← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K181030

# Rubicon Lumbar Interbody Fusion System (K181030)

_Rubicon Spine, LLC · MAX · Jul 17, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181030

## Device Facts

- **Applicant:** Rubicon Spine, LLC
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Jul 17, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Rubicon Lumbar Interbody Fusion is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be use with autograft bone.

## Device Story

The Rubicon Lumbar Interbody Fusion System consists of PEEK cages designed for lumbar or lumbosacral interbody fusion. Implants feature hollow geometry for autograft bone packing, bullet-nosed leading edges, and slanted teeth to prevent migration. Superior and inferior surfaces are convex to match vertebral endplates. Available in three configurations for posterior, transforaminal, or lateral surgical approaches. Used by surgeons in clinical settings to provide structural support and facilitate fusion between vertebral bodies. The device is implanted during surgery; it requires supplemental fixation to stabilize the spinal segment. Benefits include support and correction of the disc space, promoting bone through-growth via large cephalad/caudal and side openings, and aiding visualization of the fusion mass via tantalum radiographic markers.

## Clinical Evidence

Bench testing only. Testing performed per 'Class II Special Controls Guidance Document: Intervertebral Body Fusion Device' included static and dynamic compression (ASTM F2077-11), static and dynamic compression shear (ASTM F2077-14), subsidence (ASTM F2267-04), and expulsion testing.

## Technological Characteristics

Materials: PEEK (ASTM F2026) with tantalum radiographic markers (ASTM F560). Design: Hollow cage with convex superior/inferior surfaces, slanted teeth, and large openings for bone growth. Configurations: Posterior, transforaminal, and lateral approach variants with varying lengths, widths, heights, and lordosis angles (0°, 4°, 8°). Sterilization: Not specified. Connectivity: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Integrity Spine Lumbar Interbody Fusion System ([K151819](/device/K151819.md))
- Vertebron Interbody Fusion System ([K073502](/device/K073502.md))
- Synthes Opal and Oracle Spacer ([K072791](/device/K072791.md))
- Stryker AVS PL Spacer ([K090816](/device/K090816.md))
- Stryker AVS TL Spacer ([K083661](/device/K083661.md))
- Synthes TPAL Spacer ([K100089](/device/K100089.md))
- Stryker AVS ARIA Spacer ([K101051](/device/K101051.md))
- InFront Lumbar Interbody Fusion System ([K141443](/device/K141443.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

Rubicon Spine, LLC Mike Boyer President 2955-B Bee Ridge Road Sarasota, Florida 34239

Re: K181030

Trade/Device Name: Rubicon Lumbar Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: June 25, 2018 Received: June 28, 2018

Dear Mr. Boyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

July 17, 2018

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Melissa Hall -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181030

Device Name

Rubicon Lumbar Interbody Fusion System

#### Indications for Use (Describe)

The Rubicon Lumbar Interbody Fusion is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be use with autograft bone.

| Type of Use (Select one or both, as applicable)                                                     |
|-----------------------------------------------------------------------------------------------------|
| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Prepared April 6, 2018

| 1. Company:             | Rubicon Spine, LLC<br>2955-B Bee Ridge Road<br>Sarasota, FL 34239 Tel: 888-515-<br>2532<br>FAX: 888-292-4691                           |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| 2. Contact:             | Mike Boyer<br>President<br>Rubicon Spine, LLC<br>2955-B Bee Ridge Road<br>Sarasota, FL 34239<br>Tel: 888-515-2532<br>FAX: 888-292-4691 |
| 3. Common Name          | Intervertebral Fusion Device with Bone Graft, Lumba                                                                                    |
| 4. Trade Name           | Rubicon Lumbar Interbody Fusion System                                                                                                 |
| 5. Proprietary Name:    | Rubicon Lumbar Interbody Fusion System                                                                                                 |
| 6. Classification Name: | Intervertebral Body Fusion Device (21 CFR<br>888,3080): Class II Product Code MAX                                                      |

## 7. Product Description:

The Rubicon Lumbar Interbody Fusion System consists of cages of various lengths, widths and heights, which can be inserted between two lumbar or lumbrosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The Rubicon Lumbar Interbody Fusion System is to be used with supplemental fixation. The leading edge consists of bullet nose. The cephalad/caudal opening is large to allow for bone through growth. In addition, there are 2 large holes on each side to facilitate bone and to aid in visualizing the fusion mass. The teeth are slanted away from the direction of insertion of the implant to minimize implant migration. The superior and inferior surfaces are convex to better fit the vertebral endplates.

The device is offered in 3 general configurations, specifically shaped for posterior, transforaminal or lateral surgical approaches. The posterior approach configuration is available in 508 possible sizes, with lengths from 22mm, widths from 8mm to11mm, heights from 7mm to 16mm and lordosis angles of 0°, 4° and 8°. The transforaminal approach configuration is available in 720 possible sizes, with lengths from 22mm to 36mm, widths from 9mm to 12mm, heights from 7mm to 18mm and lordosis angles of 0°, 4° and 8°. The lateral approach configuration is available in 330 possible

{4}------------------------------------------------

sizes, with lengths from 40mm to 60mm, widths from 18mm to 22mm, heights from 8mm to 18mm and lordosis angles of 0°, 4° and 8°.

The Rubicon Lumbar Interbody Fusion devices are made from PEEK (Polyetheretherketone) per ASTM F2026, with radiographic markers made from tantalum per ASTM F560.

#### 8. Indications for Use:

The Rubicon Lumbar Interbody Fusion System is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of nonoperative treatment prior to treatment with the device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be use with autograft bone.

#### 9. Summary of Technological Characteristics

The Rubicon Lumbar Interbody Fusion System consists of a series of PEEK lumbar cages. These cages are intended to be placed between two lumbar or a sacral and lumbar vertebrae using instruments provided. The proposed devices are the same as current interbody cages already on the market with the only difference being minor variations in the shape.

Rubicon has determined that the minor differences between the proposed interbody fusion cages and the predicate devices will not impact the safety or effectiveness of the interbody fusion cages for their intended use. Analysis has shown that the proposed lumbar caqes are equivalent to the predicate lumbar cages.

#### 10. Identification of Leqally Marketed Predicate Device

Documentation was provided which demonstrates that the subject Rubicon Interbody Cage System is substantially equivalent to several predicate devices that are currently on the market in the US, including the following:

Primary Predicate

- · Integrity Spine Lumbar Interbody Fusion System (K151819. November 10. 2015)
Additional Predicates

- · Vertebron Interbody Fusion System (K073502, March 24, 2008)
- · Synthes Opal and Oracle Spacer (K072791, December 26, 2007)
- · Stryker AVS PL Spacer (K090816, April 24, 2009)
- · Stryker AVS TL Spacer (K083661, February 27, 2009)
- · Synthes TPAL Spacer (K100089, May 13, 2010)
- · Stryker AVS ARIA Spacer (K101051 August 12, 2010)
- InFront Lumbar Interbody Fusion System (K141443, August 28, 2014)

{5}------------------------------------------------

#### 11. Brief Discussion of Non-Clinical Tests Submitted

The subject Rubicon Lumbar Interbody Fusion System and the predicate cage systems are identical in their indications for use, performance specifications and fundamental technological characteristics. The subject cage system was tested as described in "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" with the following tests being performed:

- Static Compression ASTM F2077-11 ●
- Dynamic Compression ASTM F2077-11 •
- o Static Compression Shear - ASTM F2077-14
- Dynamic Compression Shear ASTM F2077-14 .
- Subsidence ASTM F2267-04 ●
- . Expulsion

## 12. Conclusions from Non-Clinical Tests

Based on the results of the testing and comparisons performed, Rubicon believes that the subject Rubicon Interbody Fusion System is substantially equivalent to the predicate systems it was compared to.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181030](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K181030)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
