← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K181380

# LnK Lumbar Interbody Fusion Cage System (K181380)

_L&K BIOMED Co., Ltd. · MAX · Sep 11, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181380

## Device Facts

- **Applicant:** L&K BIOMED Co., Ltd.
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Sep 11, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

LnK Lumbar Interbody Fusion Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. LnK Lumbar Interbody Fusion Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

## Device Story

Intervertebral body fusion device; used as an aid in spinal fixation. Implants fabricated from PEEK-OPTIMA LT1 (ASTM F2026) with Tantalum markers (ASTM F560). Available in various sizes/heights with machined teeth to engage vertebral endplates. Surgical approaches include PLIF, T-PLIF, TLIF, ALIF, DLIF/LLIF, and ATP. Implanted in hospital/healthcare settings by surgeons. Provides structural support for fusion; used with autogenous bone graft and supplemental fixation. Benefits include stabilization of spinal segments in patients with DDD.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated via static/dynamic axial compression, compression-shear, and torsion testing per ASTM F2077, plus static subsidence testing per ASTM F2267 and expulsion testing. Finite element analysis used to validate additional components against predicate data.

## Technological Characteristics

Materials: PEEK-OPTIMA LT1 (ASTM F2026), Tantalum markers (ASTM F560). Form factor: Interbody cages with machined teeth. Approaches: PLIF, T-PLIF, TLIF, ALIF, DLIF/LLIF, ATP. Sterilization: Gamma radiation. Connectivity: None. Software: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- LnK Lumbar Interbody Fusion Cage System ([K151140](/device/K151140.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

L&K BIOMED Co., Ltd. Jihyeon Seo RA Associate #201, 202, 16-25, Dongbaekjungang-ro 16 beon-gil, Giheung-gu, Yongin-si, Gyeonggi-do, 17015 Korea

## Re: K181380

Trade/Device Name: LnK Lumbar Interbody Fusion Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 23, 2018 Received: August 23, 2018

## Dear Ms. Seo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

September 11, 2018

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181380

Device Name

LnK Lumbar Interbody Fusion Cage System

### Indications for Use (Describe)

LnK Lumbar Interbody Fusion Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. LtK Lumbar Interbody Fusion Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

| 1. | Submitter:      | Gook Jin Kang<br>L&K BIOMED Co., Ltd.<br>#201, 202 16-25, Dongbaekjungang-ro 16 beon-gil<br>Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea<br>Phone. 82-2-6717-1983<br>FAX .82-2-6717-1949 |
|----|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | Contact Person: | Jihyeon Seo                                                                                                                                                                                  |
|    | Date prepared:  | August 28, 2018                                                                                                                                                                              |

#### 2. Device Identification

| Trade Name          | LnK Lumbar Interbody Fusion Cage System             |
|---------------------|-----------------------------------------------------|
| Common Name         | Lumbar Interbody Fusion Cage System                 |
| Product Code        | MAX                                                 |
| Regulatory Class    | II                                                  |
| Classification Name | Intervertebral body fusion device (21 CFR 888.3080) |

#### Predicate or legally marketed devices which are substantially equivalent 3.

The design feature and indications for use for the subject LnK Lumbar Interbody Fusion Cage System is substantially equivalent to the following predicates:

- Unmodified device (predicate): LnK Lumbar Interbody Fusion Cage System (K151140) ●
#### Description of the Device 4.

The LnK Lumbar Interbody Fusion Cage System is intervertebral body fixation devices intended for use as an aid in spinal fixation. This system is fabricated and manufactured from PEEK-OPTIMA® LT1 as described by ASTM F2026. The Tantalum marker used for this product is made to the voluntary standard of ASTM F560. The devices are available in a variety of different sizes and heights to match more closely the patient's anatomy. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates. LnK Lumbar Interbody Fusion Cage System Implants are single-use devices. They have to be used in a healthcare setting or a hospital.

{4}------------------------------------------------

The LnK PLIF PEEK Cages are to be implanted via posterior approach. Two PLIF PEEK cages are placed on each side of the interbody space (right and left).

The LnK T-PLIF PEEK Cages which is used in transforaminal posterior lumbar interbody fusion technique. One T-PLIF cage is implanted in interbody space.

The LnK TLIF PEEK Cages are a "banana" shaped. Implants are dedicated to transforaminal approach. TLIF technique involves placing only one bone graft spacer in the middle of the interbody space, without retraction of the spinal nerves.

The LnK ALIF PEEK Cages are to be implanted via anterior approach.

The LnK DLIF PEEK Cages (LnK LLIF PEEK Cages/LnK Lateral PEEK Cages) are to be implanted via direct lateral interbody fusion approach. LnK DLIF PEEK Cage (LnK LLIF PEEK Cage/LnK Lateral PEEK Cage) can be used in an open approach and a percutaneous approach with MIS instrumentation.

The LnK ATP PEEK Cages are to be implanted via anterior lateral approach.

#### 5. Indication for use

LnK Lumbar Interbody Fusion Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. LnK Lumbar Interbody Fusion Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

{5}------------------------------------------------

| No | Item          | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|----|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    |               | LnK Lumbar Interbody Fusion<br>Cage System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | LnK Lumbar Interbody Fusion<br>Cage System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| 1  | Manufacturer  | L&K BIOMED Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | L&K BIOMED Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 2  | Material      | PEEK and Tantalum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | PEEK and Tantalum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| 3  | 510(K) Number |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K151140                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| 4  | Product Code  | MAX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | MAX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 5  | Class         | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 6  | Sterility     | Non-sterile, Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| 7  | Intended Use  | LnK Lumbar Interbody<br>Fusion Cage System is<br>indicated for intervertebral<br>body fusion procedures in<br>skeletally mature patients with<br>degenerative disc disease<br>(DDD) at one or two<br>contiguous levels from L2-S1.<br>DDD is defined as discogenic<br>back pain with degeneration<br>of the disc confirmed by<br>patient history and<br>radiographic studies. These<br>DDD patients may also have<br>up to Grade 1<br>spondylolisthesis or<br>retrolisthesis at the involved<br>level(s).This device is to be<br>used with autogenous bone<br>graft. LnK Lumbar Interbody<br>Fusion Cage System is to be<br>used with supplemental<br>fixation. Patients should have<br>at least six (6) months of non-<br>operative treatment prior to<br>treatment with an<br>intervertebral cage. | LnK Lumbar Interbody<br>Fusion Cage System is<br>indicated for intervertebral<br>body fusion procedures in<br>skeletally mature patients with<br>degenerative disc disease<br>(DDD) at one or two<br>contiguous levels from L2-S1.<br>DDD is defined as discogenic<br>back pain with degeneration<br>of the disc confirmed by<br>patient history and<br>radiographic studies. These<br>DDD patients may also have<br>up to Grade 1<br>spondylolisthesis or<br>retrolisthesis at the involved<br>level(s).This device is to be<br>used with autogenous bone<br>graft. LnK Lumbar Interbody<br>Fusion Cage System is to be<br>used with supplemental<br>fixation. Patients should have<br>at least six (6) months of non-<br>operative treatment prior to<br>treatment with an<br>intervertebral cage. |

### 6. Comparison of the technology characteristics of the device to predicate and legally marketed devices

{6}------------------------------------------------

|  | Current                                                                  | Previous                                                                  |
|--|--------------------------------------------------------------------------|---------------------------------------------------------------------------|
|  | PLIF, T-PLIF, TLIF, ALIF, DLIF, ATP PEEK Cage                            | PLIF, T-PLIF, TLIF, ALIF, DLIF Cage                                       |
|  | Width: 11~65mm<br>Length: 11~33mm<br>Height: 6~19mm<br>Lordosis: 0°~ 30° | Width: 11~65mm<br>Length: 11~32mm<br>Height: 6~19mm<br>Lordosis : 0°~ 14° |

#### 7. Performance Data

Mechanical performance of additional components is same with the predicate device (K151140). We performed finite element analysis to demonstrate that the additional components do not introduce a new worst case in terms of mechanical performance. Therefore, we leveraged the mechanical test data of the primary predicate (LnK Lumbar Interbody Fusion Cage System, K151140) to demonstrate the performance of the subject components. It is widely known that gamma sterilization does not affect the performance of the device. Therefore, only mechanical test data of non-sterilized cages was provided.

The LnK Lumbar Interbody Fusion Cage System was tested according to the relevant standards, specifically, Static and Dynamic Axial Compression, Static and Dynamic Compression-Shear Testing, Static and Dynamic Torsion Testing per ASTM F2077, Static Subsidence testing under Axial Compression, per ASTM F2267, and Expulsion Testing and .

#### Conclusion 8.

The additional components of LnK Lumbar Interbody Fusion Cage System are as safe, as effective and perform as well as the predicate device. Therefore, the additional components of LnK Lumbar Interbody Fusion Cage System are substantially equivalent to the predicate device (K151140).

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181380](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181380)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
