NuVasive® MLX® – Medial Lateral Expandable Lumbar Interbody System; NuVasive® AP Expandable XLIF System

K173025 · Nu Vasive, Incorporated · MAX · Feb 8, 2018 · Orthopedic

Device Facts

Record IDK173025
Device NameNuVasive® MLX® – Medial Lateral Expandable Lumbar Interbody System; NuVasive® AP Expandable XLIF System
ApplicantNu Vasive, Incorporated
Product CodeMAX · Orthopedic
Decision DateFeb 8, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The NuVasive MLX - Medial Lateral Expandable Lumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive MLX- Medial Lateral Expandable Lumbar Interbody System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive MLX- Medial Lateral Expandable Lumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. The NuVasive AP Expandable XLIF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle screw/rod systems) cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive AP Expandable XLIF System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive AP Expandable XLIF System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.

Device Story

Expandable interbody fusion implants; used for spinal stabilization in thoracic/lumbar regions. Implants inserted surgically by physicians to facilitate fusion; used with autogenous/allogeneic bone graft and supplemental internal fixation. Devices provide structural support to disc space; promote fusion via bone graft containment. Benefits include restoration of disc height and spinal alignment in patients with degenerative conditions. No software or automated processing involved.

Clinical Evidence

No clinical data; bench testing only. Mechanical performance data from previous submissions supported the device. Bacterial endotoxin testing (BET) conducted per ANSI/AAMI ST72:2011/(R)2016.

Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F136, ISO 5832-3), Ti-6Al-4V (ASTM F1472), Nitinol SE508 (ASTM F2063). Expandable interbody fusion design. Sterilization method not specified. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients with symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at 1-2 adjacent levels (T1-S1), including thoracic disc herniation with myelopathy/radiculopathy, or as adjunct for multilevel degenerative scoliosis. Requires 6 months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" stacked on top of each other. February 8, 2018 NuVasive, Incorporated Cynthia Adams Sr. Specialist, Regulatory Affairs 7475 Lusk Blvd. San Diego, California 92121 Re: K173025 Trade/Device Name: NuVasive® MLX® - Medial Lateral Expandable Lumbar Interbody System, NuVasive® AP Expandable XLIF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, PHM Dated: January 8, 2018 Received: January 9, 2018 Dear Ms. Adams: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Katherine D. Kavlock -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173025 ### Device Name NuVasive® MLX® - Medial Lateral Expandable Lumbar Interbody System ### Indications for Use (Describe) The NuVasive MLX - Medial Lateral Expandable Lumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The used in patients who have had at least six months of non-operative treatment. The NuVasive MLX- Medial Lateral Expandable Lumbar Interded for use in interded for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive MLX- Medial Lateral Expandable Lumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K173025 ### Device Name NuVasive® AP Expandable XLIF System ### Indications for Use (Describe) The NuVasive AP Expandable XLIF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle screw/rod systems) cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive AP Expandable XLIF System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive AP Expandable XLIF System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scollosis. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided: #### Submitted by: A. Cynthia Adams Senior Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800 Date Prepared: January 8, 2018 #### B. Device Name | Trade or Proprietary Name: | NuVasive® MLX® – Medial Lateral Expandable Lumbar Interbody System NuVasive® AP Expandable XLIF System | |----------------------------|----------------------------------------------------------------------------------------------------------| | Common or Usual Name: | Intervertebral Body Fusion Device | | Classification Name: | Intervertebral Body Fusion Device; | | Device Class: | Class II | | Classification: | 21 CFR § 888.3080 | | Product Code: | MAX, PHM | #### Predicate Devices C. The subject NuVasive MLX and APX Interbody Devices are substantially equivalent to the primary predicate device, CoRoent Thoracolumbar System (K170962) and additional predicate devices, NuVasive® MLX™ – Medial Lateral Expandable Lumbar Interbody System (K153105) and AP Expandable XLIF System (K140162). #### D. Device Description The NuVasive MLX and APX Interbody Devices are manufactured from Ti-6Al-4V ELI conforming to ASTM F136 and ISO 5832-3, Ti-6A1-4V conforming to ASTM 1472, and Nitinol SE508 conforming to ASTM F2063. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. This scope of this 510(k) includes expanded indications for use and minor add-to-file changes. {5}------------------------------------------------ | K173025 | |-------------| | Page 2 of 3 | #### Indications for Use E. The NuVasive MLX- Medial Lateral Expandable Lumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive MLX- Medial Lateral Expandable Lumbar Interbody System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive MLX- Medial Lateral Expandable Lumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. The NuVasive AP Expandable XLIF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle screw/rod systems) cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive AP Expandable XLIF System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive AP Expandable XLIF System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis. ### F. Technological Characteristics As was established in this submission, the subject NuVasive MLX and APX Interbody Devices are substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject devices were shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function and sterilization method. #### G. Performance Data Mechanical performance testing data was provided as part of the previous submissions to establish substantial equivalence for their use. The previously presented mechanical testing data are sufficient to support the subject system and does not warrant the need for additional nonclinical testing. Therefore, no new mechanical testing was performed for this 510(k) submission. {6}------------------------------------------------ | K173025 | Page 3 of 3 | |---------|-------------| |---------|-------------| Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST72:2011/(R)2016. #### H. Conclusions Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject NuVasive MLX and APX Interbody Devices have been shown to be substantially equivalent to legally marketed predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...