summary

## Spinal Elements Ti-Bond coated devices

## 510(k) Number: K181837

| Manufacturer Identification<br>Submitted by: | Spinal Elements, Inc.<br>3115 Melrose Dr., Suite 200<br>Carlsbad, CA 92010<br>760-607-0121                                                                                                          |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Information:                         | Julie Lamothe<br>Senior Director, Regulatory Affairs/Quality Assurance<br>Spinal Elements, Inc.<br>3115 Melrose Dr., Suite 200<br>Carlsbad, CA 92010<br>760-607-1816<br>jlamothe@spinalelements.com |
| Date Prepared:                               | October 30th 2018                                                                                                                                                                                   |
| Proprietary Name                             | Spinal Elements Ti-Bond coated devices                                                                                                                                                              |

| Device Classification     | 21 CFR Section 888.3080 |
|---------------------------|-------------------------|
| Proposed Regulatory       | Class II                |
| Class Device Product Code | MAX, OVD, MQP, ODP, OVE |

## Device Description

Spinal Elements' Ti-Bond® is an additive plasma sprayed coating of commercially pure titanium conforming to ASTM F1580. Spinal Elements' devices are coated with Ti-Bond®f on their surfaces. Spinal Elements Ti-Bond® coating presents a macro-, a micro and a nano-surface structure.

The indications for use of the devices coated with Ti-Bond® are not being modified from the ones cleared in their respective 510(k) and are as followed:

## Indications for Use

Lucent

{5}------------------------------------------------

Lucent intervertebral body fusion devices are intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).

This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).

This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device.

## Lucent

Lucent Magnum+ is an intervertebral body fusion device intended for spinal fusion procedure at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).

This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device.

## Crystal

Crystal devices are intended for spinal fusion at one or two contiguous levels in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach from the C2-C3 disc space to the C7-T1 disc space and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

The device is intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).

Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.

## Vertu

Vertu devices are stand-alone interbody fusion devices intended for spinal fusion procedures at one or two contiguous levels from the C2/C3 disc space to the C7/T1 disc space in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

{6}------------------------------------------------

K181837 Page 3 of 3

Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.

The implant is designed to accommodate two screws should be used to ensure adequate fixation of the implant.

#### Substantial Equivalence

The proposed characterization of the Ti-Bond® coating does not modify the devices previously cleared by Spinal Elements. Devices design, device specifications, material, manufacturing process, operating principle and indication for use are identical. Ti-Bond® coating is identical to the coating of the predicate devices coating: Primary Predicate: Spinal Elements Lucent devices - K110632. Additional Predicate: Spinal Elements Lucent devices - K150061 Additional Predicate: Spinal Elements Crystal/Vertu devices - K133218 Additional Predicate: Spinal Elements Lucent XP devices - K152011 Additional Predicate: Spinal Elements Lucent Lateral devices – K122967/K170235

## Technological Characteristics

There are no changes to the predicate devices coating. All technical characteristics remain identical.

#### Performance Data

In support of this 510(k) Premarket Notification, Spinal Elements has conducted nonclinical testing demonstrating that the Spinal Elements Ti-Bond® coating present a surface topography at a macro, a micro and a nano scale.

#### Conclusion

Based on the indications for use, technological characteristics, and comparison to predicate device, the subject device with the Ti-Bond® coating has been shown to be substantially equivalent to the aforementioned predicate device cleared by FDA for commercial distribution in the United States.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181837](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181837)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
